Recall of Device Recall AIRCAST VENAFLOW ELITE SYSTEM

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by DJO, LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    65493
  • Event Risk Class
    Class 2
  • Event Number
    Z-1661-2013
  • Event Initiated Date
    2013-05-28
  • Event Date Posted
    2013-07-03
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-02-13
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Sleeve, limb, compressible - Product Code JOW
  • Reason
    Djo, llc has recently identified a product safety issue with the battery assembly for the venaflow 30b-b and 30 bi-b products. there is a potential for battery leakage which could result in thermal meltdown (fire).
  • Action
    DJO sent a Customer Notification letter dated May 28, 2013 to Domestic customers via USPS certified return receipt and to International customers via UPS on June 14, 2013. The Customer Notification informed customers of the product safety issue, product description, and instructions to follow. DJO plans to install replacement batteries at no charge. For questions contact your local DJO Customer Service Representative.

Device

  • Model / Serial
    Model Numbers: 30B-B and 30BI-B.  Serial Numbers: E00026  E00029  E00030  E00031  E00036  E00039  E00045  E00050  E00051  E00052  E00053  E00056  E00057  E00058  E00060  E00143  E00147  E00151  E00156  E00163  E00169  E00175  E00178  E00184  E00191  E00200  E00201  E00203  E00205  E00206  E00211  E00211  E00222  E00227  E00234  E00236  E00238  E00239  E00240  E00241  E00242  E00243  E00245  E00246  E00247  E00248  E00250  E00252  E00258  E00260  E00261  E00262  E00264  E00267  E00269  E00271  E00272  E00273  E00274  E00276  E00281  E00283  E00289  E00296  E00298  E00301  E00303  E00305  E00306  E00308  E00313  E00314  E00315  E00317  E00321  E00322  E00326  E00329  E00330  E00332  E00335  E00336  E00337  E00339  E00340  E00342  E00344  E00347  E00348  E00351  E00352  E00357  E00358  E00363  E00368  E00370  E00371  E00373  E00375  E00376  E00377  E00379  E00380  E00381  E00384  E00385  E00388  E00390  E00391  E00392  E00393  E00394  E00395  E00396  E00397  E00398  E00399  E00400  E00401  E00402  E00403  E00404  E00406  E00407  E00408  E00409  E00410  E00411  E00412  E00413  E00414  E00415  E00416  E00417  E00418  E00419  E00420  E00422  E00426  E00429  E00430  E00432  E00433  E00434  E00436  E00437  E00439  E00492  E00503  E00516  E00529  E00533  E00540  E00541  E00543  E00546  E00548  E00559  E00563  E00570  E00576  E00577  E00580  E00582  E00583  E00584  E00585  E00589  E00590  E00591  E00594  E00595  E00596  E00598  E00599  E00600  E00602  E00604  E00605  E00606  E00607  E00608  E00609  E00610  E00611  E00612  E00613  E00614  E00615  E00616  E00617  E00618  E00619  E00620  E00621  E00622  E00623  E00628  E00629  E00630  E00631  E00632  E00634  E00635  E00636  E00637  E00638  E00639  E00642  E00643  E00644  E00645  E00646  E00647  E00649  E00651  E00652  E00656  E00657  E00658  E00659  E00660  E00661  E00662  E00663  E00664  E00665  E00666  E00669  E00670  E00672  E00673  E00675  E00676  E00677  E00679  E00680  E00682  E00683  E00686  E00687  E00688  E00689  E00692  E00693  E00696  E00698  E00699  E00700  E00701  E00703  E00704  E00705  E00706  E00707  E00708  E00710  E00712  E00714  E00715  E00716  E00718  E00719  E00720  E00723  E00724  E00725  E00727  E00728  E00730  E00733 E00734 E00735 E00736 E00737 E00739 E00740 E00741 E00743 E00744 E00745 