Recall of Device Recall AdTech Electrode Connection System/Cables

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ad-Tech Medical Instrument Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    73685
  • Event Risk Class
    Class 2
  • Event Number
    Z-1503-2016
  • Event Initiated Date
    2016-03-29
  • Event Date Posted
    2016-04-15
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2016-08-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Electrode, depth - Product Code IKD
  • Reason
    Ad-tech is recalling unopened/unused sterile cables intended to be used only with all compatible ad-tech electrodes for the purpose of connecting these electrodes to third party monitors/stimulators. this recall has been initiated due to package failures (i.E. compromised sterile barriers) that were identified during the execution of a shipping simulation investigation.
  • Action
    Consignees were sent a "Medical Device Recall" letter dated March 29, 2016. The letter described the problem and the product being recalled. Advised consignees to immediately examine their inventory, quarantine and return the product to AD-Tech. If the product has been further distributed identify customers and notify them of the recall. Requested consignees to complete and return the "Medical Device Recall Return Response" form to their Ad-Tech Clinical Specialist at customersupport@adtechmedical.com; FAX: 262-634-5668 or Telephone 262-634-1555. For questions contact Ad-Tech Clinical Specialist. .

Device

  • Model / Serial
    Lot Numbers: 208140507, 208140552, 208140556, 208140559, 208140560, 208140561, 208140562, 208140563, 208140564R, 208140579, 208140581, 208140586, 208140588R, 208140590, 208140604.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Distributed in the states of GA and MI, and the countries of FRANCE, UNITED KINGDOM, COLOMBIA, and BRAZIL.
  • Product Description
    Ad-Tech Electrode Connection System/Cables. Lightweight TECH-ATTACH Cable Models: || 1. L-DCL-4DINX (4 contacts: 1 terminal block with 4 connectors each); || 2. L-DCL-6DINX (6 contacts: 1 terminal block with 6 connectors each); || 3. L-DCL-8DINX (8 contacts: 1 terminal block with 8 connectors each); || 4. L-DCL-16BDINX (16 contacts: 1 terminal block with 16 connectors each); || 5. L-DCL-32BDINX (32 contacts: 2 terminal blocks with 16 connectors each); || 6. L-DCL-64BDINX (64 contacts: 4 terminal blocks with 16 connectors each). || FO TECH-ATTACH Connection Cable model FO-LDC-4DINX (4 contacts: 1 terminal block with 4 connectors each). || Tech-Attach: these cables require the use of an additional component - the connector block. The tail of the electrode is inserted into the block and the block in turn mates with the cable. A key-pin is used to ensure proper orientation. The cable terminated in EEG compatible connector sockets. || Lightweight CABRIO Cable, model L-SRL-64BDINX (64 contacts: 4 terminal blocks with 16 connectors each). Cabrio cables are similar to Tech-Attach cables except that the block portion is pre-attached. The tail of an electrode is inserted into this portion which then hinges shut.