Recall of Device Recall AdvantageSim MD

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE Healthcare, LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    57083
  • Event Risk Class
    Class 2
  • Event Number
    Z-1473-2011
  • Event Initiated Date
    2010-11-24
  • Event Date Posted
    2011-03-02
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-08-03
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel data.
  • Extra notes in the data
    System, simulation, radiation therapy - Product Code KPG
  • Reason
    Ge healthcare has become aware that inconsistent drr (digitally reconstructed radiographic) visualization may occur when the blended drr feature or the isocenter move mode feature is used in combination with specified other functions of the advantagesim application on advantage workstation. these inconsistencies may affect patient safety. other advantagesim functions are not affected.
  • Action
    GE Healthcare sent an "Urgent Medical Device Correction" letter dated November 23, 2010. The letter was addressed to Hospital Administrators / Risk mangers, Oncology Department Managers / Radiation Oncologists, & Radiation Physicists / Dosimetrists / Therapists. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. For any questions or concerns regarding this recall, please call one of the following phone numbers: United States: 800 437 1171 Japan: 0120 - 055 - 919 For other countries, please contact your local GE Healthcare Service Representative.

Device

  • Model / Serial
    Serial Number 00000132364GE1 00000009851GE7 00000184342GE4 00000186390GE1 00000184339GE0 00000186389GE3 00000184341GE6 00000184340GE8 00000184338GE2 00000063115GE0 00000067040GE6 00000069353GE1 00000098315VP5 00000044657GE5 00000099698VP3 00000028881GE1 00000030810HP6 00000028878GE7 00000025475GE5 00000009900GE2 00000009850GE9 00000002678GE1 00000068373GE0 00000112962GE6 00000006609GE2 00000006605GE0 00000041805GE3 00000041804GE6 00000046030GE3 00000008134GE9 00000140656GE0 00000152559GE1 00000152558GE3 00000123784GE1 00000030995HP5 00000035199GE9 00000112965GE9 00000098362VP7 00000063886GE6 00000098721VP4 00000018769GE0 00000109868GE0 00000124937GE4 00000030846HP0 00000025485GE4 00000018779GE9 00000152560GE9 00000041800GE4 00000113881GE7 00000018781GE5 00000056503GE6 00000061414DP9 00000010204GE6 00000061413DP1 00000018673AW4 00000085514GE8 00000127013GE1 00000030793HP4 00000030807HP2 00000001442GE3 00000085513GE0 00000076520GE6 00000142885GE3 263 00000028877GE9 00000108671GE9 00000101181VP6 00000106448GE4 00000106867GE5 00000118326GE8 00000192700GE3 00000001511GE5 00000082780GE8 00000001446GE4 00000005679GE6 00000005492GE4 00000018671AW8 00000099859VP1 00000138886GE7 00000080674GE5 00000017037VE3 00000052577GE4 00000090481GE3 00000006416GE2 00000157089GE4 00000007463GE3 00000006418GE8 00000007465GE8 00000090488GE8 00000022887VE4 00000098603GE4 00000010197GE2 00000051673GE2 00000003616GE0 00000165784GE0 00000005563GE2 00000025484GE7 00000116935GE8 00000068390GE4 00000068391GE2 00000000ASM009 00000159393GE8 00000098598GE6 00000076531GE3 00000001451GE4 00000165782GE4 00000152556GE7 00000008137GE2 00000100000VP9 00000098596GE0 00000098600GE0 00000194611GE0 00000127667GE4 00000152562GE5 00000025472GE2 00000001448GE0 00000114870GE9 00000114871GE7 