Recall of Device Recall AUTOHDL CHOLESTEROL REAGENT SET

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Pointe Scientific, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    51044
  • Event Risk Class
    Class 2
  • Event Number
    Z-1200-2009
  • Event Initiated Date
    2009-01-29
  • Event Date Posted
    2009-04-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2010-10-20
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Ldl & Vdl precipitation, Hdl - Product Code LBR
  • Reason
    The r1 reagent has microbial contamination.
  • Action
    Consignees were notified by "Urgent: Medical Device Recall" letter dated January 29, 2009, to conduct the recall to the retail level, to examine inventory and dispose of affected stocks on hand, and to complete the "Recall Return Response Form". Consignees were instructed to return their completed forms via mail, fax (1-734-483-1592) or e-mail to Pointe Scientific, Inc. (PS). Direct questions about the recall to PS Technical Support Department at 1-800-757-5313 or 1-734-487-8300.

Device

  • Model / Serial
    Lot Number: 810703, exp. 10/09.
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution -- US, Chile, Dominican Republic, Egypt, Ghana, Greece, India, Israel, Italy, Jamaica-West Indies, Korea, Malaysia, Mexico, Philippines, Poland, Romania, Serbia, Trinidad-West Indies, Turkey and Venezuela.
  • Product Description
    Pointe Scientific Liquid Auto HDL Cholesterol Reagent Set, Pointe Scientific, Inc., Canton, MI, Catalog Numbers: 10-H7545-80, 12-H7545-162, 3-H7545-L, 7-H7545-1000, 7-H7545-320, 7-H7545-40, 7-H7545-80, H7545-1000, H7545-320, H7545-40, H7545-80, HH945-240, and HH945-480. || The product is used for In vitro diagnostic for the quantitative determination of high-density lipoprotein cholesterol in human serum or plasma.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Pointe Scientific, Inc., 5449 Research Dr, Canton MI 48188
  • Manufacturer Parent Company (2017)
  • Source
    USFDA