Recall of Device Recall ARIA Radiation Oncology

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Varian Medical Systems, Inc. Oncology Systems.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    60830
  • Event Risk Class
    Class 2
  • Event Number
    Z-0713-2012
  • Event Initiated Date
    2011-12-16
  • Event Date Posted
    2012-01-11
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-08-06
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Accelerator, linear, medical - Product Code IYE
  • Reason
    Varian received customer complaints involving aria oncology information systems, where the planned total dose did not calculate correctly in physicians intent. it was identified that incorrect values were being stored for planned no. fractions, leading to an unexpected calculation.
  • Action
    Varian sent an "Urgent Medical Device Correction - Urgent Field Safety Notice" letter dated December 16, 2011 to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter identified the product, problem and recommended user actions to be taken. Customers were instructed to notify all appropriate personnel of the correction notice. The notice instructs customers to complete and return the attached Proof of Notification or Receipt Verification Card to Varian Medical Systems. For questions contact your Varian Customer Support District or Reginal Manager.

Device

  • Model / Serial
    ,HIT0001 ,HIT0120 ,HIT0214 ,HIT0317 ,HIT0422 ,HIT0516 ,HIT0622 ,HIT0737 ,HIT0838 ,HIT0933 ,HIT1030 ,HIT1131 ,HIT0002 ,HIT0121 ,HIT0215 ,HIT0318M ,HIT0424 ,HIT0517 ,HIT0624 ,HIT0738M ,HIT0839M ,HIT0937 ,HIT1035 ,HIT1132M ,HIT0003 ,HIT0122 ,HIT0217 ,HIT0319M ,HIT0426 ,HIT0519 ,HIT0625 ,HIT0739M ,HIT0840 ,HIT0938 ,HIT1036 ,HIT1134 ,HIT0004M ,HIT0123M ,HIT0219 ,HIT0323 ,HIT0427 ,HIT0520M ,HIT0627 ,HIT0740 ,HIT0842 ,HIT0940 ,HIT1039 ,HIT1135 ,HIT0005 ,HIT0124 ,HIT0223 ,HIT0326 ,HIT0428 ,HIT0522 ,HIT0629M ,HIT0742M ,HIT0843 ,HIT0942 ,HIT1041 ,HIT1137 ,HIT0011 ,HIT0125 ,HIT0225 ,HIT0328M ,HIT0429M ,HIT0523 ,HIT0630 ,HIT0746 ,HIT0844 ,HIT0944 ,HIT1043 ,HIT1139 ,HIT0013 ,HIT0126 ,HIT0227 ,HIT0329 ,HIT0430 ,HIT0525M ,HIT0631 ,HIT0748 ,HIT0846 ,HIT0946 ,HIT1044 ,HIT1141 ,HIT0021 ,HIT0127 ,HIT0229 ,HIT0331 ,HIT0432M ,HIT0527 ,HIT0632M ,HIT0749 ,HIT0848 ,HIT0948A1 ,HIT1046 ,HIT1142M ,HIT0023 ,HIT0130 ,HIT0231 ,HIT0333 ,HIT0434M ,HIT0528 ,HIT0633 ,HIT0750 ,HIT0849 ,HIT0948M ,HIT1047 ,HIT1144 ,HIT0025 ,HIT0132 ,HIT0234 ,HIT0335 ,HIT0436M ,HIT0531 ,HIT0637 ,HIT0754 ,HIT0850 ,HIT0949M ,HIT1049 ,HIT1145M ,HIT0027 ,HIT0134 ,HIT0236 ,HIT0337M ,HIT0437 ,HIT0532 ,HIT0639 ,HIT0756M ,HIT0851 ,HIT0950 ,HIT1050 ,HIT1149M ,HIT0028 ,HIT0137 ,HIT0238 ,HIT0339M ,HIT0441 ,HIT0533 ,HIT0641M ,HIT0757 ,HIT0852 ,HIT0951M ,HIT1052 ,HIT1151 ,HIT0029 ,HIT0138 ,HIT0241 ,HIT0341M ,HIT0442 ,HIT0535 ,HIT0642 ,HIT0759 ,HIT0854 ,HIT0952 ,HIT1053 ,HIT1152 ,HIT0030 ,HIT0139M ,HIT0242 ,HIT0342 ,HIT0443 ,HIT0537 ,HIT0643 ,HIT0760 ,HIT0856 ,HIT0954 ,HIT1054 ,HIT1155 ,HIT0031M ,HIT0140 ,HIT0244 ,HIT0344 ,HIT0444 ,HIT0539 ,HIT0645 ,HIT0762 ,HIT0857 ,HIT0956 ,HIT1056M ,HIT1157 ,HIT0032 ,HIT0143M ,HIT0246 ,HIT0347 ,HIT0446 ,HIT0540 ,HIT0647 ,HIT0764 ,HIT0858 ,HIT0958 ,HIT1058 ,HIT1159 ,HIT0038 ,HIT0144 ,HIT0247 ,HIT0348 ,HIT0448 ,HIT0541 ,HIT0649 ,HIT0766 ,HIT0860 ,HIT0959 ,HIT1059 ,HIT1160 ,HIT0040 ,HIT0145 ,HIT0249 ,HIT0350 ,HIT0450 ,HIT0542 ,HIT0650 ,HIT0767 ,HIT0863 ,HIT0960 ,HIT1060 ,HIT1163 ,HIT0041 ,HIT0147 ,HIT0251M ,HIT0351 ,HIT0452 ,HIT0544 ,HIT0651 ,HIT0769 ,HIT0864 ,HIT0962 ,HIT1062 ,HIT1164 ,HIT0043M ,HIT0148 ,HIT0254 ,HIT0353 ,HIT0454 ,HIT0547 ,HIT0652 ,HIT0771 ,HIT0866 ,HIT0963 ,HIT1064 ,HIT1166 ,HIT0045 ,HIT0150 ,HIT0255 ,HIT0355M ,HIT0456 ,HIT0549 ,HIT0653 ,HIT0773M ,HIT0868 ,HIT0965 ,HIT1065 ,HIT1168 ,HIT0047 ,HIT0152 ,HIT0259 ,HIT0357 ,HIT0458 ,HIT0551 ,HIT0655 ,HIT0776 ,HIT0869 ,HIT0967 ,HIT1066 ,HIT1170 ,HIT0049 ,HIT0154 ,HIT0261 ,HIT0359 ,HIT0459 ,HIT0553 ,HIT0657 ,HIT0777M ,HIT0870M ,HIT0968 ,HIT1069 ,HIT1172M ,HIT0050 ,HIT0155 ,HIT0262 ,HIT0361M ,HIT0461 ,HIT0555 ,HIT0658 ,HIT0779M ,HIT0872 ,HIT0970 ,HIT1070 ,HIT1174 ,HIT0052 ,HIT0157 ,HIT0264M ,HIT0363 ,HIT0462 ,HIT0556 ,HIT0660 ,HIT0781 ,HIT0874 ,HIT0972 ,HIT1072 ,HIT1176 ,HIT0055 ,HIT0158 ,HIT0265 ,HIT0366M ,HIT0464 ,HIT0558 ,HIT0662 ,HIT0782 ,HIT0876 ,HIT0975 ,HIT1074 ,HIT1179M ,HIT0057 ,HIT0162 ,HIT0266 ,HIT0370 ,HIT0465 ,HIT0563 ,HIT0663 ,HIT0784M ,HIT0879 ,HIT0977 ,HIT1076M ,HIT1181 ,HIT0059 ,HIT0165M ,HIT0267M ,HIT0372 ,HIT0467 ,HIT0565 ,HIT0665 ,HIT0787 ,HIT0880 ,HIT0978 ,HIT1077 ,HIT1184A1 ,HIT0061 ,HIT0167 ,HIT0268 ,HIT0373M ,HIT0468M ,HIT0566 ,HIT0667 ,HIT0789 ,HIT0881 ,HIT0979 ,HIT1080 ,HIT1184M ,HIT0065 ,HIT0170M ,HIT0269 ,HIT0374 ,HIT0470 ,HIT0568M ,HIT0668 ,HIT0790 ,HIT0884 ,HIT0981 ,HIT1082 ,HIT1186 ,HIT0066 ,HIT0171 ,HIT0270 ,HIT0376 ,HIT0474M ,HIT0570 ,HIT0671 ,HIT0792 ,HIT0885 ,HIT0982M ,HIT1084 ,HIT1188 ,HIT0068 ,HIT0174 ,HIT0274 ,HIT0378 ,HIT0475 ,HIT0572 ,HIT0673 ,HIT0793 ,HIT0886 ,HIT0983M ,HIT1085 ,HIT1189 ,HIT0070 ,HIT0177M ,HIT0279 ,HIT0380 ,HIT0477M ,HIT0573 ,HIT0681 ,HIT0795 ,HIT0887 ,HIT0986M ,HIT1087 ,HIT1190 ,HIT0072 ,HIT0178 ,HIT0280 ,HIT0381 ,HIT0479 ,HIT0579M ,HIT0682 ,HIT0797 ,HIT0889 ,HIT0988 ,HIT1089 ,HIT1200 ,HIT0074M ,HIT0179 ,HIT0282M ,HIT0383 ,HIT0480M ,HIT0581 ,HIT0683 ,HIT0799 ,HIT0892 ,HIT0989 ,HIT1090 ,HIT1202 ,HIT0076 ,HIT0181 ,HIT0284 ,HIT0385 ,HIT0481 ,HIT0583 ,HIT0685M ,HIT0801 ,HIT0894 ,HIT0991 ,HIT1091 ,HIT1203 ,HIT0077 ,HIT0182 ,HIT0286 ,HIT0388 ,HIT0483 ,HIT0585M ,HIT0687 ,HIT0803 ,HIT0896 ,HIT0993 ,HIT1092 ,HIT1206 ,HIT0079M ,HIT0183 ,HIT0288 ,HIT0390 ,HIT0485 ,HIT0587 ,HIT0689M ,HIT0804 ,HIT0898 ,HIT0994 ,HIT1093 ,HIT1207 ,HIT0082 ,HIT0185 ,HIT0290 ,HIT0393 ,HIT0486 ,HIT0588 ,HIT0691M ,HIT0805 ,HIT0900M ,HIT0996 ,HIT1095 ,HIT1208 ,HIT0084 ,HIT0187 ,HIT0292 ,HIT0394 ,HIT0487 ,HIT0589M ,HIT0693M ,HIT0807 ,HIT0901M ,HIT0998 ,HIT1098 ,HIT1211 ,HIT0088 ,HIT0190M ,HIT0293 ,HIT0396 ,HIT0488 ,HIT0590 ,HIT0696 ,HIT0808M ,HIT0905M ,HIT1000 ,HIT1102 ,HIT1212 ,HIT0090 ,HIT0191M ,HIT0295 ,HIT0397 ,HIT0489M ,HIT0592 ,HIT0697 ,HIT0809 ,HIT0907 ,HIT1001 ,HIT1103 ,HIT1213 ,HIT0092 ,HIT0193 ,HIT0297 ,HIT0399M ,HIT0490 ,HIT0593 ,HIT0699 ,HIT0812 ,HIT0909 ,HIT1002 ,HIT1106 ,HIT1215 ,HIT0095 ,HIT0195M ,HIT0300 ,HIT0401 ,HIT0492 ,HIT0594 ,HIT0701 ,HIT0813 ,HIT0910 ,HIT1004 ,HIT1107 ,HIT1216 ,HIT0097M ,HIT0196 ,HIT0301 ,HIT0402 ,HIT0494 ,HIT0598 ,HIT0703M ,HIT0814M ,HIT0911 ,HIT1006 ,HIT1110M ,HIT1220M ,HIT0100 ,HIT0198M ,HIT0303 ,HIT0404 ,HIT0495 ,HIT0599 ,HIT0705 ,HIT0816 ,HIT0912 ,HIT1007 ,HIT1111 ,HIT1225 ,HIT0101 ,HIT0199 ,HIT0304 ,HIT0405 ,HIT0498 ,HIT0600 ,HIT0711M ,HIT0818M ,HIT0913 ,HIT1010 ,HIT1112 ,HIT1227 ,HIT0103 ,HIT0200 ,HIT0305 ,HIT0406 ,HIT0499 ,HIT0602 ,HIT0712 ,HIT0820 ,HIT0916 ,HIT1013 ,HIT1113 ,HIT1229 ,HIT0105M ,HIT0202 ,HIT0306 ,HIT0407 ,HIT0501 ,HIT0604 ,HIT0714 ,HIT0823 ,HIT0918M ,HIT1014 ,HIT1115 ,HIT1231 ,HIT0107 ,HIT0203 ,HIT0307 ,HIT0408 ,HIT0503M ,HIT0606M ,HIT0716 ,HIT0825 ,HIT0920 ,HIT1015 ,HIT1116 ,HIT1234 ,HIT0109 ,HIT0204 ,HIT0308M ,HIT0409 ,HIT0505 ,HIT0608 ,HIT0718 ,HIT0827 ,HIT0922 ,HIT1016 ,HIT1117 ,HIT1235M ,HIT0111 ,HIT0205 ,HIT0310 ,HIT0411M ,HIT0507 ,HIT0610 ,HIT0721 ,HIT0828 ,HIT0924 ,HIT1018 ,HIT1119M ,HIT1236 ,HIT0112 ,HIT0206 ,HIT0311 ,HIT0413 ,HIT0509 ,HIT0613 ,HIT0723M ,HIT0829 ,HIT0925 ,HIT1020M ,HIT1120 ,HIT1241 ,HIT0114 ,HIT0207 ,HIT0312 ,HIT0414 ,HIT0510 ,HIT0614 ,HIT0726 ,HIT0830 ,HIT0926 ,HIT1022 ,HIT1121 ,HIT1243 ,HIT0115 ,HIT0210 ,HIT0313 ,HIT0416 ,HIT0511 ,HIT0615M ,HIT0728 ,HIT0832 ,HIT0928 ,HIT1024 ,HIT1122 ,HIT1245 ,HIT0116M ,HIT0211 ,HIT0314 ,HIT0417 ,HIT0512 ,HIT0616 ,HIT0729 ,HIT0833 ,HIT0929 ,HIT1025 ,HIT1124 ,HIT1247 ,HIT0117 ,HIT0212 ,HIT0315 ,HIT0419 ,HIT0513 ,HIT0617 ,HIT0733 ,HIT0835 ,HIT0930 ,HIT1027 ,HIT1128 ,HIT1250 ,HIT0118 ,HIT0213 ,HIT0316 ,HIT0420 ,HIT0515 ,HIT0620 ,HIT0736 ,HIT0837 ,HIT0932M ,HIT1029 ,HIT1129 ,HIT1252M   ,HIT1260 ,HIT1385M ,HIT1490 ,HIT1599 ,HIT1704 ,HIT1810 ,HIT1914 ,HIT2012 ,HIT2118 ,HIT2238 ,HIT2367 ,HIT2501 ,HIT1264 ,HIT1386 ,HIT1492 ,HIT1600M ,HIT1707 ,HIT1811 ,HIT1916 ,HIT2013 ,HIT2119 ,HIT2241 ,HIT2369 ,HIT2502 ,HIT1267 ,HIT1389 ,HIT1493 ,HIT1601 ,HIT1709 ,HIT1813 ,HIT1918 ,HIT2014 ,HIT2123 ,HIT2245 ,HIT2370 ,HIT2504 ,HIT1269 ,HIT1390 ,HIT1494 ,HIT1602M ,HIT1711 ,HIT1817 ,HIT1921 ,HIT2016 ,HIT2124 ,HIT2247M ,HIT2374 ,HIT2505 ,HIT1271 ,HIT1392 ,HIT1496 ,HIT1604 ,HIT1712 ,HIT1820 ,HIT1922 ,HIT2017 ,HIT2126 ,HIT2252 ,HIT2376 ,HIT2506 ,HIT1273 ,HIT1394 ,HIT1497M ,HIT1606 ,HIT1714 ,HIT1822 ,HIT1924M ,HIT2018 ,HIT2129M ,HIT2256 ,HIT2378 ,HIT2508 ,HIT1275 ,HIT1396 ,HIT1498 ,HIT1608M ,HIT1716 ,HIT1824 ,HIT1925 ,HIT2020 ,HIT2131 ,HIT2263 ,HIT2380 ,HIT2509 ,HIT1277 ,HIT1397 ,HIT1499M ,HIT1609M ,HIT1718 ,HIT1825 ,HIT1926 ,HIT2021 ,HIT2132 ,HIT2265 ,HIT2382 ,HIT2510 ,HIT1278 ,HIT1398 ,HIT1500M ,HIT1610M ,HIT1719M ,HIT1827 ,HIT1928 ,HIT2022 ,HIT2135 ,HIT2267M ,HIT2383 ,HIT2512M ,HIT1281M ,HIT1399 ,HIT1505 ,HIT1611M ,HIT1720 ,HIT1829 ,HIT1930 ,HIT2023M ,HIT2136 ,HIT2269 ,HIT2385 ,HIT2513 ,HIT1284 ,HIT1400 ,HIT1506 ,HIT1612M ,HIT1723M ,HIT1830M ,HIT1932 ,HIT2025 ,HIT2138 ,HIT2271 ,HIT2388 ,HIT2516 ,HIT1288 ,HIT1401 ,HIT1507 ,HIT1613 ,HIT1724 ,HIT1831M ,HIT1934M ,HIT2027 ,HIT2140 ,HIT2272 ,HIT2390 ,HIT2519 ,HIT1290 ,HIT1402 ,HIT1509 ,HIT1614 ,HIT1725 ,HIT1833 ,HIT1938M ,HIT2028 ,HIT2145 ,HIT2273M ,HIT2392 ,HIT2521 ,HIT1293 ,HIT1404 ,HIT1510 ,HIT1616 ,HIT1726 ,HIT1837M ,HIT1941 ,HIT2029 ,HIT2146 ,HIT2275M ,HIT2394 ,HIT2522 ,HIT1294 ,HIT1406 ,HIT1511 ,HIT1617 ,HIT1729 ,HIT1838 ,HIT1942 ,HIT2031 ,HIT2149 ,HIT2277 ,HIT2397 ,HIT2523 ,HIT1296 ,HIT1408 ,HIT1512M ,HIT1619 ,HIT1731 ,HIT1839 ,HIT1943M ,HIT2034 ,HIT2151 ,HIT2278 ,HIT2399 ,HIT2526 ,HIT1298 ,HIT1410 ,HIT1514 ,HIT1620M ,HIT1733 ,HIT1840 ,HIT1944M ,HIT2036 ,HIT2152 ,HIT2279 ,HIT2400 ,HIT2528 ,HIT1300 ,HIT1412 ,HIT1516M ,HIT1622 ,HIT1735 ,HIT1842 ,HIT1945 ,HIT2037M ,HIT2154 ,HIT2282 ,HIT2402 ,HIT2529 ,HIT1301 ,HIT1413 ,HIT1517M ,HIT1623 ,HIT1736 ,HIT1843 ,HIT1947 ,HIT2039 ,HIT2156 ,HIT2283M ,HIT2404 ,HIT2531 ,HIT1303 ,HIT1414M ,HIT1518M ,HIT1624 ,HIT1738 ,HIT1845 ,HIT1948 ,HIT2041 ,HIT2158 ,HIT2285 ,HIT2405 ,HIT2533 ,HIT1306 ,HIT1415 ,HIT1520 ,HIT1625 ,HIT1741 ,HIT1848 ,HIT1951 ,HIT2042 ,HIT2159 ,HIT2287 ,HIT2409M ,HIT2534 ,HIT1308 ,HIT1418 ,HIT1521 ,HIT1626 ,HIT1742M ,HIT1850 ,HIT1952 ,HIT2044 ,HIT2160 ,HIT2289 ,HIT2411 ,HIT2535 ,HIT1310 ,HIT1419 ,HIT1522 ,HIT1628 ,HIT1744 ,HIT1852 ,HIT1953 ,HIT2045M ,HIT2162 ,HIT2291M ,HIT2414M ,HIT2537 ,HIT1311 ,HIT1420 ,HIT1523M ,HIT1629M ,HIT1745 ,HIT1854 ,HIT1955M ,HIT2047M ,HIT2164 ,HIT2293 ,HIT2417 ,HIT2539 ,HIT1313 ,HIT1421 ,HIT1526 ,HIT1632 ,HIT1747M ,HIT1855 ,HIT1956 ,HIT2049 ,HIT2166 ,HIT2296 ,HIT2420 ,HIT2541 ,HIT1315 ,HIT1423 ,HIT1529 ,HIT1633 ,HIT1748 ,HIT1857 ,HIT1958 ,HIT2051 ,HIT2168 ,HIT2298 ,HIT2421 ,HIT2543 ,HIT1317 ,HIT1426 ,HIT1531 ,HIT1635 ,HIT1750 ,HIT1859 ,HIT1961 ,HIT2053 ,HIT2171 ,HIT2299M ,HIT2422 ,HIT2544 ,HIT1318 ,HIT1428 ,HIT1536 ,HIT1639 ,HIT1751 ,HIT1860 ,HIT1963 ,HIT2056 ,HIT2172 ,HIT2301 ,HIT2424 ,HIT2546 ,HIT1321 ,HIT1429 ,HIT1537 ,HIT1642 ,HIT1753 ,HIT1861 ,HIT1964 ,HIT2058 ,HIT2173 ,HIT2303 ,HIT2425 ,HIT2554 ,HIT1323 ,HIT1430M ,HIT1539 ,HIT1644 ,HIT1757 ,HIT1863 ,HIT1965 ,HIT2063 ,HIT2178 ,HIT2305 ,HIT2429 ,HIT2555 ,HIT1325 ,HIT1432 ,HIT1541 ,HIT1649 ,HIT1758 ,HIT1864 ,HIT1967 ,HIT2064 ,HIT2180 ,HIT2314 ,HIT2430 ,HIT2557 ,HIT1327 ,HIT1433 ,HIT1544 ,HIT1651 ,HIT1760 ,HIT1865 ,HIT1968 ,HIT2065 ,HIT2182 ,HIT2316 ,HIT2431 ,HIT2559 ,HIT1329 ,HIT1434M ,HIT1546 ,HIT1653 ,HIT1761 ,HIT1866 ,HIT1969 ,HIT2067 ,HIT2183 ,HIT2317 ,HIT2432M ,HIT2561 ,HIT1331M ,HIT1435 ,HIT1548 ,HIT1655 ,HIT1763 ,HIT1867 ,HIT1970M ,HIT2068 ,HIT2185 ,HIT2318 ,HIT2434 ,HIT2563M ,HIT1332 ,HIT1436 ,HIT1549 ,HIT1658 ,HIT1764M ,HIT1870M ,HIT1973 ,HIT2069 ,HIT2187 ,HIT2319 ,HIT2435 ,HIT2564 ,HIT1336 ,HIT1438 ,HIT1551 ,HIT1660M ,HIT1765 ,HIT1871M ,HIT1975 ,HIT2074 ,HIT2188 ,HIT2320 ,HIT2437 ,HIT2565 ,HIT1338 ,HIT1440 ,HIT1553 ,HIT1662 ,HIT1767 ,HIT1874 ,HIT1976 ,HIT2076 ,HIT2190M ,HIT2321 ,HIT2439 ,HIT2567 ,HIT1339M ,HIT1443 ,HIT1554 ,HIT1665 ,HIT1768 ,HIT1875 ,HIT1977 ,HIT2078 ,HIT2191 ,HIT2323 ,HIT2441 ,HIT2568 ,HIT1342 ,HIT1444 ,HIT1556 ,HIT1667 ,HIT1770 ,HIT1876 ,HIT1978 ,HIT2079 ,HIT2193 ,HIT2325 ,HIT2443 ,HIT2570A1 ,HIT1343 ,HIT1446 ,HIT1557 ,HIT1669 ,HIT1771 ,HIT1877 ,HIT1980M ,HIT2080 ,HIT2196 ,HIT2326 ,HIT2444 ,HIT2570M ,HIT1344M ,HIT1448M ,HIT1558 ,HIT1671 ,HIT1773 ,HIT1879 ,HIT1981 ,HIT2081 ,HIT2198 ,HIT2327 ,HIT2445 ,HIT2572 ,HIT1346M ,HIT1450 ,HIT1560M ,HIT1674 ,HIT1778 ,HIT1880 ,HIT1982 ,HIT2083 ,HIT2199 ,HIT2329 ,HIT2446 ,HIT2574A1 ,HIT1347 ,HIT1451 ,HIT1563 ,HIT1675 ,HIT1780 ,HIT1882 ,HIT1986 ,HIT2085 ,HIT2202 ,HIT2331 ,HIT2448 ,HIT2574A2 ,HIT1348 ,HIT1452 ,HIT1565M ,HIT1677 ,HIT1781 ,HIT1885 ,HIT1987 ,HIT2087 ,HIT2204 ,HIT2333 ,HIT2449M ,HIT2574M ,HIT1350 ,HIT1454 ,HIT1567 ,HIT1678 ,HIT1782 ,HIT1887 ,HIT1988 ,HIT2088 ,HIT2205 ,HIT2335 ,HIT2451 ,HIT2576 ,HIT1353M ,HIT1456 ,HIT1571 ,HIT1679 ,HIT1787 ,HIT1888 ,HIT1989 ,HIT2089 ,HIT2208M ,HIT2336 ,HIT2453 ,HIT2577 ,HIT1354 ,HIT1458M ,HIT1573 ,HIT1680 ,HIT1788 ,HIT1889 ,HIT1991 ,HIT2091 ,HIT2209 ,HIT2337M ,HIT2455 ,HIT2584 ,HIT1356 ,HIT1460 ,HIT1574 ,HIT1681 ,HIT1790M ,HIT1890 ,HIT1994 ,HIT2093 ,HIT2212 ,HIT2341 ,HIT2461 ,HIT2588 ,HIT1359 ,HIT1462 ,HIT1575 ,HIT1683 ,HIT1794 ,HIT1893 ,HIT1996 ,HIT2095 ,HIT2215 ,HIT2342 ,HIT2463M ,HIT2589 ,HIT1360 ,HIT1464 ,HIT1577 ,HIT1684 ,HIT1795 ,HIT1894 ,HIT1997M ,HIT2098 ,HIT2218 ,HIT2343 ,HIT2464 ,HIT2591 ,HIT1361 ,HIT1466 ,HIT1579M ,HIT1686 ,HIT1797 ,HIT1895 ,HIT1998 ,HIT2100 ,HIT2219 ,HIT2345 ,HIT2467 ,HIT2593 ,HIT1363M ,HIT1467 ,HIT1580 ,HIT1687 ,HIT1798 ,HIT1896 ,HIT2000 ,HIT2101M ,HIT2221 ,HIT2346M ,HIT2469 ,HIT2596 ,HIT1365M ,HIT1469 ,HIT1581 ,HIT1688 ,HIT1799 ,HIT1898 ,HIT2002 ,HIT2102 ,HIT2222 ,HIT2348 ,HIT2471M ,HIT2598 ,HIT1369 ,HIT1470M ,HIT1582 ,HIT1690 ,HIT1800 ,HIT1899 ,HIT2003 ,HIT2105 ,HIT2223 ,HIT2350 ,HIT2473 ,HIT2600 ,HIT1371 ,HIT1473 ,HIT1584 ,HIT1691 ,HIT1801 ,HIT1900 ,HIT2004 ,HIT2107 ,HIT2225 ,HIT2353 ,HIT2475 ,HIT2602 ,HIT1373 ,HIT1475 ,HIT1585 ,HIT1693 ,HIT1802 ,HIT1901 ,HIT2005M ,HIT2109 ,HIT2226 ,HIT2355 ,HIT2477M ,HIT2603 ,HIT1375 ,HIT1478 ,HIT1587M ,HIT1696 ,HIT1803 ,HIT1903 ,HIT2006 ,HIT2110 ,HIT2228 ,HIT2357 ,HIT2480 ,HIT2605 ,HIT1376 ,HIT1481 ,HIT1589M ,HIT1697M ,HIT1804M ,HIT1904M ,HIT2007 ,HIT2111M ,HIT2229 ,HIT2358 ,HIT2482 ,HIT2606 ,HIT1379 ,HIT1483 ,HIT1591 ,HIT1698 ,HIT1805 ,HIT1905 ,HIT2008 ,HIT2112 ,HIT2231 ,HIT2359 ,HIT2484 ,HIT2609 ,HIT1380 ,HIT1486 ,HIT1592 ,HIT1700 ,HIT1806 ,HIT1907 ,HIT2009 ,HIT2113M ,HIT2233 ,HIT2361 ,HIT2491 ,HIT2612 ,HIT1381 ,HIT1488 ,HIT1594 ,HIT1701M ,HIT1808 ,HIT1911 ,HIT2010M ,HIT2115 ,HIT2236 ,HIT2362M ,HIT2493 ,HIT2615 ,HIT1383 ,HIT1489 ,HIT1597 ,HIT1703 ,HIT1809 ,HIT1913 ,HIT2011 ,HIT2117 ,HIT2237M ,HIT2364 ,HIT2498 ,HIT2620  ,HIT2621 ,HIT2741 ,HIT2891 ,HIT3018 ,HIT3140 ,HIT3274 ,HIT3404M ,HIT3527 ,HIT3671 ,HIT3796 ,HIT3936 ,HIT4112 ,HIT2622 ,HIT2742 ,HIT2892 ,HIT3019 ,HIT3145 ,HIT3277 ,HIT3406 ,HIT3529 ,HIT3675 ,HIT3798 ,HIT3939 ,HIT4114 ,HIT2623A1 ,HIT2744 ,HIT2894M ,HIT3020 ,HIT3147 ,HIT3279 ,HIT3407 ,HIT3531 ,HIT3677 ,HIT3799 ,HIT3941 ,HIT4116 ,HIT2623M ,HIT2747 ,HIT2895 ,HIT3021 ,HIT3153 ,HIT3282 ,HIT3409 ,HIT3535 ,HIT3679 ,HIT3801 ,HIT3943 ,HIT4117 ,HIT2625 ,HIT2749 ,HIT2896 ,HIT3025 ,HIT3154 ,HIT3288 ,HIT3411 ,HIT3537 ,HIT3680 ,HIT3802 ,HIT3945 ,HIT4118M ,HIT2627 ,HIT2752 ,HIT2897 ,HIT3027 ,HIT3157M ,HIT3290 ,HIT3412 ,HIT3539 ,HIT3682 ,HIT3803 ,HIT3946 ,HIT4122 ,HIT2629 ,HIT2754 ,HIT2899 ,HIT3028 ,HIT3160 ,HIT3293 ,HIT3413 ,HIT3541 ,HIT3684 ,HIT3804 ,HIT3952 ,HIT4123 ,HIT2630 ,HIT2757 ,HIT2900 ,HIT3032 ,HIT3162 ,HIT3295 ,HIT3423 ,HIT3546 ,HIT3686 ,HIT3807 ,HIT3954 ,HIT4129 ,HIT2633 ,HIT2759 ,HIT2902 ,HIT3033 ,HIT3165 ,HIT3296 ,HIT3424 ,HIT3549 ,HIT3690 ,HIT3813 ,HIT3955 ,HIT4130 ,HIT2635 ,HIT2761 ,HIT2903 ,HIT3034 ,HIT3168 ,HIT3298M ,HIT3425 ,HIT3551 ,HIT3691 ,HIT3815 ,HIT3957 ,HIT4132 ,HIT2637 ,HIT2763 ,HIT2906 ,HIT3039 ,HIT3172 ,HIT3299M ,HIT3427 ,HIT3555 ,HIT3692 ,HIT3816 ,HIT3962 ,HIT4133 ,HIT2639 ,HIT2764 ,HIT2908M ,HIT3040 ,HIT3173 ,HIT3302 ,HIT3428 ,HIT3560 ,HIT3697 ,HIT3822 ,HIT3964 ,HIT4136 ,HIT2640M ,HIT2766 ,HIT2911 ,HIT3042 ,HIT3174 ,HIT3304 ,HIT3432 ,HIT3563M ,HIT3700 ,HIT3823 ,HIT3965 ,HIT4139 ,HIT2643 ,HIT2767 ,HIT2915 ,HIT3044 ,HIT3177 ,HIT3308 ,HIT3434 ,HIT3564 ,HIT3702 ,HIT3826 ,HIT3967 ,HIT4144 ,HIT2645 ,HIT2768 ,HIT2918M ,HIT3045 ,HIT3179 ,HIT3309 ,HIT3435 ,HIT3573 ,HIT3705 ,HIT3828 ,HIT3969 ,HIT4146 ,HIT2647 ,HIT2769 ,HIT2919 ,HIT3048 ,HIT3184 ,HIT3311 ,HIT3438 ,HIT3575 ,HIT3706 ,HIT3829 ,HIT3970 ,HIT4149 ,HIT2649M ,HIT2770 ,HIT2923 ,HIT3050 ,HIT3186 ,HIT3314 ,HIT3439 ,HIT3578 ,HIT3708 ,HIT3832 ,HIT3973 ,HIT4151 ,HIT2650 ,HIT2772 ,HIT2929 ,HIT3052 ,HIT3187 ,HIT3315 ,HIT3441 ,HIT3579 ,HIT3710 ,HIT3837 ,HIT3975 ,HIT4154 ,HIT2652 ,HIT2774 ,HIT2930 ,HIT3054 ,HIT3188 ,HIT3319 ,HIT3446 ,HIT3580 ,HIT3713 ,HIT3841 ,HIT3977 ,HIT4156 ,HIT2655 ,HIT2777 ,HIT2932 ,HIT3055 ,HIT3192M ,HIT3320 ,HIT3448 ,HIT3582 ,HIT3714 ,HIT3843 ,HIT3980 ,HIT4158 ,HIT2657 ,HIT2779 ,HIT2934 ,HIT3057 ,HIT3194 ,HIT3322 ,HIT3449M ,HIT3583 ,HIT3717 ,HIT3844 ,HIT3984 ,HIT4159 ,HIT2658 ,HIT2780 ,HIT2935 ,HIT3058 ,HIT3195 ,HIT3324 ,HIT3454 ,HIT3585 ,HIT3718 ,HIT3845 ,HIT3988 ,HIT4160M ,HIT2659 ,HIT2783 ,HIT2937 ,HIT3059 ,HIT3198 ,HIT3326 ,HIT3455 ,HIT3587 ,HIT3719 ,HIT3847 ,HIT3989 ,HIT4163M ,HIT2660 ,HIT2788 ,HIT2939 ,HIT3061 ,HIT3199 ,HIT3327 ,HIT3456 ,HIT3588 ,HIT3720 ,HIT3849 ,HIT3992 ,HIT4165 ,HIT2662 ,HIT2790 ,HIT2941 ,HIT3063 ,HIT3201 ,HIT3328 ,HIT3457 ,HIT3591 ,HIT3722 ,HIT3852 ,HIT3994 ,HIT4167 ,HIT2666 ,HIT2793 ,HIT2942 ,HIT3064 ,HIT3204 ,HIT3329 ,HIT3459 ,HIT3592 ,HIT3724 ,HIT3854 ,HIT3997 ,HIT4168 ,HIT2668 ,HIT2794 ,HIT2943 ,HIT3070 ,HIT3206 ,HIT3333 ,HIT3460 ,HIT3595 ,HIT3725 ,HIT3856 ,HIT3998 ,HIT4169 ,HIT2670 ,HIT2795 ,HIT2945 ,HIT3072 ,HIT3209 ,HIT3336 ,HIT3461 ,HIT3598M ,HIT3726 ,HIT3860 ,HIT4001 ,HIT4171 ,HIT2676 ,HIT2797 ,HIT2950 ,HIT3073 ,HIT3216 ,HIT3338 ,HIT3463 ,HIT3601 ,HIT3727 ,HIT3862 ,HIT4003 ,HIT4178 ,HIT2681 ,HIT2812 ,HIT2951 ,HIT3074 ,HIT3218 ,HIT3340 ,HIT3464 ,HIT3603 ,HIT3731 ,HIT3863 ,HIT4007 ,HIT4181 ,HIT2686 ,HIT2813 ,HIT2953 ,HIT3077 ,HIT3219 ,HIT3341 ,HIT3466M ,HIT3605 ,HIT3732 ,HIT3865 ,HIT4009 ,HIT4188 ,HIT2687 ,HIT2816 ,HIT2954 ,HIT3081 ,HIT3220 ,HIT3342 ,HIT3468 ,HIT3607 ,HIT3735 ,HIT3867 ,HIT4010 ,HIT4190 ,HIT2688 ,HIT2818 ,HIT2961 ,HIT3084 ,HIT3221M ,HIT3344 ,HIT3469 ,HIT3609 ,HIT3736 ,HIT3869 ,HIT4011 ,HIT4191 ,HIT2689 ,HIT2823 ,HIT2965 ,HIT3087 ,HIT3222 ,HIT3349 ,HIT3470M ,HIT3612 ,HIT3738 ,HIT3871 ,HIT4013 ,HIT4193 ,HIT2690 ,HIT2825 ,HIT2966 ,HIT3088M ,HIT3224 ,HIT3354 ,HIT3473 ,HIT3614 ,HIT3739M ,HIT3873 ,HIT4014 ,HIT4196 ,HIT2692 ,HIT2826 ,HIT2968 ,HIT3089 ,HIT3226 ,HIT3355 ,HIT3474 ,HIT3615 ,HIT3740 ,HIT3875 ,HIT4015 ,HIT4204 ,HIT2693 ,HIT2828 ,HIT2970 ,HIT3091 ,HIT3227 ,HIT3356 ,HIT3476 ,HIT3618 ,HIT3741M ,HIT3876 ,HIT4016 ,HIT4206M ,HIT2694 ,HIT2830 ,HIT2973 ,HIT3092 ,HIT3230 ,HIT3359 ,HIT3478 ,HIT3624 ,HIT3742 ,HIT3879 ,HIT4017 ,HIT4232 ,HIT2695 ,HIT2831 ,HIT2974 ,HIT3093 ,HIT3232 ,HIT3360 ,HIT3479 ,HIT3625 ,HIT3744 ,HIT3881 ,HIT4020 ,HIT4234 ,HIT2699 ,HIT2834 ,HIT2975 ,HIT3095 ,HIT3233 ,HIT3362M ,HIT3481 ,HIT3627 ,HIT3752 ,HIT3883 ,HIT4027 ,HIT4235 ,HIT2701M ,HIT2835 ,HIT2977 ,HIT3097 ,HIT3234 ,HIT3364 ,HIT3483 ,HIT3628 ,HIT3753 ,HIT3887 ,HIT4030 ,HIT4237 ,HIT2702 ,HIT2836 ,HIT2980 ,HIT3100 ,HIT3236 ,HIT3365 ,HIT3484 ,HIT3629 ,HIT3754 ,HIT3888 ,HIT4034 ,HIT4240 ,HIT2703 ,HIT2840 ,HIT2981 ,HIT3102 ,HIT3237 ,HIT3366 ,HIT3486 ,HIT3630 ,HIT3755 ,HIT3891M ,HIT4038 ,HIT4242 ,HIT2704 ,HIT2842 ,HIT2982 ,HIT3106 ,HIT3239 ,HIT3367 ,HIT3489 ,HIT3631 ,HIT3757 ,HIT3896 ,HIT4043 ,HIT4245 ,HIT2705 ,HIT2843 ,HIT2984 ,HIT3109 ,HIT3240 ,HIT3369 ,HIT3492 ,HIT3633 ,HIT3758M ,HIT3897 ,HIT4045 ,HIT4251 ,HIT2706 ,HIT2852 ,HIT2985 ,HIT3110 ,HIT3241 ,HIT3370 ,HIT3494 ,HIT3635 ,HIT3759 ,HIT3899 ,HIT4047 ,HIT4252 ,HIT2710M ,HIT2854M ,HIT2987 ,HIT3112M ,HIT3242 ,HIT3373 ,HIT3495 ,HIT3638 ,HIT3764M ,HIT3901 ,HIT4051 ,HIT4253 ,HIT2711 ,HIT2856M ,HIT2988 ,HIT3116M ,HIT3244 ,HIT3376 ,HIT3498 ,HIT3640 ,HIT3766 ,HIT3902 ,HIT4053 ,HIT4254 ,HIT2715 ,HIT2860 ,HIT2991 ,HIT3117 ,HIT3245 ,HIT3379 ,HIT3499 ,HIT3641 ,HIT3767 ,HIT3904 ,HIT4057 ,HIT4255 ,HIT2716 ,HIT2863 ,HIT2992 ,HIT3118 ,HIT3250 ,HIT3384 ,HIT3501 ,HIT3642 ,HIT3769 ,HIT3906 ,HIT4062M ,HIT4257 ,HIT2717 ,HIT2865 ,HIT2993 ,HIT3119 ,HIT3251 ,HIT3386 ,HIT3503 ,HIT3644 ,HIT3771 ,HIT3908 ,HIT4065 ,HIT4262 ,HIT2720 ,HIT2867 ,HIT2995 ,HIT3121 ,HIT3254 ,HIT3388 ,HIT3505 ,HIT3646 ,HIT3775 ,HIT3910 ,HIT4071 ,HIT4263 ,HIT2722 ,HIT2868 ,HIT2997 ,HIT3125 ,HIT3257 ,HIT3391 ,HIT3509 ,HIT3648 ,HIT3778 ,HIT3911 ,HIT4074M ,HIT4264 ,HIT2724 ,HIT2875 ,HIT2999 ,HIT3126 ,HIT3259 ,HIT3393M ,HIT3513 ,HIT3652 ,HIT3780 ,HIT3913 ,HIT4075 ,HIT4270 ,HIT2726 ,HIT2878 ,HIT3000 ,HIT3127 ,HIT3260 ,HIT3394 ,HIT3514 ,HIT3653 ,HIT3782 ,HIT3918 ,HIT4078 ,HIT4271 ,HIT2728 ,HIT2879 ,HIT3004 ,HIT3128 ,HIT3261 ,HIT3395 ,HIT3515 ,HIT3658 ,HIT3784 ,HIT3920 ,HIT4080 ,HIT4272 ,HIT2731 ,HIT2880 ,HIT3006 ,HIT3130 ,HIT3262 ,HIT3396 ,HIT3516 ,HIT3659 ,HIT3786 ,HIT3923 ,HIT4082 ,HIT4273 ,HIT2732 ,HIT2883 ,HIT3008 ,HIT3131M ,HIT3264 ,HIT3397 ,HIT3517 ,HIT3660 ,HIT3787 ,HIT3925 ,HIT4087 ,HIT4276 ,HIT2737 ,HIT2884 ,HIT3010 ,HIT3134 ,HIT3265 ,HIT3400 ,HIT3519 ,HIT3661 ,HIT3788 ,HIT3927M ,HIT4090 ,HIT4277 ,HIT2738 ,HIT2887 ,HIT3012 ,HIT3135 ,HIT3267 ,HIT3401 ,HIT3521M ,HIT3664 ,HIT3790 ,HIT3928 ,HIT4105 ,HIT4282 ,HIT2739 ,HIT2888 ,HIT3013 ,HIT3136 ,HIT3268 ,HIT3402 ,HIT3524 ,HIT3667 ,HIT3792 ,HIT3931 ,HIT4107 ,HIT4283 ,HIT2740 ,HIT2889 ,HIT3016 ,HIT3139 ,HIT3270 ,HIT3403 ,HIT3526 ,HIT3668 ,HIT3795 ,HIT3934 ,HIT4108 ,HIT4285  ,HIT4286 ,HIT4313 ,HIT4348 ,HIT4391 ,HIT4436 ,HIT4455 ,HIT4287 ,HIT4317 ,HIT4349 ,HIT4392 ,HIT4437  ,HIT4289 ,HIT4323 ,HIT4350 ,HIT4394 ,HIT4438  ,HIT4293 ,HIT4324M ,HIT4356 ,HIT4395 ,HIT4439M  ,HIT4299 ,HIT4326 ,HIT4357 ,HIT4396 ,HIT4449  ,HIT4303 ,HIT4328 ,HIT4358 ,HIT4397 ,HIT4450  ,HIT4306 ,HIT4331 ,HIT4360 ,HIT4400 ,HIT4451  ,HIT4309 ,HIT4332 ,HIT4381 ,HIT4402 ,HIT4452  ,HIT4310 ,HIT4334 ,HIT4385 ,HIT4404 ,HIT4453  ,HIT4311 ,HIT4335 ,HIT4390 ,HIT4434 ,HIT4454   ***12-12-12, AMENDED INFO ***: ,HIT0004R1 ,HIT0434R1 ,HIT0818R2 ,HIT1414R3 ,HIT1723R1 ,HIT2337R2 ,HIT3131R1 ,HIT4084 ,HIT4426 ,HIT4539 ,HIT4650 ,HIT4774 ,HIT4989 ,HIT0004R2 ,HIT0436R1 ,HIT0839R1 ,HIT1414R4 ,HIT1723R2 ,HIT2343 ,HIT3192R1 ,HIT4092 ,HIT4428 ,HIT4541 ,HIT4652 ,HIT4776 ,HIT4998 ,HIT0031R1 ,HIT0459M ,HIT0854R1 ,HIT1434R1 ,HIT1742R1 ,HIT2362R1 ,HIT3199M ,HIT4097 ,HIT4429 ,HIT4543 ,HIT4655 ,HIT4778 ,HIT4999 ,HIT0043R1 ,HIT0459R1 ,HIT0870R1 ,HIT1434R2 ,HIT1747R1 ,HIT2362R2 ,HIT3199R1 ,HIT4110 ,HIT4431 ,HIT4545 ,HIT4658M ,HIT4788 ,HIT5000M ,HIT0043R2 ,HIT0474R1 ,HIT0870R2 ,HIT1434R3 ,HIT1747R2 ,HIT2362R3 ,HIT3221R1 ,HIT4120 ,HIT4439R1 ,HIT4547 ,HIT4658R1 ,HIT4789 ,HIT5000R1 ,HIT0043R3 ,HIT0474R2 ,HIT0870R3 ,HIT1434R4 ,HIT1790R1 ,HIT2362R4 ,HIT3221R2 ,HIT4121M ,HIT4440 ,HIT4549 ,HIT4658R2 ,HIT4791 ,HIT5000R2 ,HIT0049R1 ,HIT0474R3 ,HIT0870R4 ,HIT1448R1 ,HIT1790R2 ,HIT2362R5 ,HIT3221R3 ,HIT4121R1 ,HIT4442 ,HIT4551 ,HIT4659 ,HIT4793 ,HIT5000R3 ,HIT0074R1 ,HIT0476 ,HIT0900R1 ,HIT1458R1 ,HIT1804R1 ,HIT2362R6 ,HIT3221R4 ,HIT4134 ,HIT4443 ,HIT4552 ,HIT4660 ,HIT4796 ,HIT5000R4 ,HIT0079R2 ,HIT0477R1 ,HIT0905R1 ,HIT1458R2 ,HIT1831R1 ,HIT2409R1 ,HIT3221R5 ,HIT4147 ,HIT4446 ,HIT4553 ,HIT4664 ,HIT4802M ,HIT5000R5 ,HIT0079R3 ,HIT0479M ,HIT0918R1 ,HIT1469M ,HIT1837R1 ,HIT2414R1 ,HIT3292 ,HIT4156M ,HIT4447 ,HIT4558 ,HIT4665 ,HIT4802R1 ,HIT5000R6 ,HIT0079R4 ,HIT0479R1 ,HIT0925M ,HIT1469R1 ,HIT1870R1 ,HIT2432R1 ,HIT3298R1 ,HIT4156R1 ,HIT4448 ,HIT4559 ,HIT4666 ,HIT4804M ,HIT5000R7 ,HIT0079R5 ,HIT0480R1 ,HIT0925R1 ,HIT1497R1 ,HIT1871R1 ,HIT2432R2 ,HIT3298R2 ,HIT4160R1 ,HIT4459 ,HIT4560M ,HIT4669 ,HIT4804R1 ,HIT5002 ,HIT0097R1 ,HIT0480R2 ,HIT0925R2 ,HIT1499R1 ,HIT1871R2 ,HIT2432R3 ,HIT3298R3 ,HIT4163R1 ,HIT4462 ,HIT4560R1 ,HIT4671 ,HIT4809 ,HIT5006 ,HIT0097R2 ,HIT0489R1 ,HIT0932R1 ,HIT1500R1 ,HIT1904R1 ,HIT2432R4 ,HIT3299R1 ,HIT4163R2 ,HIT4464 ,HIT4561M ,HIT4674 ,HIT4813 ,HIT5010 ,HIT0097R3 ,HIT0503R1 ,HIT0948R1 ,HIT1500R2 ,HIT1924R1 ,HIT2449R1 ,HIT3299R3 ,HIT4177 ,HIT4465 ,HIT4561R1 ,HIT4675 ,HIT4817 ,HIT5011 ,HIT0097R4 ,HIT0520R1 ,HIT0949R1 ,HIT1500R3 ,HIT1932M ,HIT2471R1 ,HIT3362R1 ,HIT4184 ,HIT4467 ,HIT4561R2 ,HIT4677 ,HIT4819 ,HIT5013 ,HIT0097R5 ,HIT0525R1 ,HIT0949R2 ,HIT1500R4 ,HIT1932R1 ,HIT2496 ,HIT3375 ,HIT4201 ,HIT4468 ,HIT4564 ,HIT4679 ,HIT4822 ,HIT5014 ,HIT0105R1 ,HIT0525R2 ,HIT0951R1 ,HIT1512R1 ,HIT1934R1 ,HIT2570R1 ,HIT3393R1 ,HIT4206R1 ,HIT4470 ,HIT4568 ,HIT4681 ,HIT4827 ,HIT5015 ,HIT0123R1 ,HIT0544M ,HIT0982R1 ,HIT1516R1 ,HIT1938R1 ,HIT2570R2 ,HIT3404R1 ,HIT4206R2 ,HIT4471 ,HIT4571 ,HIT4683 ,HIT4828 ,HIT5022 ,HIT0123R3 ,HIT0544R1 ,HIT0986R1 ,HIT1518R1 ,HIT1943R1 ,HIT2570R3 ,HIT3425M ,HIT4209 ,HIT4472M ,HIT4573 ,HIT4686 ,HIT4830 ,HIT5032 ,HIT0139R1 ,HIT0568R1 ,HIT1020R1 ,HIT1518R2 ,HIT1944R1 ,HIT2574R2 ,HIT3425R1 ,HIT4212 ,HIT4472R1 ,HIT4574 ,HIT4689 ,HIT4835 ,HIT5035 ,HIT0143R1 ,HIT0579R1 ,HIT1056R1 ,HIT1560R1 ,HIT1955R1 ,HIT2574R3 ,HIT3438R1 ,HIT4221 ,HIT4472R2 ,HIT4577 ,HIT4692 ,HIT4836 ,HIT5039 ,HIT0165R1 ,HIT0583M ,HIT1056R2 ,HIT1560R2 ,HIT1970R1 ,HIT2574R4 ,HIT3449R1 ,HIT4229 ,HIT4472R3 ,HIT4578 ,HIT4695 ,HIT4837 ,HIT5041 ,HIT0177R1 ,HIT0583R1 ,HIT1076R1 ,HIT1560R3 ,HIT1980A1 ,HIT2574R5 ,HIT3449R2 ,HIT4247 ,HIT4474 ,HIT4579 ,HIT4698 ,HIT4840 ,HIT5042 ,HIT0190R1 ,HIT0585R1 ,HIT1076R2 ,HIT1565R1 ,HIT1980R1 ,HIT2574R6 ,HIT3449R3 ,HIT4292 ,HIT4475 ,HIT4581 ,HIT4700 ,HIT4846 ,HIT5044 ,HIT0190R2 ,HIT0589R1 ,HIT1076R3 ,HIT1565R2 ,HIT1985 ,HIT2574R7 ,HIT3449R4 ,HIT4294 ,HIT4478 ,HIT4584 ,HIT4702 ,HIT4849 ,HIT5046 ,HIT0190R3 ,HIT0606R1 ,HIT1076R4 ,HIT1565R3 ,HIT1997R1 ,HIT2574R8 ,HIT3466R1 ,HIT4297 ,HIT4479 ,HIT4586 ,HIT4707 ,HIT4851 ,HIT5058 ,HIT0195R1 ,HIT0615R1 ,HIT1076R5 ,HIT1579R1 ,HIT2005R1 ,HIT2623R1 ,HIT3470R1 ,HIT4301 ,HIT4484M ,HIT4587 ,HIT4708 ,HIT4852 ,HIT5062 ,HIT0198R1 ,HIT0615R2 ,HIT1132R1 ,HIT1579R2 ,HIT2005R2 ,HIT2623R2 ,HIT3521R1 ,HIT4308 ,HIT4484R1 ,HIT4589 ,HIT4709 ,HIT4855 ,HIT5065 ,HIT0217 ,HIT0622M ,HIT1142R1 ,HIT1600R1 ,HIT2005R3 ,HIT2633M ,HIT3521R2 ,HIT4322 ,HIT4484R2 ,HIT4590 ,HIT4712 ,HIT4857 ,HIT5066 ,HIT0251R1 ,HIT0622R1 ,HIT1142R2 ,HIT1600R2 ,HIT2005R4 ,HIT2633R1 ,HIT3525 ,HIT4324R1 ,HIT4485 ,HIT4594 ,HIT4713 ,HIT4859 ,HIT5067 ,HIT0255M ,HIT0629R1 ,HIT1142R3 ,HIT1600R3 ,HIT2010R1 ,HIT2640R1 ,HIT3563R1 ,HIT4324R2 ,HIT4486 ,HIT4596 ,HIT4715 ,HIT4864 ,HIT5068 ,HIT0255R1 ,HIT0629R2 ,HIT1142R4 ,HIT1602R1 ,HIT2023R1 ,HIT2640R2 ,HIT3597 ,HIT4325 ,HIT4487 ,HIT4597 ,HIT4717 ,HIT4866 ,HIT5072 ,HIT0264R1 ,HIT0635 ,HIT1145R1 ,HIT1608R1 ,HIT2045R1 ,HIT2649R1 ,HIT3598R1 ,HIT4341 ,HIT4488 ,HIT4598 ,HIT4719 ,HIT4867 ,HIT5073 ,HIT0264R2 ,HIT0641R1 ,HIT1145R2 ,HIT1608R2 ,HIT2047 ,HIT2649R2 ,HIT3598R2 ,HIT4346 ,HIT4491 ,HIT4602 ,HIT4722 ,HIT4873 ,HIT5074 ,HIT0264R3 ,HIT0685R1 ,HIT1145R3 ,HIT1608R3 ,HIT2101R1 ,HIT2700 ,HIT3598R3 ,HIT4347 ,HIT4493 ,HIT4611 ,HIT4723 ,HIT4878 ,HIT5082 ,HIT0267R1 ,HIT0689R1 ,HIT1172R1 ,HIT1608R4 ,HIT2111R1 ,HIT2701R1 ,HIT3636 ,HIT4355 ,HIT4494 ,HIT4612 ,HIT4725 ,HIT4889 ,HIT5084 ,HIT0279 ,HIT0693R1 ,HIT1216M ,HIT1609R1 ,HIT2111R2 ,HIT2701R2 ,HIT3739R1 ,HIT4359M ,HIT4496 ,HIT4613 ,HIT4726 ,HIT4890 ,HIT5086 ,HIT0282R1 ,HIT0703R1 ,HIT1216R1 ,HIT1610R2 ,HIT2111R3 ,HIT2710R1 ,HIT3739R2 ,HIT4359R1 ,HIT4497 ,HIT4614 ,HIT4727M ,HIT4899 ,HIT5090 ,HIT0290M ,HIT0711R1 ,HIT1220R1 ,HIT1611R1 ,HIT2113R1 ,HIT2710R2 ,HIT3739R3 ,HIT4363 ,HIT4499 ,HIT4615 ,HIT4727R1 ,HIT4902 ,HIT5091 ,HIT0290R1 ,HIT0723R1 ,HIT1220R2 ,HIT1612R1 ,HIT2113R2 ,HIT2714 ,HIT3741R1 ,HIT4365 ,HIT4502 ,HIT4621 ,HIT4728 ,HIT4904 ,HIT5095 ,HIT0308R1 ,HIT0738R1 ,HIT1220R3 ,HIT1620R1 ,HIT2129R1 ,HIT2743 ,HIT3758R1 ,HIT4367 ,HIT4503 ,HIT4624 ,HIT4730 ,HIT4909 ,HIT5098 ,HIT0318R1 ,HIT0739R1 ,HIT1220R4 ,HIT1620R2 ,HIT2129R2 ,HIT2745 ,HIT3764R1 ,HIT4371 ,HIT4504 ,HIT4625 ,HIT4734 ,HIT4911 ,HIT5100 ,HIT0319R1 ,HIT0742R1 ,HIT1220R5 ,HIT1620R3 ,HIT2129R3 ,HIT2746 ,HIT3788M ,HIT4373 ,HIT4505 ,HIT4626 ,HIT4737 ,HIT4916 ,HIT5101 ,HIT0320R1 ,HIT0742R2 ,HIT1235R1 ,HIT1620R5 ,HIT2129R4 ,HIT2854R1 ,HIT3788R1 ,HIT4376 ,HIT4514 ,HIT4629 ,HIT4739 ,HIT4923 ,HIT5102 ,HIT0328R1 ,HIT0756R1 ,HIT1235R2 ,HIT1620R6 ,HIT2129R5 ,HIT2856R1 ,HIT3848 ,HIT4377 ,HIT4516 ,HIT4631 ,HIT4740 ,HIT4927 ,HIT5106 ,HIT0328R2 ,HIT0773R1 ,HIT1252R1 ,HIT1629R1 ,HIT2190R1 ,HIT2908R1 ,HIT3891R1 ,HIT4387 ,HIT4518 ,HIT4632 ,HIT4748 ,HIT4933 ,HIT5110 ,HIT0355R1 ,HIT0777R1 ,HIT1331R1 ,HIT1629R2 ,HIT2247R1 ,HIT2918R1 ,HIT3983 ,HIT4398 ,HIT4519 ,HIT4634 ,HIT4749 ,HIT4935 ,HIT5116 ,HIT0361R1 ,HIT0779R1 ,HIT1339R1 ,HIT1697R1 ,HIT2247R2 ,HIT2935M ,HIT4024 ,HIT4400M ,HIT4520 ,HIT4635 ,HIT4756 ,HIT4936 ,HIT5126 ,HIT0361R2 ,HIT0784R1 ,HIT1339R2 ,HIT1697R2 ,HIT2273R1 ,HIT2935R1 ,HIT4034M ,HIT4400R1 ,HIT4523 ,HIT4636 ,HIT4757 ,HIT4940 ,HIT5135 ,HIT0366R1 ,HIT0784R2 ,HIT1339R3 ,HIT1701R1 ,HIT2283R1 ,HIT2954M ,HIT4049 ,HIT4405 ,HIT4524 ,HIT4638 ,HIT4758 ,HIT4941 ,HIT5137 ,HIT0373R1 ,HIT0784R3 ,HIT1344R1 ,HIT1701R2 ,HIT2299R1 ,HIT2954R1 ,HIT4059 ,HIT4406 ,HIT4526M ,HIT4640 ,HIT4759 ,HIT4945 ,HIT5151 ,HIT0399R1 ,HIT0784R4 ,HIT1344R2 ,HIT1701R3 ,HIT2299R3 ,HIT3002 ,HIT4062R1 ,HIT4408 ,HIT4526R1 ,HIT4642 ,HIT4760 ,HIT4949 ,HIT5176 ,HIT0411R1 ,HIT0795M ,HIT1346R1 ,HIT1701R4 ,HIT2299R4 ,HIT3072M ,HIT4063 ,HIT4410 ,HIT4527 ,HIT4644 ,HIT4764 ,HIT4953  ,HIT0411R2 ,HIT0795R1 ,HIT1353R1 ,HIT1704M ,HIT2299R6 ,HIT3072R1 ,HIT4067 ,HIT4414 ,HIT4530 ,HIT4646 ,HIT4765 ,HIT4972  ,HIT0414M ,HIT0808R1 ,HIT1365R1 ,HIT1704R1 ,HIT2336M ,HIT3112R1 ,HIT4074R1 ,HIT4419 ,HIT4531 ,HIT4647 ,HIT4769 ,HIT4973  ,HIT0414R1 ,HIT0814R1 ,HIT1414R1 ,HIT1719R1 ,HIT2336R1 ,HIT3116R1 ,HIT4074R2 ,HIT4421 ,HIT4535 ,HIT4648 ,HIT4771 ,HIT4977  ,HIT0429R1 ,HIT0818R1 ,HIT1414R2 ,HIT1719R2 ,HIT2337R1 ,HIT3123 ,HIT4074R3 ,HIT4423 ,HIT4538 ,HIT4649 ,HIT4773 ,HIT4985***
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA (nationwide)
  • Product Description
    Varian brand ARIA Radiation Oncology, ARIA versions 10.0.39 and below, Reference/FSCA Identifier: CP-06070, Model Number: HIT, Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA || Product Usage: || The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA also includes optional tools to manage the oncology departments clinical and business information.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Varian Medical Systems, Inc. Oncology Systems, 911 Hansen Way, Palo Alto CA 94304-1028
  • Manufacturer Parent Company (2017)
  • Source
    USFDA