00000124938GE2 00000138887GE5 00000162306GE5 00000037513GE9 00000018760AW9 00000030922HP9 00000098609GE1 00000165783GE2 00000056504GE4 00000174250GE1 00000064212GE4 00000106872GE5 00000019273VE2 00000051667GE4 00000073484GE8 00000112968GE3 00000134785GE5 00000068374GE8 00000102304VP3 00000028895GE1 00000127665GE8 00000006611GE8 00000069358GE0 00000099606VP6 00000030906HP2 00000098604GE2 00000025481GE3 00000102315VP9 00000100838GE2 00000174248GE5 00000111715GE9 00000114868GE3 00000102229VP2 00000030982HP3 00000100867VP1 00000001449GE8 00000025486GE2 00000018783GE1 00000100836GE6 00000073486GE3 00000073485GE5 00000021235GE7 00000144923GE0 00000073490GE5 00000001444GE9 00000106870GE9 00000028884GE5 00000121719GE9 00000028886GE0 00000028876GE1 00000018778GE1 00000145027GE9 00000106450GE0 00000031035HP9 00000106451GE8 00000138881GE8 00000028894GE4 00000098296VP7 00000182511GE6 00000068376GE3 00000098592GE9 00000069361GE4 00000009893GE9 00000061425GE5 00000145023GE8 00000090486GE2 00000163709GE9 00000079708GE4 00000069348GE1 00000174247GE7 00000073491GE3 00000073492GE1 00000001520GE6 00000019279VE9 00000190302GE0 00000010190GE7 00000186396GE8 00000041013GE4 00000082787GE3 00000096084GE9 00000080675GE2 00000109865GE6 00000186391GE9 00000144925GE5 00000100841GE6 00000121721GE5 00000056732GE1 00000081295GE8 00000068381GE3 00000046034GE5 00000177829GE9 00000030875HP9 00000111105GE3 00000034438GE2 00000114866GE7 00000001516GE4 00000006088GE9 00000111716GE7 00000102863VP8 00000140657GE8 00000051587GE4 00000051674GE0 00000007467GE4 00000018782GE3 00000001184GE1 00000081290GE9 00000069351GE5 00000001452GE2 00000108104VE1 00000001455GE5 00000046033GE7 00000118328GE4 00000018766GE6 00000114867GE5 00000006414GE7 00000006962HP5 00000154701GE7 00000108121VE5 00000108120VE7 00000063123GE4 00000124940GE8 00000100837GE4 00000138889GE1 00000144926GE3 00000169274GE8 00000106875GE8 00000118332GE6 00000085515GE5 00000061428GE9 00000034440GE8 00000048215GE8 00000014291GE9 00000157090GE2 00000031062HP3 00000102652VP5 00000073488GE9 00000108119VE9 00000197193GE6 00000030918HP7 00000174246GE9 00000014287GE7 00000046045GE1 00000061436GE2 00000028891GE0 00000046041GE0 00000098597GE8 00000046040GE2 00000174249GE3 00000001447GE2 00000079709GE2 00000006089GE7 00000056733GE9 00000144933GE9 00000025480GE5 00000031063HP1 00000037511GE3 00000001527GE1 00000010208GE7 00000041802GE0 00000111713GE4 00000160237GE4 00000098602GE6 00000197191GE0 00000197192GE8 00000186398GE4 00000068380GE5 00000190301GE2 00000138884GE2 00000160233GE3 00000068388GE8 00000082778GE2 00000051590GE8 00000082779GE0 00000051581GE7 00000109863GE1 00000109862GE3 00000108674GE3 00000109864GE9 00000046046GE9 00000109869GE8 00000056499GE7 00000157088GE6 00000144924GE8 00000009901GE0 00000353688BU5 00000010209GE5 00000154702GE5 00000140662GE8 00000081298GE2 00000081297GE4 00000001443GE1 00000068389GE6 00000108672GE7 00000121722GE3 00000106868GE3 00000082781GE6 00000082786GE5 00000145030GE3 00000190298GE0 00000063117GE6 00000076523GE0 00000035200GE5 00000051672GE4 00000030904HP7 00000006966HP6 00000001525GE5 00000192701GE1 00000018664AW3 00000069363GE0 00000090494GE6 00000052585GE7 00000080676GE0 00000001513GE1 00000030945HP0 00000138882GE6 00000154703GE3 00000108665GE1 00000113883GE3 00000111106GE1 00000121725GE6 