Recall of Device Recall 3.0T GEM Flex Coil

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by NeoCoil, LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    72976
  • Event Risk Class
    Class 2
  • Event Number
    Z-1027-2016
  • Event Initiated Date
    2015-12-18
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2016-10-12
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Coil, magnetic resonance, specialty - Product Code MOS
  • Reason
    Potential for higher than specified surface temperatures. use of 3.0t gem flex system may result in patient warming or burns.
  • Action
    Distributor was sent a NeoColi "Urgent Medical Device Recall" letter dated 18 December 2015. The letter described the problem and the product involved in the recall. Consignee was advised to quarantine the product and to identify their customers and provide a list to NeoCoil. NeoCoil will contact each customer for correction and removal of the devices. Requested consignee to complete and return the response form. For questions call Michael Leigh at 261-522-6127.

Device

  • Model / Serial
    Mfg Lot or Serial # System ID 0000034-4-0130 608263MR5W 0000040-4-0135 608263MR5W 000032E-4-0151 608263MR5W 033-4-0151 608263MR5W 0000040-4-0202 602863MR 0000033-4-0204 602863MR 034-4-0265 602863MR 040-4-0076 602863MR 0000034-4-0250 703478MR2DV 0000040-4-0270 703478MR2DV 033-4-022 703478MR2DV 0000033-4-0272 703478MR2DV 034-4-0255 506648MR4 040-4-0275 506648MR4 040-4-0020 506648MR4 033-4-0277 506648MR4 034-4-0071 608265MR750W 0000040-4-0381 608265MR750W 0000033-4-0384 608265MR750W 034-4-0345 608265MR750W 0000034-4-0093 708346MR3 0000040-4-0096 708346MR3 0000033-4-0088 708346MR3 034-4-0352 269DVMR 040-4-0392 269DVMR 033-4-0395 269DVMR 0000034-4-0144 808873MR750W 0000033-4-0151 808873MR750W 0000040-4-0150 808873MR750W 0000034-4-0112 270798MR2 0000040-4-0118 270798MR2 0000033-4-0118 270798MR2 034-4-0089 650723NCMR1 033-4-0258 650723NCMR1 033-4-0134 650723NCMR1 0000034-4-0086 281274MR2 0000040-4-0089 281274MR2 0000033-4-0090 281274MR2 0000034-4-0076 757827MR2 0000033-4-0081 757827MR2 0000SGH225PYJJ 757827MR2 0000034-4-0007 305461MR1 0000040-4-0070 305461MR1 0000033-4-0070 305461MR1 0000034-4-0110 847872ZMR2 0000040-4-0114 847872ZMR2 0000033-4-0114 847872ZMR2 0000034-4-0095 330344MR750W 0000040-4-0098 330344MR750W 0000033-4-0099 330344MR750W 0000034-4-0094 616774MR750W 0000040-4-0097 616774MR750W 0000033-4-0098 616774MR750W 0000040-4-0276 386FNSBMR3T 0000033-4-0279 386FNSBMR3T 0000034-4-0260 386FNSBMR3T 0000034-4-0140 217876DMH3T 0000040-4-0145 217876DMH3T 0000033-4-0146 217876DMH3T 0000034-4-0216 410328MR3T 0000040-0-0224 410328MR3T 0000033-4-0227 410328MR3T 034-4-0313 716740MR1 040-4-0329 716740MR1 033-4-0329 716740MR1 0000034-4-0177 415CBMRP 0000040-4-0182 415CBMRP 0000033-4-0183 415CBMRP 0000034-4-0043 808485MR2 0000040-4-0047 808485MR2 033-4-0323 808485MR2 034-4-0267 