E00746 E00747 E00748 E00749 E00750 E00752 E00759 E00760 E00761 E00762 E00763 E00764 E00765 E00766 E00767 E00768 E00769 E00770 E00771 E00772 E00773 E00774 E00775 E00778 E00779 E00786 E00787 E00789 E00791 E00795 E00796 E00797 E00798 E00800 E00806 E00809 E00811 E00812 E00813 E00815 E00822 E00839 E01016 E01018 E01019 E01020 E01021 E01022 E01023 E01024 E01025 E01027 E01028 E01029 E01030 E01032 E01033 E01035 E01036 E01037 E01038 E01041 E01042 E01043 E01044 E01045 E01046 E01047 E01048 E01051 E01053 E01054 E01055 E01057 E01058 E01059 E01060 E01061 E01062 E01063 E01064 E01065 E01066 E01067 E01068 E01071 E01072 E01074 E01076 E01079 E01080 E01081 E01082 E01083 E01084 E01085 E01086 E01087 E01088 E01089 E01090 E01092 E01093 E01094 E01095 E01097 E01098 E01099 E01101 E01102 E01103 E01104 E01105 E01166 E01167 E01168 E01169 E01170 E01171 E01172 E01173 E01174 E01175 E01176 E01177 E01179 E01180 E01183 E01186 E01188 E01189 E01190 E01191 E01192 E01193 E01194 E01195 E01196 E01198 E01200 E01201 E01204 E01205 E01206 E01207 E01208 E01209 E01211 E01212 E01213 E01217 E01219 E01221 E01224 E01226 E01228 E01229 E01230 E01231 E01232 E01234 E01235 E01238 E01239 E01240 E01242 E01245 E01247 E01248 E01249 E01250 E01251 E01252 E01253 E01254 E01456 E01457 E01458 E01459 E01460 E01461 E01462 E01463 E01464 E01466 E01467 E01468 E01469 E01470 E01472 E01474 E01475 E01476 E01477 E01479 E01480 E01481 E01482 E01483 E01484 E01485 E01486 E01488 E01490 E01491 E01494 E01495 E01496 E01500 E01503 E01505 E01507 E01508 E01511 E01512 E01513 E01514 E01516 E01517 E01518 E01520 E01522 E01524 E01525 E01529 E01531 E01532 E01535 E01536 E01537 E01540 E01542 E01543 E01544 E01545 E01546 E01547 E01548 E01549 E01550 E01551 E01552 E01554 E01556 E01557 E01558 E01560 E01561 E01562 E01563 E01566 E01568 E01570 E01571 E01576 E01579 E01581 E01582 E01584 E01585  E01586  E01589  E01590  E01591  E01593  E01594  E01595  E01597  E01598  E01601  E01602  E01605  E01606  E01610  E01611  E01612  E01615  E01617  E01618  E01620  E01624  E01627  E01629  E01630  E01631  E01633  E01634  E01636  E01637  E01638  E01639  E01640  E01641  E01642  E01643  E01646  E01647  E01648  E01649  E01650  E01651  E01653  E01654  E01655  E01656  E01657  E01658  E01659  E01660  E01661  E01662  E01663  E01664  E01666  E01667  E01670  E01671  E01673  E01676  E01680  E01888  E03292  E03295  E03296  E03297  E03299  E03301  E03302  E03303  E03304  E03306  E03307  E03308  E03309  E03311  E03312  E03313  E03314  E03316  E03565  E03566  E04488  E04626  E05033  E05036  E05037  E05038  E05040  E05041  E05042  E05043  E05044  E05045  E05048  E05049  E05050  E05051  e05053  E05055  E05056  E05057  E05058  E05059  E05060  E05062  E05063  E05065  E05067  E05070  E05082  E05083  E05088  E05093  E05099  E05100  E05102  E05141  E05143  E05145  E05146  E05147  E05149  E05150  E05151  E05184  E05185  E05186  E05187  E05189  E05190  E05191  E05192  E05193  E05195  E05202  E05210  E05212  E05213  E05215  E05216  E05217  E05219  E05220  E05221  E05222  E05223  E05224  E05237  E05239  E05241  E05244  E05245  E05246  E05248  E05249  E05254  E05258  E05263  E05264  E05265  E05266  E05267  E05269  E05345  E05346  E05348  E05349  E05350  E05351  E05378  E05380  E05381  E05382  E05383  E05385  E05386  E05388  E05389  E05391  E05392  E05393  E05394  E05396  E05398  E05400  E05402  E05405  E05409  E05411  E05414  E05415  E05418  E05419  E05420  E05422  E05423  E05424  E05425  E05547  E05549  E05550  E05551  E05552  E05553  E05554  