00000037514GE7 00000124935GE8 00000014289GE3 00000061439GE6 00000194612GE8 00000192705GE2 00000194614GE4 00000098430VP2 00000140655GE2 00000090484GE7 00000079706GE8 00000069350GE7 00000030859HP3 00000090480GE5 00000123791GE6 00000108116VE5 00000157092GE8 00000132367GE4 00000018776GE5 00000061424GE8 00000061396DP8 00000006960HP9 00000123790GE8 00000096087GE2 00000194613GE6 00000080276GE9 00000100843GE2 00000090498GE7 00000112967GE5 00000041799GE8 00000145033GE7 00000152565GE8 00000067041GE4 00000145031GE1 00000114862GE6 00000134788GE9 00000079710GE0 00000080277GE7 00000162301GE6 00000017087GE8 00000153088GE0 00000030933HP6 00000123786GE6 00000113882GE5 00000106452GE6 00000030838HP7 00000159390GE4 00000159391GE2 00000102707VP7 00000152563GE3 00000030771HP0 00000106449GE2 00000001519GE8 00000113885GE8 00000188323GE0 00000121718GE1 00000056501GE0 00000052630GE1 00000127668GE2 00000031043HP3 00000061385DP1 00000030787HP6 00000112960GE0 00000030864HP3 00000076524GE8 00000169272GE2 00000008842GE7 00000186395GE0 00000041358VE3 00000108110VE8 00000046032GE9 00000021229GE0 00000160234GE1 00000021234GE0 00000062503GE8 00000080479GE9 00000100842GE4 00000132365GE8 00000052580GE8 00000081293GE3 00000186394GE3 00000121726GE4 00000081291GE7 00000100846GE5 00000056728GE9 00000030831HP2 00000030779HP3 00000030818HP9 00000046039GE4 00000064017GE7 00000005687GE9 00000010205GE3 00000041015GE9 00000006612GE6 00000006610GE0 00000051679GE9 00000051588GE2 00000025473GE0 00000085875VE3 00000052583GE2 00000017091GE0 00000082784GE0 00000030856HP9 00000026292HP3 00000182512GE4 00000121724GE9 00000034444GE0 00000056724GE8 00000010203GE8 00000030855HP1 00000014294GE3 00000031028HP4 00000134787GE1 00000090482GE1 00000100845GE7 00000160239GE0 00000162300GE8 00000127012GE3 00000153086GE4 00000081294GE1 00000127009GE9 00000102234VP2 00000017044VE9 00000144927GE1 00000145034GE5 00000073487GE1 00000030978HP1 00000076530GE5 00000090490GE4 00000006614GE2 00000030828HP8 00000018771GE6 00000001450GE6 00000095305VP9 00000111718GE3 00000035201GE3 00000006094GE7 00000066231GE2 00000096085GE6 00000051676GE5 00000096086GE4 00000006091GE3 00000134784GE8 00000069352GE3 00000124939GE0 00000108669GE3 00000076525GE5 00000123785GE8 00000034442GE4 00000069356GE4 00000069364GE8 00000025482GE1 00000118329GE2 00000112966GE7 00000025470GE6 00000041801GE2 00000098594GE5 00000061427GE1 00000030803HP1 00000030861HP9 00000076522GE2 00000159106GE4 00000090492GE0 00000031038HP3 00000159392GE0 00000018665AW0 00000046044GE4 00000034445GE7 00000051671GE6 00000019284VE9 00000061391DP9 00000152561GE7 00000186392GE7 00000096093GE0 00000162305GE7 00000098601GE8 00000019274VE0 00000063726GE4 00000101421VP6 00000100849GE9 00000031034HP2 00000018780GE7 00000018770GE8 00000124941GE6 00000100026VP4 00000169275GE5 00000169270GE6 00000061429GE7 00000162303GE2 00000030884HP1 00000145029GE5 00000165778GE2 00000030896HP5 00000009857GE4 00000157095GE1 00000123789GE0 00000030868HP4 00000144932GE1 00000069349GE9 00000111712GE6 00000082782GE4 00000068372GE2 00000046048GE5 00000009044GE9 00000001434GE0 00000076518GE0 00000106871GE7 00000098297VP5 00000061386DP9 00000056725GE5 00000018777GE3 00000030902HP1 00000073483GE0 00000081289GE1 00000090496GE1 00000157093GE6 00000109866GE4 00000121720GE7 00000096088GE0 00000118327GE6 