Recall of Device Recall ARROWgard Blue LargeBore MultiLumen CVC Kit

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Arrow International Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    76937
  • Event Risk Class
    Class 2
  • Event Number
    Z-2265-2017
  • Event Initiated Date
    2017-03-20
  • Event Status
    Open, Classified
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
  • Reason
    Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
  • Action
    Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.

Device

  • Model / Serial
    Material Number CDA-25123-1A
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide Distribution
  • Product Description
    ARROWg+ard Blue Large-Bore Multi-Lumen CVC Kit || The large-bore multiple-lumen catheter permits venous access to the central circulation for rapid fluid administration
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Arrow International Inc, 2400 Bernville Rd, Reading PA 19605-9607
  • Manufacturer Parent Company (2017)
  • Source