Recall of Device Recall Abbott

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Abbott Laboratories HPD/ADD.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    26450
  • Event Risk Class
    Class 2
  • Event Number
    Z-1134-03
  • Event Initiated Date
    2003-06-09
  • Event Date Posted
    2003-06-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2004-08-27
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Set, Administration, Intravascular - Product Code FPA
  • Reason
    The drip chambers and/or adapters in the sets were manufactured with non-radiation grade plastic resin, resulting in discoloration of the resin.
  • Action
    The IV administration sets were recalled by letters dated 6/9/03, advising the accounts of the discoloration of the sight chamber and adapters due to the use of non-radiation grade plastic resin in the manufacture of these components of the sets and requesting the return of the sets for credit. These letters were accompanied by a customer ship history/business reply form, which listed the specific set list numbers and lot numbers shipped to the direct account. The direct accounts were requested to send copies of the enclosed recall letter with the listing of all affected sets to their subaccounts. Abbott sent recall letters dated 6/13/03 to those accounts who received I.V. set list 018810458, lot 910824W packed in cases with Liposyn II 20% Kits, list 97920403, lot 92552DF and list 97930403, lot 94553DF. These accounts were informed of the problem with the administration sets and were requested to remove the sets from the Liposyn II cases and return the sets for credit. Abbott sent recall letters dated 7/7/03 to distributors and veterinary clinics receiving the 12 additional lots of primary I.V. sets, list numbers 018810448 and 089620448, informing them of the problem with the drip chambers and requesting them to return the affected lots for credit. The distributors were requested to notify their customers. Abbott sent recall letters dated 7/7/03 to the third party manufacturers receiving the affected drip chamber components, informing them of the problem with the drip chambers and requesting the return of the affected lots. Abbott sent recall letters dated 8/19/03 to hospitals and distributors receiving the 9 additional lots of I.V. sets, list numbers 119440402, 119480402, 120300412, 122590402 and 123120401, informing them of the problem with the drip chambers and requesting them to return the affected lots for credit. The distributors were requested to notify their customers. Abbott sent recall letters dated 5/13/04 to hospitals and distributor

Device

  • Model / Serial
    Abbott Commodity 96-8366, Ashland Product #85-55960, lot numbers 85C51LBMP, 89A195IMP, 89B335IMP, 89B345IMP, 90A105IMP, 90A125IMP, 91A145IMP, 91A155IMP, 91A165IMP, 92A125IMP
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide and Internationally to Abbott affiliates in Argentina, Aruba, Australia, the Bahamas, Barbados, Bermuda, Bolivia, Brazil, Canada, Cayman Islands, Chile, Columbia, Costa Rica, Croatia, Curacao, Dominican Republic, Ecuador, Egypt, El Salvador, France, Greece, Guam, Guatemala, Haiti, Honduras, Hong Kong, Hungary, Italy, Jordan, Kuwait, Mexico, New Zealand, Nicaragua, Pakistan, Panama, Paraguay, Peru, Suadia Arabia, Singapore, Taiwan, Trinidad, Turkey, United Arab Emirates and Uruguay
  • Product Description
    96-8366 Drip Chambers, Convertible with Pierce Pin .138 Assembly; Abbott, Ashland, Ohio 44805; 650 drip chambers per case; Abbott Commodity 968366, Ashland Product #85-55960
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Abbott Laboratories HPD/ADD, 100/200 Abbott Park Road, Abbott Park IL 60064
  • Source
    USFDA