00000173431GE8 00000138888GE3 00000134783GE0 00000005681GE2 00000052579GE0 00000069347GE3 00000112963GE4 00000037510GE5 00000192704GE5 00000153089GE8 00000140659GE4 00000138880GE0 00000142878GE8 00000144931GE3 00000165780GE8 00000182513GE2 00000144922GE2 00000142879GE6 00000030936HP9 00000085512GE2 00000102720VP0 00000154704GE1 00000108668GE5 00000034436GE6 00000153091GE4 00000109867GE2 00000061380DP2 00000114863GE4 00000109919GE1 00000052631GE9 00000051584GE1 00000051583GE3 00000090495GE3 00000025477GE1 00000100854VP9 00000046049GE3 00000061315GE8 00000100844GE0 00000001521GE4 00000051666GE6 00000051665GE8 00000019290VE6 00000138883GE4 00000069362GE2 00000069359GE8 00000134789GE7 00000132360GE9 00000163712GE3 00000030784HP3 00000134782GE2 00000134781GE4 00000080481GE5 00000080672GE9 00000076519GE8 00000134790GE5 00000090489GE6 00000090499GE5 00000061373DP7 00000184337GE4 00000028880GE3 00000073489GE7 00000124934GE1 00000124943GE2 00000124942GE4 00000080150GE6 00000113886GE6 00000041016GE7 00000051586GE6 00000142884GE6 00000001431GE6 00000073497GE0 00000001454GE8 00000068379GE7 00000127664GE1 00000108666GE9 00000051589GE0 00000111103GE8 00000142882GE0 00000142883GE8 00000106874GE1 00000028879GE5 00000102053VP6 00000127663GE3 00000005680GE4 00000140661GE0 00000111714GE2 00000025474GE8 00000009849GE1 00000169271GE4 00000160235GE8 00000145025GE3 00000098593GE7 00000020862HP7 00000127010GE7 00000006411GE3 00000144936GE2 00000006608GE4 00000073494GE7 00000008136GE4 00000102850VP5 00000118331GE8 00000010201GE2 00000080275GE1 00000010191GE5 00000010195GE6 00000142881GE2 00000001523GE0 00000009902GE8 00000100942VP2 00000051585GE8 00000112961GE8 00000132366GE6 00000030842HP9 00000030860HP1 00000098595GE2 00000030857HP7 00000001514GE9 00000018761AW7 00000030798HP3 00000006867HP6 00000157097GE7 00000144934GE7 00000076527GE1 00000103056VP8 00000028889GE4 00000123788GE2 00000063122GE6 00000041806GE1 00000160232GE5 00000010193GE1 00000085847VE2 00000029510AW5 00000056730GE5 00000202073GE3 00000009894GE7 00000010200GE4 00000177830GE7 00000101250VP9 00000190299GE8 00000190300GE4 00000001526GE3 00000006410GE5 00000009899GE6 00000197194GE4 00000052581GE6 00000099604VP1 00000173429GE2 00000202072GE5 237 238 241 00000006863HP5 00000013369VE4 00000016991VE2 00000016999VE5 00000017032VE4 00000017034VE0 00000017036VE5 00000017040VE7 00000017041VE5 00000017046VE4 00000017049VE8 00000019291VE4 00000019292VE2 00000019295VE5 00000022208VE3 00000022879VE1 00000098278VP5 00000098279VP3 00000098298VP3 00000098587VP9 00000098823VP8 00000099834VP4 00000099863VP3 00000099980VP5 00000100062VP9 00000100070VP2 00000100936VP4 00000101146VP9 00000101909VP0 00000102281VP3 00000102623VP6 00000102874VP5 00000102999VP0 00000103017VP0 00000160236GE6 00000160238GE2 00000165785GE7 00000174245GE1 00000190303GE8 00000197195GE1 00000001445GE6 00000001512GE3 00000005688GE7 00000006087GE1 00000006090GE5 00000007464GE1 00000007471GE6 00000008139GE8 00000009046GE4 00000014290GE1 00000017088GE6 00000018772GE4 00000028883GE7 00000028885GE2 00000056727GE1 00000063118GE4 00000108102VE5 00000162307GE3 00000163708GE1 00000009854GE1 00000202074GE1 00000202075GE8 00000202076GE6 00000169273GE0 00000085845VE6 00000030955HP9 00000K10111006 00000031559AW8 00000031561AW4 00000031557AW2 00000030919HP5 