415UCBHMR1 040-4-0284 415UCBHMR1 033-4-0285 415UCBHMR1 033-4-0352 843815750W 040-4-0055 843815750W 0000034-4-0221 843815750W 034-4-0272 415UCBHMR2 040-4-0288 415UCBHMR2 033-4-0290 415UCBHMR2 0000034-4-0108 856581MR3T 4040109 856581MR3T 0000033-4-0112 856581MR3T 034-4-0380 423493MR750W 040-4-0432 423493MR750W 033-4-0436 423493MR750W 0000034-4-0099 RADNETMR1648 0000040-4-0102 RADNETMR1648 0000033-4-0102 RADNETMR1648 034-4-0369 480342MR3 044-4-0403 480342MR3 033-4-0405 480342MR3 3440372 617726MR1A2 4040420 617726MR1A2 3340423 617726MR1A2 034-4-0069 503251PMR3 040-4-0322 503251PMR3 033-4-0144 503251PMR3 3440200 702382MR750W 4040192 702382MR750W 3340199 702382MR750W 0000-34-4-0235 207MMC3TMR 0000040-4-0249 207MMC3TMR 0000033-4-0252 207MMC3TMR 0000040-4-0291 703573MR1DV 0000033-4-0292 703573MR1DV 0000034-4-0275 703573MR1DV 0000034-4-0189 513636MR4 0000040-4-0196 513636MR4 0000033-4-0198 513636MR4 0000034-4-0188 708346MR4 0000040-4-0195 708346MR4 0000033-4-0197 708346MR4 0000034-4-0055 559734VIMR1 0000040-4-0056 559734VIMR1 0000033-4-0057 559734VIMR1 034-4-0384 720848MR4 040-4-0425 720848MR4 033-4-0428 720848MR4 0000034-4-0143 214645UT3TMR1 0000034-4-0149 214645UT3TMR1 0000034-4-0150 214645UT3TMR1 0000034-4-0115 757983MR 0000033-4-0120 757983MR 0000034-4-0015 757983MR 0000034-4-0037 604875MR750 0000040-4-0033 604875MR750 0000033-4-0044 604875MR750 0000034-4-0033 808545MR 0000040-4-0068 808545MR 0000033-4-0012 808545MR 0000034-4-0282 608262PETMR1 0000040-4-0299 608262PETMR1 0000033-4-0300 608262PETMR1 0000034-4-0138 808YMMR 0000040-4-0148 808YMMR 033-4-0081 808YMMR 034-4-0365 215829MR750W 040-4-0416 215829MR750W 033-4-0420 215829MR750W 034-4-0361 847316MR3T 040-4-0412 847316MR3T 033-4-0414 847316MR3T 0000034-4-0245 914750W 0000040-4-0264 914750W 0000033-4-0267 914750W 0000034-4-0147 856247MR1 0000040-4-0155 856247MR1 0000033-4-0155 856247MR1 0000034-4-0199 937641MR2 0000040-4-0204 937641MR2 0000033-4-0207 937641MR2 034-4-0379 859334MR 040-4-0448 859334MR 033-4-0451 859334MR 0000034-4-0166 COR526967 0000040-4-0179 COR526967 0000033-4-0191 COR526967 0000034-4-0058 RADNETMR1631 0000040-4-0059 RADNETMR1631 0000033-4-0059 RADNETMR1631 034-4-0405 CHOPPETMR 040-4-0467 CHOPPETMR 033-4-4070 CHOPPETMR 034-4-0338 608890MR750 040-4-0367 608890MR750 033-4-0370 608890MR750 034-4-0296 E416249MR2 040-4-0315 E416249MR2 033-4-0314 E416249MR2 0000034-4-0180 608CHILDMR1 0000040-4-0184 608CHILDMR1 0000033-4-0185 608CHILDMR1 0000034-4-0009 RADNETMR1632 0000040-4-0016 RADNETMR1632 0000033-4-0018 RADNETMR1632 0000034-4-0264 6142343T 0000040-4-0281 6142343T 0000033-4-0282 6142343T 034-4-0284 614293MR9 040-4-0304 614293MR9 033-4-0302 614293MR9 034-4-0394 650PH3T 040-4-0451 650PH3T 033-4-0454 650PH3T 0000033-4-0253 843750WGEM 00000040-40250 843750WGEM 033-4-0410 212288MR4 040-4-0407 212288MR4 0000033-4-0040 608WIMRMR2 0000040-4-0038 608WIMRMR2 034-4-0269 415UCBHMR3 040-4-0286 415UCBHMR3 