E05555  E05950  E05952  E05953-1  E05954-1  E05956-1  E05957-1  E05958  E05959-1  E05960-1  E05961-1  E05962-1  E05963-1  E05964-1  E05965-1  E05966  E05968-1  E05969-1  E05970-1  E05971-1  E05973-1  E05974-1  E05975-1  E05976-1  E05977-1  E05978-1  E05980-1  E05982-1  E06273  E06278  E06281  E06283  E06284  E06285  E06290  E06291  E06292  E06369  E06445-1  E06446  E06598  E06598  E06599  E06600  E06601  E06603  e06606  E06607  E06608  E06609  E06610  E06612  E06613  E06615  E06616  E06617  E06618  E06620  E06622  E06625  E06627  E06630  E06631  E06634  E06636 E06637 E06637-1 E06638-1 E06639-1 E06715 E06718 E06719 E06720 E06721 E06722 E06723 E06724 E06725 E06745 E06747 E06975 E06976 E06977 E07350-1 E07351-1 E07352 E07353-1 E07355-1 E07356-1 E07360 E07361 E07362 E07363 E07365 E07366-1 E07367 E07368-1 E07369 E07370 E07371-1 E07373-1 E07374-1 E07375 E07376 E07377 E07378-1 E07379 E07380 E07381 E07382 E07383-1 E07384-1 E07385-1 E07386 E07387-1 E07388-1 E07389 E07390 E07393-1 E07394 E07395 E07396-1 E07397-1 E07398 E07399 E07400-1 E07401-1 E07402 E07403-1 E07404 E07405-1 E07406-1 E07407 E07408-1 E07409-1 E07410 E07411-1 E07412 E07413-1 E07414-1 E07415-1 E07416 E07417 E07418 E07419 E07495 E07496 E07497 E07498 E07499 E07500 E07501 E07502 E07503 E07506 E07507 E07508 E07509 E07510 E07511 E07513 E07514 E07686 E07687 E07689 E07690 E07691 E07692 E07693 E07694 E07695 E07696 E07697 E07698 E07699 E07700 E07702 E07703 E07704 E07780 E07781 E07782 E07783 E07784 E07860 E07861 E07862 E07863 E07864 E07865 E07866 E07867 E07868 E07869 E08020 E08021 E08022 E08023 E08024 E08025 E08026 E08027 E08028 E08029 E08030 E08031 E08033 E08034 E08035 E08036 E08037 E08038 E08039 E08040 E08041 E08044 E08045 E08046 E08047 E08048 E08049 E08050 E08052 E08053 E08054 E08055 E08056 E08057 E08058 E08059 E08060 E08061 E08063 E08064 E08065 E08066 E08067 E08068 E08069 E08070 E08071 E08072 E08073 E08074 E08075 E08076 E08077 E08078 E08079 E08081 E08082 E08083 E08084 E08086 E08087 E08089 E08090 E08091 E08092 E08093 E08094 E08095 E08096 E08097 E08098 E08099 E08100 E08101 E08102 E08103 E08104 E08105 E08106 E08107 E08108 E08109 E08110 E08111 E08113 E08114 E08115  E08115  E08117  E08118  E08119  E08120  E08121  E08122  E08124  E08131  E08132  E08133  E08134  E08135  E08137  E08139  E08140  E08142  E08143  E08145  E08146  E08147  E08148  E08149  E08150  E08151  E08152  E08153  E08154  E08155  E08156  E08157  E08158  E08160  E08161  E08162  E08163  E08164  E08165  E08755  E08756  e08757  E08758  E08759  E08760  E08761  E08762  E08763  E08764  E08765  E08766  E08767  E08769  e08770  E08771  E08772  E08773  E08775  E08776  E08777  e08778  E08779  E08780  E08781  E08782  E08783  E08859  E08860  E08861  E08864  E08865  E08866  E08868  E08870  E08871  E08872  E08873  E08874  E08878  E09125  E09126  E09127  E09128  E09129  E09130  E09131  E09207  E09208  E09209  E09210  E09211  E09212  E09213  E09214  E09215  E09216  E09217  E09218  E09219  E09220  E09221  E09222  E09225  E09226  E09227  E09228  E09229  E09231 E09232  E09233  E09234  E09235  E09236  E09237  E09238  E09239  E09240  E09241  E09242  E09243  E09244  E09245  E09246  E09248  E09249  E09250  E09251  E09252  E09253  E09254  E09255  E09256  E09334  E09336  E09338  E09339  E09345  E09346  E09347  E09349  E09354  E09355  E09356  E09358  E09359  E09360  E09361  E09362  E09363  E09364  E09365  E09369  E09370  E09371  E09372  E09374  E09375  E09376  E09378  E09454  E09455  E09457  E09460  E09461  E09464  E09466  