00000031600AW0 000000M9062001 00000031599AW4 00000031598AW6 00000031601AW8 00000031594AW5 00000031605AW9 00000031603AW4 00000031604AW2 00000031602AW6 00000018674AW2 00000018670AW0 00000108087VE8 00000054159AW9 000000D8015004 000000D8015003 000000W7360002 00000031573AW9 00000078498VE3 00000078495VE9 00000031574AW7 000000D8015002 000000Q7351006 000000L7235003 000000U8203003 000000Q9348001 00000029556AW8 000000X9356003 000000X9356002 00000C10014013 00000C10014011 000000X9356001 00000030929HP4 00000051843VE1 00000030937HP7 00000030959HP1 000000E7298003 000000N7271001 000000E7298001 000000N7271002 000000E7298002 00000054241AW5 00000053664AW9 00000L10008001 00000054151AW6 00000053660AW7 000000127S0004 00000054235AW7 00000031596AW0 00000031575AW4 00000031568AW9 00000031567AW1 00000054150AW8 00000053666AW4 00000108099VE3 000000T9173001 00000029534AW5 00000AG9247001 00000C10014018 00000S10064003 000000Q9348004 00000017042VE3 00000029516AW2 00000029517AW0 00000029528AW7 00000108078VE7 00000108092VE8 00000029537AW8 000000U9243003 00000C10014001 00000S10064006 00000078492VE6 00000M10181001 0000AB10246001 00000Z10249002 00000Z10249003 00000030984HP9 00000054166AW4 000000S9146001 000000AB920201 000000S8207001 000000V8109002 00000100847GE3 00000029506AW3 00000029542AW8 00000447516BU6 00000029551AW9 00000085822VE5 00000085824VE1 00000029550AW1 00000029554AW3 000000J7254003 00000054242AW3 00000054240AW7 00000054161AW5 00000108083VE7 00000108079VE5 00000108100VE9 00000447518BU2 00000108073VE8 00000108082VE9 00000085809VE2 00000022882VE5 323 000000L7269001 000000B8324003 00000085832VE4 00000078505VE5 00000108107VE4 000000D7295001 00000108075VE3 00000029535AW2 00000108076VE1 00000051825VE8 00000031587AW9 000000Z9253001 000000Z9253003 000000Z9253002 00000031558AW0 00000108108VE2 000000Z9321001 00000056707VE3 00000031002HP9 00000031040HP9 00000022888VE2 00000019294VE8 00000051834VE0 000000J8114001 00000085855VE5 00000051838VE1 00000051826VE6 00000031016HP9 00000051836VE5 00000006913HP8 00000006833HP8 00000006911HP2 00000054186AW2 00000C10014008 00000C10014023 00000Q10189001 00000085876VE1 00000059763VE3 00000030964HP1 000000B7310001 00000031018HP5 00000030841HP1 00000054193AW8 000000W7360001 00000108106VE6 000000A7178003 00000447514BU1 00000053671AW4 00000053654AW0 000000L7269003 00000M10204002 00000085806VE8 000000L7235006 00000AE9329005 00000AE9329006 00000AG9343001 0000AB10246002 00000085825VE8 00000085853VE0 00000018679AW1 00000018678AW3 000000AB920202 000000D9167001 000000D9167004 000000Z9253004 00000108109VE0 00000029531AW1 00000085801VE9 00000085854VE8 00000085865VE4 000000R9155001 00000N10221001 00000Z10249001 00000C10014024 00000085848VE0 00000030920HP3 00000030925HP2 000000K8211001 00000030924HP5 000000K8220001 00000030907HP0 00000N10202001 000000U9243002 000000Q8148002 00000085846VE4 00000085842VE3 000000R9313001 000000U8203004 00000031588AW7 000000P8059003 00000085819VE1 00000078501VE4 00000085805VE0 000000J7254001 00000019298VE9 000000S8343001 00000053672AW2 00000085834VE0 000000P9071001 000000T8038001 000000R9168003 000000V8109004 000000Q8140006 000000X8063002 00000054198AW7 00000054192AW0 00000054237AW3 00000054215AW9 00000054170AW6 00000029543AW6 00000029513AW9 00000G10154003 00000C10014010 000000K8295006 000000V8023001 000000G7346001 000000G7346002 