Recall of Device Recall adaPT insight (12C), v1.3.2

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ion Beam Applications S.A..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69781
  • Event Risk Class
    Class 2
  • Event Number
    Z-0836-2015
  • Event Initiated Date
    2014-11-07
  • Event Date Posted
    2014-12-19
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-06-29
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, radiation therapy, charged-particle, medical - Product Code LHN
  • Reason
    Two issues were identified: 1. in the release of adapt insight v1.3.2., the kv/kv single source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. a gantry collision incident required the iba service team to verify the gantry alignment and protons i x-ray beams colinearity, du.
  • Action
    The firm, IBA, sent a " User Notice" dated November 7, 2014, to the sole affected site-UPHS Customer. The notice describes the product, problem and actions to be taken. The customer was instructed that the KV/kV source acquisition mode has been disabled and shall not be used for patient alignment until system is officially confirmed to be back within tolerances. IBA will change the kV/kV Single Source acquisition workflow configuration file to be linked to the correct geometrical calibration file and will update the kV/kV Single Source calibration file that includes the colinearity parameters, with the correct values. If you need assistance, contact the IBA Site Staff or the Operations manager at +32 10 475 997 or email: Sophie.dessart@iba-group.com.

Device

  • Model / Serial
    PAT.109 (US), Treatment room 4
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    US Distribution in PA only.
  • Product Description
    adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ion Beam Applications S.A., Chemin du Cyclotron, 3, Louvain La Neuve Belgium
  • Manufacturer Parent Company (2017)
  • Source
    USFDA