E09467  E09468  E09470  E09471  E09472  E09473  E09474  E09476  E09477  E09478  E09856  E09857  E09859  E09860  E09861  E09862  E09863  E09945  E09946  E09947  E09948  E09949  E09950  E09951  E09952  E09953  E09954  E09955  E09956  E09957  E09958  E09959  E09960  E09961  E09962  E09963  E09964  E09965  E09966  E09967  E09968  E09969  E10115  E10116  E10117  E10119  E10120  E10121  E10122  E10123  E10124  E10275  E10277  E10278  E10279  E10280  E10281  E10282  E10283  E10284  E10285  E10286  E10287  E10288  E10289  E10290  E10291  E10292  E10293  E10294  E10295  E10296  E10297  E10298  E10299  E10576  E10578  E10579  E10656  E10657  E10659  E10661  E10662  E10666  E10667  E10668  E10670  E10671  E10673  E10674  E10711  E10712  E10713  E10714  E10716  E10717  E10718  E10721  E10722  E10723  E10746  E10750  E10754  E10755 E10756 E10796 E10797 E10799 E10800 E10801 E10803 E10804 E10806 E10808 E10826 E10827 E10829 E10831 E10832 E10833 E10834 E10835 E10899 E10900 E10901 E10902 E10903 E10904 E10905 E10906 E10907 E10909 E10910 E10911 E10912 E10913 E10914 E10915 E10916 E10917 E10918 E10919 E10920 E10921 E10923 E10924 E10925 E10926 E10927 E10928 E10929 E10930 E10933 E10934 E10935 E10936 E10937 E10939 E10940 E10941 E10942 E10943 E10944 E10945 E10946 E10947 E10948 E10951 E10952 E10953 E10954 E10955 E10956 E10957 E10958 E10959 E10960 E10961 E11062 E11063 E11064 E11065 E11066 E11068 E11245 E11246 E11247 E11248 E11249 E11251 E11252 E11253 E11254 E11255 E11256 E11257 E11258 E11259 E11260 E11261 E11262 E11263 E11264 E11265 E11266 E11268 E11269 e11270 E11271 E11273 E11274 E11275 E11278 E11279 E11280 E11281 E11282 E11283 E11284 E11285 E11286 E11287 E11288 E11289 E11290 E11291 E11292 E11293 E11294 E11295 e11296 E11297 E11298 E11299 E11300 E11301 E11302 E11303 E11304 E11305 E11306 E11307 E11308 E11309 E11310 E11311 E11312 E11313 E11314 E11315 E11316 E11317 E11318 E11319 E11320 E11321 E11323 E11326 E11327 E11328 E11329 E11330 E11331 E11332 E11333 E11334 E11336 E11337 E11338 E11339 E11340 E11341 E11342 E11343 E11344 E11345 E11346 E11347 E11348 E11349 E11350 E11351 E11352 E11353 E11354 E11430 E11432 E11433 E11434 E11435 E11437 E11438 E11439 E11440 E11441 E11442 E11443 E11446 E11447 E11448 E11449 E11452 E11459 E11460 E11461 E11515 E11516 E11517 E11518 E11519 E11520 E11521 E11522 E11524 E11525 E11526 E11527 E11528 E11529 E11530 E11531  E11532  E11533  E11534  E11535  E11536  E11537  E11538  E11539  E11540  E11541  E11542  E11543  E11544  E11545  E11546  E11547  E11548  E11549  E11550  E11551  E11552  E11553  E11554  E11555  E11556  E11557  E11558  E11559  E11560  E11561  E11562  E11563  E11564  E11565  E11567  E11568  E11569  E11570  E11571  E11572  E11573  E11574  E11575  E11576  E11577  E11578  E11579  E11580  E11581  E11582  E11583  E11584  E11585  E11586  E11587  E11588  E11589  E11590  E11591  E11592  E11593  E11594  E11595  E11596  E11597  E11598  E11599  E11600  E11601  E11602  E11603  E11604  E11605  E11607  E11608  E11609  E11610  E11613  E11614  E11645  E11646  E11647  E11648  E11649  E11650  E11651  E11652  E11653  E11654  E11655  E11656  E11658  E11659  E11660  E11661  E11662  E11663  E11664  E11665  E11666  E11667  E11668  E12010  E12011  E12012  E12013  E12089  E12092  E12093  E12094  E12095  E12096  E12097  E12098  E12675  E12676  E12677  E12678  E12679  E12680  E12681  E12684  E12685  E12686  E12687  E12688  E12689  E12689  E12690  E12691  E12692  E12693  E12695  E12696  E12698  E12699  E12700  E12703  E12704  E12705  