000000J7254002 00000054185AW4 00000054184AW7 00000054244AW9 00000054155AW7 00000029547AW7 00000054249AW8 00000054173AW0 00000K10111001 00000AE9329009 000000Z9253008 00000041012GE6 00000106876GE6 00000098599GE4 00000025468GE0 00000192703GE7 00000007470GE8 00000101352VP3 00000053657AW3 00000054153AW2 000000S7318001 00000AE9329010 00000AC9357001 000000C7253001 000000X8063003 355 00000085849VE8 000000C7253002 000000N7313001 000000Y7169004 00000054174AW8 00000054176AW3 00000054175AW5 00000054177AW1 00000054239AW9 00000054234AW0 00000108088VE6 00000031578AW8 00000031565AW5 000000N8281002 000000T8252001 000000P8248001 000000N8281001 000000U8060001 000000S8066001 000000X8283002 000000X8283001 00000018675AW9 00000140658GE6 000000R9183001 000000N8301001 000000H8298001 000000U8332001 000000Q7351004 00000054169AW8 00000108095VE1 000000C8242001 000000X9350001 000000Z9253006 00000AE9329007 000000M9338001 000000N9051002 000000B8324002 000000Q8340001 00000C10014004 00000C10014003 00000C10014006 00000C10014012 00000C10014017 000000D9107001 00000AE9329001 00000AB9231001 000000Q9197001 000000D9167003 000000U8203001 000000Q9184001 000000M9061001 000000U8332004 000000N9051001 000000K8021002 000000K9030001 000000N8350002 000000Q8340002 000000P8354001 000000K7316001 000000N8352001 000000K8213001 000000G8256002 000000V8109001 000000L7235002 000000K8021001 000000X8063001 000000K8036001 000000D7222002 000000K8021004 000000N8028001 000000N7311002 000000N7311003 000000L7269004 00000054233AW2 00000053670AW6 000000H7174001 000000Y7169001 00000054191AW2 00000054171AW4 00000054231AW6 00000054232AW4 00000053662AW3 00000054245AW6 00000031597AW8 00000054221AW7 00000031579AW6 00000054160AW7 00000031608AW3 00000031589AW5 00000031556AW4 00000031576AW2 00000031585AW3 00000078509VE7 000000K8021003 00000030871HP8 222 213 00000031580AW4 00000031560AW6 00000031572AW1 00000C10014007 00000AC9357002 00000G10154001 00000H10166001 00000H10174002 00000M10153004 000000E9177001 000000R9336002 00000018683AW3 00000029519AW6 0000AB10246003 00000W10013001 000000J9056001 00000029530AW3 00000054189AW6 000000AC833301 00000018663AW5 00000085872VE0 00000054220AW9 00000054247AW2 00000031571AW3 000000Q8148001 00000108098VE5 00000C10014009 00000C10014002 000000ZA912601 000000S9146003 00000006931HP0 00000447513BU3 00000018682AW5 00000G10154002 00000029525AW3 00000085864VE7 00000C10014014 00000085813VE4 00000054164AW9 00000C10014005 00000M10168001 00000H10169001 00000S10188001 00000029540AW2 000000V9153002 000000X9163001 000000V9153001 000000R9149001 000000S9146002 000000X9163002 00000054183AW9 00000108091VE0 000000A9132002 000000A9132003 00000H10174001 00000054168AW0 000000D7222001 000000Q8140005 000000T8266001 000000L7235001 000000Q8140002 000000A8049001 000000H9098001 00000144928GE9 000000ZB808501 00000029538AW6 000000S8259001 000000B8348001 000000D9167002 000000U8332003 000000J8135002 000000L7235005 00000029548AW5 000000N9240001 000000Q9348006 00000006874HP2 00000085829VE0 00000108080VE3 00000054200AW1 00000029555AW0 00000018689AW0 00000054204AW3 000000F8227003 000000Q9348005 000000H9098002 00000M10204001 00000085871VE2 00000085837VE3 00000C10014019 00000C10014020 00000M10153001 00000S10064007 00000078506VE3 00000S10167002 00000085844VE9 000000X9250001 00000085802VE7 000000V9251002 000000Q9348008 00000K10111008 