Recall of Device Recall Bone Marrow Biopsy Needle

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Argon Medical Devices, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    71533
  • Event Risk Class
    Class 2
  • Event Number
    Z-2189-2015
  • Event Initiated Date
    2015-06-18
  • Event Date Posted
    2015-07-21
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2016-06-23
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Instrument, biopsy - Product Code KNW
  • Reason
    Potential defect in the forming of the primary package of the blister that may compromise the sterility of the product inside the tray.
  • Action
    The recalling firm's sales force was notified of the recall on 6/17/15. The recalling firm will begin to send out Urgent Medical Device Recall letters by registered mail to affected customers on 6/18/15. Customers are instructed to inspect and quarantine and return any affected product and complete the Recal Response Form.

Device

  • Model / Serial
    Lot Numbers: 11088295, 11088596, 11088597, 11089105, 11089366, 11094182, 11094681, 11094912, 11095381, 11095844, 11095845, 11095951, 11096558, 11097278, 11097504, 11098024, 11098647, 11099173
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    US
  • Product Description
    T-Lok Bone Marrow Biopsy Needle 8G x 4 Ext Can, Part Number DBMNJ0804TL; T-Lok Bone Marrow Biopsy Needle 11G x 4 Ext Can, Part Number DBMNJ1104TL; Bone Marrow Biopsy Needle 13G x 4, Part Number DBMNJ1304
  • Manufacturer

Manufacturer