E12706  E12707  E12708  E12709  E12710  E12711  E12712  E12713  E12714  E12715  E12717  E12718  E12720  E12721  E12723  E12725  E12726  E12727  E12728  E12729  E12731  E12732  E12733  E12734  E12735  E12736  E12737  E12738  E12739  E12740  E12741  E12742  E12743  E12744  E12745  E12746  E12747  E12748  E12749  E12750  E12751  E12752  E12753  E12754  E12755  E12756  E12757  E12759  E12760  E12761  E12762  E12763  E12764  E12765  E12766  E12768  E12769  E12770  E12772  E12773  E12774  E12775  E12777  E12778  E12779  E12780  E12781  E12782  E12783  E12784  E12785  E12786  E12787  E12790  E12791  E12792  E12793  E12796  E12797  E12798  E12799  E12800  E12801  E12802  E12923  E12924  E12925  E12926  E12927-1  E12928  E12929  E12930  E12931  E12932  E13150  E13151  E13152  E13153  E13154  E13155  E13156  E13157  E13158  E13159  E13160  E13161  E13162  E13163  E13164  E13165  E13166  E13167  E13168  E13169  E13370  E13371  E13372  E13373  E13379  E13380  E13382  E13383  E13384  E13385  E13386  E13387  E13388  E13389  E13390  E13391 E13492 E13493 E13494 E13495 E13496 E13497 E13498 E13499 E13500 E13501 E13502 E13503 E13504 E13505 E13507 E13508 E13509 E13510 E13511 E13513 E13514 E13515 E13516 E13518 E13519 E13520 E13521 E13522 E13523 E13524 E13525 E13526 E13527 E13528 E13529 E13530 E13531 E13532 E13533 E13534 E13535 E13536 E13538 E13539 E13540 E13541 E13542 E13543 E13545 E13546 E13547 E13548 E13549 E13550 E13551 E13552 E13553 E13555 E13556 E13557 E13558 E13559 E13560 E13561 E13712 E13713 E13714 E13715 E13716 E13717 E13718 E13719 E13720 E13721 E13722 E13723 E13724 E13725 E13726 E13727 E13728 E13729 E13730 E13932 E13933 E13934 E13935 E13936 E13937 E13938 E13939 E13940 E13941 E13942 E13943 E13944 E13945 E14481 E14482 E14483 E14484 E14485 E14486 e14487 E14488 E14489 E14490 e14491 E14492 E14493 E14495 E14496 E14498 E14499 E14500 E14501 e14504 E14505 E14506 E14507 E14508 e14509 E14510 E14511 E14512 E14513 E14514 E14515 E14516 E14517 E14518 E14519 E14520 E14621 E14624 E14626 E14628 E14630 E14631 E14632 E14636 E14637 E14639 E14641 E14642 E14643 E14644 E14645 E14646 E14647 E14648 E14650 E14651 E14652 E14653 E14654 E14655 E14656 E14657 E14658 E14659 E14661 E14662 E14663 E14664 E14665 E14666 E14667 E14668 E14669 E14670 E14671 E14672 E14673 E14674 E14675 E14677 E14678 E14679 E14680 E14682 E14683 E14684 E14686 E14687 E14688 E14690 E14691 E14692 E14693 E14694 E14695 E14697 e14698 E14699 E14700 E14701 E14702 E14703 E14704 E14705 E14706 E14707 E14709 E14710 E14862 E14865 E14866 E14867 E14868 E14870 E14871 E14872 E14873 E14875 E14876  E14877  E14878  E14881  E14882  E14883  E14886  E14887  E14888  E14891  E14892  E14894  E14895  E14896  E14898  E14899  E14900  E14902  E14904  E14905  E14906  E14907  E14908  E14909  E14910  E14911  E14912  E14915  E14916  E14918  E14919  E14920  E14922  E14923  E14924  E14925  E14926  E14927  E14929  E14930  E14931  E14932  E14933  E14935  E14937  E14938  E14939  E14939  E14940  E14941  E14942  E14943  E14944  E14945  E14946  E14947  E14949  E14950  E14953  E14954  E14955  E14956  E14957  E14959  E14960  E14961  E15062  E15066  E15067  E15070  E15075  E15079  E15081  E16472  E16473  E16474  E16475  E16476  E16578  E16581  E16582  E16584  E16586  E16687  E16688  E16689  E16690  E16691  E16692  E16693  E16695  E16697  E16698  E16699  E16700  E16701  E16702  E16703  E16705  E16706  E16707  E16709  E16711  E16792  E16793  E16794  E16795  E16796  E16797  E16798  E16799  E16800  E16801  E16802  E16803  