00000K10111003 00000M10153003 00000029549AW3 00000085823VE3 00000085858VE9 00000085862VE1 00000085870VE4 00000108084VE5 00000108085VE2 00000108097VE7 00000447515BU8 000000V9251001 000000Q8140004 000000F9065001 00000018687AW4 00000108089VE4 00000030854HP4 00000J10232001 00000101117VP0 00000054158AW1 000000P9047001 000000B8348003 00000006801HP5 00000085866VE2 00000030927HP8 00000054190AW4 000000R9149002 000000F8227001 000000U8203002 000000N8311001 000000C8242002 00000085860VE5 00000085859VE7 000000Q9348007 00000M10153002 00000085836VE5 00000029532AW9 00000085812VE6 00000078508VE9 00000053652AW4 000000L9057002 00000018708AW8 00000AE9329008 00000085878VE7 00000056718VE0 00000085873VE8 00000051822VE5 00000030991HP4 000000A9132001 00000085838VE1 00000029521AW2 00000029514AW7 00000C10014015 00000S10064004 0000ZA10020001 000000W7360004 000000C7324001 000000W7360003 00000L10008002 00000056716VE4 000000B8348002 000000Q7351005 000000M9078001 00000018681AW7 00000056725VE5 000000R9022001 00000031569AW7 000000L9352001 000000Q8140001 00000053650AW8 00000030988HP0 00000031005HP2 000000K7316002 00000029536AW0 000000N8350001 00000029557AW6 00000031015HP1 000000Q9348003 00000054154AW0 000000J8114003 00000C10014025 000000V8109003 000000J8114002 000000J7254004 000000V8023002 000000C8234001 000000L7235004 000000S9048001 00000018677AW5 000000Y7169003 00000054188AW8 00000026279HP0 000000T8309001 000000N7271004 00000054163AW1 00000053658AW1 00000085833VE2 000000U8060002 00000031583AW8 00000054218AW3 00000053659AW9 000000ZB733801 000000U9243001 000000H8235001 000000R9168004 00000Y10190003 00000Y10190004 00000085828VE2 00000108090VE2 00000085850VE6 00000C10014016 00000B10238002 00000S10064001 00000S10064008 000000R9336001 267 00000019270VE8 00000029507AW1 00000085808VE4 00000085817VE5 00000085863VE9 00000AE9329003 00000AE9329004 00000Y10190001 00000Y10190002 00000029539AW4 00000029515AW4 00000F10217001 00000F10217002 00000N10202004 000000N9189001 00000085811VE8 00000085804VE3 00000S10064005 000000P8059004 00000U10209001 0000AB10246004 00000B10238001 00000F10256001 00000N10202005 00000C10014022 000000B7310002 00000030944HP3 00000112959GE2 000000Y7169006 000000Y7169002 00000111719GE1 00000197196GE9 00000098735VP4 000000L7269002 000000B8324001 00000054156AW5 00000053669AW8 000000R9168002 000000R9187001 000000Z9253005 00000030895HP7 000000K8214001 00000108105VE8 00000K10111005 00000078497VE5 00000R10183001 00000078500VE6 00000D10230001 00000054248AW0 00000029545AW1 00000085840VE7 000000H9041001 00000029520AW4 00000054211AW8 00000054210AW0 00000056728VE9 000000W9239001 00000029526AW1 00000085867VE0 00000018762AW5 00000447512BU5 00000085857VE1 00000085839VE9 000000J9014001 00000054243AW1 00000054165AW6 00000108117VE3 000000N7311001 00000447517BU4 00000K10111009 00000K10111010 00000K10111002 00000054194AW6 00000054222AW5 00000054227AW4 00000085826VE6 00000108074VE6 00000F10176001 00000K10111004 000000K9138001 00000054226AW6 000000J7254005 000000Q9348002 000000U8332002 000000K9034001 00000054195AW3 00000085814VE2 00000054228AW2 00000085807VE6 00000AE9329002 000000G8256001 000000K8295005 00000085816VE7 00000C10014021 000000P9316001 000000Q9076001 00000054180AW5 00000054201AW9 00000054236AW5 00000054172AW2 00000108081VE1 00000056715VE6 00000085830VE8 00000059759VE1 