Recall of Device Recall ABX PENTRA Reagent Container

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Horiba Instruments, Inc dba Horiba Medical.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69773
  • Event Risk Class
    Class 2
  • Event Number
    Z-0902-2015
  • Event Initiated Date
    2014-11-13
  • Event Date Posted
    2015-01-02
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-03-19
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Reason
    Horiba medical is recalling the abx pentra 400 reagent container because they do not always sit properly into the reagent rack. in low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. the defect would result in a false high or false low results for hba1c and false high results for magnes.
  • Action
    A recall letter was sent on 11/13/14 to all customers who purchased the ABX PENTRA 400 Clinical Chemistry Analyzer. The letter informs the customer of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact their local HORIBA Medical representative.