E16804  E16805  E16807  E16808  E16809  E16810  E16811  E16912  E16913  E16914  E16915  E16916  E16917  E16918  E16919  E16920  E16921  E16922  E16923  E16924  E16925  E16926  E16927  E16928  E16929  E16930  E16931  E16932  E16933  E16934  E16935  E16936  E16937  E16938  E16939  E16940  E16941  E16942  E16943  E16944  E16945  E16946  E16947  E16948  E16949  E16950  E16951  E16952  E16953  E16954  E16955  E16956  E16957  E16958  E16959  E16960  E16961  E16962  E16963  E16964  E16965  E16966  E16967  E16968  E16969  E16970  E16971  E16972  E16973  E16974  E16975  E16976  E16977  E16978  E16979  E16980  E16981  E16982  E16983  E16984  E16985  E16986  E16987  E16988  E16989  E16990  E16991  E16992  E16993  E16994  E16995  E16996  E16997  E16998  E16999  E17000  E17001  E17002  E17003  E17005  E17006  E17007  E17010  E17011  E17014  E17020  E17029  E17034  E17036  E17046  E17047  E17055  E17060  E17061  E17062  E17113  E17114  E17115  E17116  E17117  E17118  E17119  E17129  E17130  E17131  E17132  E17383  E17384  E17385  E17386  E17387  E17388  E17389  E17390  E17391  E17393  E17394  E17395  E17396  E17397  E17398  E17399  E17400  E17401  E17402  E17403  E17404  E17405  E17406  E17407  E17408  E17409  E17410  E17411  E17413  E17414 E17415 E17416 E17417 E17418 E17419 E17420 E17421 E17422 E17423 E17424 E17425 E17426 E17427 E17428 E17429 E17430 E17431 E17432 E17693 E17694 E17695 E17696 E17697 E17698 E17699 E17700 E17701 E17702 E17703 E17704 E17705 E17706 E17707 E17708 E17710 E17711 E17712 E17763 E17764 E17766 E17767 E17768 E17769 E17770 E17771 E17772 E17773 E17774 E17775 E17776 E17777 E17778 E17779 E17780 E17781 E17782 E17783 E17784 E17785 E17786 E17787 E17789 E17790 E17791 E17792 E17793 E17794 E17795 E17797 E17798 E17799 E17800 E17801 E17802 E17803 E17804 E17805 E17807 E17808 E17809 E17810 E17811 E17812 E18013 E18016 E18020 E18021 E18022 E18023 E18024 E18025 E18026 E18028 E18029 E18030 E18032 E19367 E19368 E19369 E19370 E19371 E19372 E19373 E19374 E19375 E19376 E19377 E19378 E19379 E19380 E19381 E19382 E19383 E19384 E19385 E19386 E19437 E19438 E19439 E19440 E19441 E19442 E19443 E19444 E19445 E19446 E19447 E19448 E19449 E19450 E19451 E19452 E19453 E19454 E19455 E19456 E19758 E19759 E19760 E19761 E19762 E19763 E19764 E19765 E19767 E19768 E19769 E19770 E19771 E19772 E19773 E19774 E19775 E19776 E19777 E19778 E19779 E19780 E19781 E19782 E19800 E19801 E19802 E19803 E19804 E19805 E19807 E19808 E19809 E19823 E19824 E20319 E20320 E20321 E20324 E20325 E20329 E20330 E20331 E20357 E20359 E20360 E20361 E20362 E20363 E20364 E20365 E20366 E20367 E20368 E20369 E20370 E20371 E20372 E20373 E20374 E20375 E20376 E20377 E20378 E20379 E20381 E20745 E20747 E20749 E20752 E20781 E20785 E20788 EO4491 EO5095 EO5384 EO5408 E10710 E10719 E10748  E13374  E13375  E13376  E13377  E14676  E14874
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA Nationwide and countries of: Australia, Belgium, China, Guatemala CentroAmeria, France, Hong Kong, India,Italy, Korea, Lebanon, Malaysia, Morocco, Pakistan, Paraguay, The Netherlands, Romania, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom and Venezula.
  • Product Description
    VenaFlow Elite 30B-B and 30BI-B || Product Usage: The VenaFlow Elite System is a compression pump used for the prevention of Deep Vein Thrombosis.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    DJO, LLC, 1430 Decision St, Vista CA 92081-8553
  • Manufacturer Parent Company (2017)
  • Source
    USFDA