Recall of Device Recall Artis zee Angiographic xray system

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Siemens Medical Solutions USA, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    76869
  • Event Risk Class
    Class 2
  • Event Number
    Z-1824-2017
  • Event Initiated Date
    2017-03-28
  • Event Date Posted
    2017-03-30
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-08-07
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, x-ray, angiographic - Product Code IZI
  • Reason
    Software error - as a result of a software fault in artis zee systems with software version vd11 and an a100 generator, the possibility exists that following the failure of a tube assembly focus, the artis system functions properly until the next reactivation. following reactivation, the system no longer initializes as intended.
  • Action
    Siemens mailed a Customer Safety Advisory Notice (CSAN) AX017/17/S to affected customers on 03/28/2017 to inform them of the corrective action that will be performed to prevent a possible hazard to persons and equipment. The notice also explained the underlying issue, the impact on system operation, potential risk to patients and actions they are taking to correct the issue.

Device

  • Model / Serial
    Device Model Numbers: 10280959, 10094137, 10094141, Serial numbers: 160449 147208, 153744
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Distributed to: VA, CA, WI
  • Product Description
    Artis zee, Angiographic x-ray system
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Siemens Medical Solutions USA, Inc, 40 Liberty Blvd, Malvern PA 19355-1418
  • Manufacturer Parent Company (2017)
  • Source
    USFDA