Recall of Device Recall Access Ostase QC

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    61214
  • Event Risk Class
    Class 2
  • Event Number
    Z-1239-2012
  • Event Initiated Date
    2011-11-10
  • Event Date Posted
    2012-03-21
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-06-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Single (specified) analyte controls (assayed and unassayed) - Product Code JJX
  • Reason
    The recall was initiated because beckman confirmed that the access ostase calibrator and qc lots do not meet expiration date claims. the access ostase reagent kit lots identified were quality control tested and released using access ostase calibrator and qc lots that contain the implicated raw material.
  • Action
    Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated November 9, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard all remaining inventory. In addition, a Response Form was enclosed for customers to complete and return. Contact Beckman's Customer Support Center at (800) 854-3633 for questions regarding this notice.

Device

  • Model / Serial
    Lot Numbers: Expiration dates: 009091 10/10/2010  109636 09/30/2011 013491 10/31/2010  113465 10/31/2011 013492 11/30/2010  113969 10/31/2011 014488 12/31/2010  114261 11/30/2011 015211 02/28/2011  115485 12/31/2011 018855 03/31/2011  116660 01/31/2012 020420 04/30/2011  118211 02/28/2012 021285 05/31/2011
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution-USA (nationwide) and the countries of Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, France, Georgia, Germany, Greece, Hong Kong, Hungary, India,Ireland, Italy, Japan, Jordan, Lebanon, Liechtenstein, Mexico, Morocco, Netherlands, New Zealand, Panama, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela.
  • Product Description
    Access Ostase QC, Part Number: 37309. || The Access Ostase QC is intended for monitoring system performance of the Access Ostase assay.