Recall of Device Recall Baxter System 1000 Delivery System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Baxter Healthcare Renal Div.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    30486
  • Event Risk Class
    Class 2
  • Event Number
    Z-0395-05
  • Event Initiated Date
    2003-05-28
  • Event Date Posted
    2004-12-02
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2008-02-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, Dialysate Delivery, Single Patient - Product Code FKP
  • Reason
    The air detector may not detect air bubbles consistently at the selected limit.
  • Action
    Important Product Information letters were sent to customers on 5/28/03, informing them of the potential for the air detector not detecting air bubbles consistently. The letters instructed the accounts to measure the air detector sensor voltage output on each of their instruments, using the enclosed instructions, to determine whether the installation of an air detector attenuator connector is needed. The instructions also included directions for installation of the attenuator connector. Since the instrument requires re-qualification after the testing and installation, the accounts were given the choice to have the testing and installation done during the next service call or as part of the instrument''s next normally scheduled maintenance.

Device

  • Model / Serial
    All System 1000 units with serial numbers prior to 22224.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, the Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Korea, Lebanon, Malaysia, Morocco, the Netherlands, Oman, Palestine, Panama, Peru, the Philippines, Poland, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and the United Kingdom.
  • Product Description
    Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Baxter Healthcare Renal Div, 1620 Waukegan Rd Bldg R, Mc Gaw Park IL 60085-6730
  • Source
    USFDA