Recall of Device Recall BioPlex 2200 Syphilis IgG Control Set

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Bio-Rad Laboratories.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    57404
  • Event Risk Class
    Class 2
  • Event Number
    Z-1650-2011
  • Event Initiated Date
    2010-11-23
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-08-04
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Enzyme linked immunoabsorption assay, treponema, pallidum - Product Code LIP
  • Reason
    An upward shift in the control values can result in a control value that falls outside of the current manufacturing ranges.
  • Action
    The firm, Bio-Rad, sent a "Voluntary Field Correction: QC Ranges for Syphilis IgG Control Set Lot#46636" letter to Domestic customers via FedEx on November 24, 2010 and to the International customers via e-mail on November 23, 2010. The letter described the product, problem and actions to be taken by customers. The Customers were advised to temporarily establish their own means and acceptable ranges for Control Lot 46636 for the r15 and r47 markers. The firm will provide adjusted recommended means and acceptable ranges. Additionally, the customers were to complete and return the Customer Response Form for USA and/or the Subsidiary Response Form (Outside US) via fax to (510) 741-4846, Attn: Bio-Rad BPD-Product Support. If you have any questions about this notification, contact your local Bio-Rad Technical Support.

Device

  • Model / Serial
    Catalog #: 663-1430 Lot #: 46636
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide distribution: USA and countries including: Canada, Czech Republic, France, Germany, and Italy.
  • Product Description
    BIO-RAD BioPlex 2200 Syphilis IgG Control Set. Lot # 46636 || The BioPlex 2200 Syphilis IgG Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 Instrument and BioPlex 2200 Syphilis IgG Reagent Pack in the clinical laboratory.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Bio-Rad Laboratories, 6565 185th Ave NE, Redmond WA 98052-5039
  • Manufacturer Parent Company (2017)
  • Source
    USFDA