Recall of Device Recall BECKMAN COULTER1Q200 Series Urine Microscopy Analyzer

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    79566
  • Event Risk Class
    Class 2
  • Event Number
    Z-1362-2018
  • Event Initiated Date
    2018-03-02
  • Event Status
    Open, Classified
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Automated urinalysis system - Product Code KQO
  • Reason
    Beckman coulter has determined that there is a potential for under-reporting casts. this can occur if per high-power field (/hpf) units of measurement for casts are selected in the iq200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/lpf)* or *per microliter. this may occur during initial method validation or if settings are altered after the initial validation.
  • Action
    The customers were contacted initially by the Urgent Medical Device Recall letter on 03/028/2018 delivered by email and US Mail. Beckman Coulter will schedule a sites visit to verify the reporting units of measurement for casts for the laboratory within 90 days of this recall initiation. The use of Casts/HPF is not a common setting worldwide.

Device

  • Model / Serial
    All
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Worldwide
  • Product Description
    BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370