Recall of Device Recall AccuCheK Inform II Blood Glucose Monitor

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Diagnostics Operations, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    64460
  • Event Risk Class
    Class 2
  • Event Number
    Z-1113-2013
  • Event Initiated Date
    2013-03-06
  • Event Date Posted
    2013-04-15
  • Event Status
    Terminated
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, test, blood glucose, over the counter - Product Code NBW
  • Reason
    Roche investigation into complaints found accu-chek inform ii and coaguchek xs pro system barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.
  • Action
    Consignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a day, seven days a week at 1-800-440-3638 for Accu check questions and 1-800-428-4674 for CoaguChek questions about the information contained in this UMDC.

Device

Manufacturer

  • Manufacturer Address
    Roche Diagnostics Operations, Inc., 9115 Hague Rd, Indianapolis IN 46256-1025
  • Manufacturer Parent Company (2017)
  • Source