Recall of Device Recall Brainlab

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Brainlab AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    70657
  • Event Risk Class
    Class 2
  • Event Number
    Z-1316-2015
  • Event Initiated Date
    2015-02-16
  • Event Date Posted
    2015-03-25
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-02-28
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Accelerator, linear, medical - Product Code IYE
  • Reason
    The exactrac 6.X patient positioning system may incorrectly position the patient when using the exactrac cone beam ct (cbct) with a truebeam-specific optional subvolume-cbct.
  • Action
    BrainLab sent an FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated February 16, 2015, to all affected customers. The letters included instructions for customers to: 1) Do not use any actively re-reconstructed CBCT subvolumes from the Varian TrueBeam system with the Brainlab ExacTrac 6.x CBCT Import & Alignment Software module; 2) Import exclusively original, not modified, CBCT volumes into ExacTrac to use for patient positioning; and, 3) Continue to verify any ExacTrac CBCT-based correction using the ExacTrac X-ray verification and/or an external IGRT procedure as mandatory. Brainlab provided existing potentially affected ExacTrac v.6.x CBCT Import & Alignment Software customers (with a Varian TrueBeam system) with the product notification information. Brainlab will provide a software update with this issue solved to affected customers. Brainlab will actively contact the customer starting August 2015 to schedule the update installation. For further questions please call Customer Hotline + 49 89 99 15 68 44 or 1-800 597-5911 ( For US Customers only)

Device

  • Model / Serial
    ExacTrac versions 6.x including versions 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0.   Model/Catalog Numbers:   49971, CBCT IMPORT & ALIGNMENT SOFTWARE;  20836, EXACTRAC 6.0 VARIAN CBCT LICENSE;   The software may be contained in configured items:   48320, ET UPGRADE SOFTWARE 6.0 TO 6.1;  48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA;  49973B, ET DATA PREP / REVIEW SYSTEM;  49973A, ET DATA PREP / REVIEW SYSTEM;  49973, ET DATA PREP / REVIEW SYSTEM;  49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT;  48302, ET DATA PREP / REVIEW SYSTEM
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA and to the countries of : Australia, Austria, Canada Denmark, France, Germany, Hong Kong, Japan, Russia, Singapore, South Korea and United Kingdom.
  • Product Description
    ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures..