Recall of Device Recall Beckman Coulter FP1000 Cell Preparation System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    35778
  • Event Risk Class
    Class 2
  • Event Number
    Z-1466-06
  • Event Initiated Date
    2006-06-26
  • Event Date Posted
    2006-08-31
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-04-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    cell preparation system - Product Code LXG
  • Reason
    During the cleaning cycle performed during the shutdown procedure of the beckman coulter fp1000 cell preparation system, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results.
  • Action
    A Product Corrective Action letter was sent on June 26, 2006 informing users that during the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results. To reduce the risk of incorrect results from dripping fluid, users were advised to remove all patient samples, reagents, controls, and calibrators before initiating the shutdown cycle of the Beckman Coulter FP1000. Software will be modified in the next version to prevent the possibility of dripping bleach during the shutdown cycle. The Product Corrective Action letter is to be maintained as part of the customer''s laboratory Quality System documentation. The enclosed Response Form is to be completed and returned within 10 days to the recalling firm. Customers are also advised to contact the firm or their local Beckman Coulter Representative if there are questions.

Device

Manufacturer

  • Manufacturer Address
    Beckman Coulter Inc, 200 S Kraemer Blvd, Brea CA 92822-6208
  • Source
    USFDA