Recall of Device Recall BECKMAN COULTER iChemVELOCITY

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    79575
  • Event Risk Class
    Class 2
  • Event Number
    Z-1643-2018
  • Event Initiated Date
    2018-03-02
  • Event Status
    Open, Classified
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Automated urinalysis system - Product Code KQO
  • Reason
    Beckman coulter has determined that there is a potential for incorrect settings to be installed on north american ichemvelocity urine chemistry analyzers. 1. the installation of international settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations. 2. the installation of colors different from those stated in the instructions for use (colorless, straw, yellow, amber, red, blue) for output settings will result in incorrect reporting of colors. for example, if green is the color choice for the output setting instead of blue, green will be reported.
  • Action
    Beckman Coulter sent an Urgent Field Safety Notice letter dated February 22, 2018 to the customers. The letter identified the affected product, problem and actions to be taken. Beckman Coulter will schedule a site visit to verify the reporting units of measurement for the laboratory within 90 days of this recall initiation. For questions call 800-854-3633.

Device

  • Model / Serial
    All Fielded Serial Numbers with North American Configuration settings
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Malaysia, Singapore, Taiwan, Thailand, Turkey
  • Product Description
    BECKMAN COULTER iChemVELOCITY, Urine Chemistry System, Catalog Numbers: 800-3530, 800-7167, 800-7719, 800-7720, 800- || 7106, 800-3049, 800-3050, 800-3079, 800-3080, || 800-7190, 800-7713, 800-7714, 800-7715, 700- || 7176-001, 700-7177-001 || Product Usage: || The iChemVELOCITY automated urine chemistry system is an in vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterade, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChemVELOCITY strips are intended for use only with the iChemVELOCITY analyzer. In particular they are not intended for visual reading. The iChemVELOCITY is not intended to be used as a Point of Care (POC) analyzer. The measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver finction. Tests performed using the iChemVELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Beckman Coulter Inc., 250 S Kraemer Blvd, Brea CA 92821-6232
  • Manufacturer Parent Company (2017)
  • Source
    USFDA