00000078490VE0 00000085820VE9 00000097979VP9 00000099816VP1 00000054238AW1 000000K8295004 00000108086VE0 000000H9005001 00000059762VE5 00000097192VP9 00000108093VE6 000000A7178002 00000078494VE2 000000J8135001 00000054187AW0 00000054214AW2 00000085843VE1 00000059756VE7 00000056721VE4 00000022893VE2 000000X9350002 000000T8309002 00000108077VE9 00000K10111007 000000L9352002 00000108103VE3 00000029546AW9 00000127011GE5 00000008142GE2 242 00000006412GE1 00000157094GE4 00000030989HP8 00000019296VE3 00000121723GE1 00000009852GE5 00000182510GE8 00000140664GE4 00000140663GE6 00000159394GE6 00000152557GE5 00000096090GE6 00000096091GE4 00000098608GE3 00000098607GE5 00000098606GE7 00000030822HP1 00000030850HP2 00000144929GE7 00000144930GE5 00000056708VE1 00000019297VE1 00000162302GE4 00000145022GE0 00000145032GE9 215 00000029508AW9 00000025476GE3 00000018774GE0 00000192702GE9 00000177831GE5 00000177832GE3 00000100835GE8 00000080148GE0 00000080482GE3 00000063728GE0 00000090493GE8 00000046043GE6 504 318 320 00000031037HP5 00000163713GE1 00000080671GE1 00000080487GE2 00000080484GE9 00000056502GE8 00000056726GE3 00000177833GE1 00000S10064002 00000013368VE6 00000031581AW2 000000Z9253007 00000053651AW6 00000108113VE2 000000N9051003 00000078493VE4
  • Product Classification
  • Distribution
    Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MI, MA, MI, MN, MS,MO, NT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY and the countries of VENEZUELA, URUGUAY, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SINGAPORE, SAUDI ARABIA, RUSSIAN FEDERATION, REPUBLIC OF KOREA, PORTUGAL, POLAND, PAKISTAN, NOWAY, NIGERIA, NETHERLANDS, MEXICO, MALAYSIA, LITHUANIA, LATVIA, JAPAN, ITALY, ISRAEL, IRELAND, INDIA, HUNGARY, HONG KONG, GREECE, GERMANY, FRANCE, FINLAND, EGYPT, ECUADOR, DOMINICA REPUBLIC, DENMARK, CZECH REPUBLIC, CHINA, CHILE, CANADA, BRAZIL BELGIUM, AUSTRIA, AUSTRALIA.
  • Product Description
    GE ADVANTAGE SIM MD versions 5.x, 6.x, 7.x up to and including version 7.7.0 || Advantage Sim MD is used to prepare geometric and anatomical data relating to a proposed external beam radiotherapy treatment prior to dosimetry planning. Anatomical volumes can be defined automatically or manually in three dimensions using a set of CT images acquired with the patient in the proposed treatment position. Definition of the anatomical volumes may be assisted by additional CT, MR, PET or SPECT studies that have been co-registered with the planning CT scan. Additionally, CT & PET data from a respiratory tracked examination may be used to allow the user define the target or treatment volume over a defined range of the respiratory cycle. The geometric parameters of a proposed treatment field are selected to allow non dosimetric, interactive optimization of field coverage. Defined anatomical structures and geometric treatments fields are displayed on transverse images, on reformatted sagittal, coronal or oblique images, on 3 D views created from the images, or on a beam eye's view display with or without the display of defined structures with or without the display of digitally reconstructed radiograph
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GE Healthcare, LLC, 3000 N Grandview Blvd, Waukesha WI 53188-1615
  • Manufacturer Parent Company (2017)
  • Source
    USFDA