0000033-4-0124 414647MR1 0000040-4-0116 414647MR1 034-4-0150 301570MR 0000034-4-0150 301570MR 033-4-0417 608HERIMR 040-4-0413 608HERIMR 033-3-0484 303492MR 040-4-0482 303492MR 00000040-4-289 865984MR750 0000033-4-0289 865984MR750 0000033-4-0180 713798MR1 0000040-4-0180 713798MR1 033-4-0382 970624MR1 040-4-0380 970624MR1 0000033-4-0224 715847MR3T 0000040-4-0216 715847MR3T 033-4-0430 205592750 040-4-0427 205592750 0000033-4-0268 309655MR750W 0000040-4-0265 309655MR750W 0000033-4-0159 337289MR1 0000040-4-0160 337289MR1 033-4-0372 718204MR750W 040-4-0369 718204MR750W 0000040-4-0274 702731MR750W 0000033-4-0276 702731MR750W 0000034-4-0142 312355DVMR 0000033-4-0140 312355DVMR 033-4-0446 860446MR750W 040-4-0443 860446MR750W 033-4-0391 734845VA750MR 040-4-0390 734845VA750MR 033-4-0361 904483MR3T 040-4-0358 904483MR3T 0000033-4-0075 314577CGMR3T 0000040-4-0077 314577CGMR3T 0000033-4-0246 949760MR3 0000040-4-0243 949760MR3 040-4-0039 317865MR4 000032E-4-0302 317865MR4 0000034-4-0160 BRIGHAM750WMR1 3440160 BRIGHAM750WMR1 033-4-0444 773665SJMR2 040-4-0440 773665SJMR2 0000040-4-0172 212543MR750 0000033-4-0173 212543MR750 0000033-4-0288 317887MR3 0000040-4-0287 317887MR3 0000033-4-0121 671FHPMR750W 0000040-4-0122 671FHPMR750W 033-4-0360 318443MR2 040-4-0357 318443MR2 033-4-0398 608VAMR750W 040-4-0397 608VAMR750W 034-4-0294 MR229049 0000040-4-0066 206292SR750W 0000034-4-0080 RADNETMR1639 4040197 201445750W 0000033-4-0287 415UCBHMR3 0000034-4-0198 913322MISHMRW 0000034-4-0070 6033343TW 0000034-4-0049 405271OUCMR 0000034-4-0109 RADNETMR1652 034-4-0274 412914MR 0000034-4-0116 202966MR3TW 0000040-4-0104 703698MR4 0000034-4-0269 415UCBHMR3 0000034-4-0233 954355MR750W 0000034-4-0262 PRESBYMR1 0000040-4-0075 856247MR3 040-4-0169 864529MR 0000034-4-0100 BOSTONMEDMR1 040-4-0198 301295MR3T2 0000034-4-0124 VAWESTROX750W 3340162 RADNETMR613 0000034-4-0196 217528MR1 040-4-0349 817433HSWMR2 0000034-4-0156 858622HDMR1 0000040-4-0076 0920279031 0000033-4-0032 0920279031 0000034-4-0040 0920279031 0000040-4-0022 0920279031 0000040-4-0248 EM0253 0000033-4-0251 EM0253 0000034-4-0232 EM0253 0000034-4-0290 EM0253 0000040-4-0185 EM0271 0000033-4-0162 EM0271 0000033-4-0186 EM0271 0000034-4-0181 EM0271 0000040-4-0101 EM0223 0000033-4-0101 EM0223 0000034-4-0098 EM0223 0000040-4-0055 M2322246 0000033-4-0016 M2322246 0000034-4-0054 M2322246 0000040-4-0094 0850270411 0000033-4-0095 0850270411 0000034-4-0091 0850270411 0000040-4-0159 0850270430 0000033-4-0158 0850270430 0000034-4-0151 0850270430 0000040-4-0079 PC9595MR01 0000033-4-0079 PC9595MR01 0000034-4-0074 PC9595MR01 0000040-4-0194 0850270436 0000033-4-0196 0850270436 0000034-4-0187 0850270436 0000040-4-0158 EM0234 0000033-4-0176 EM0234 0000034-4-0170 EM0234 0000040-4-0272 0850270455 0000033-4-0274 0850270455 0000034-4-0256 0850270455 0000040-4-0019 HU1098MR02 0000033-4-0020 HU1098MR02 0000034-4-0020 HU1098MR02 0000040-4-0472 0850270478 0000033-4-0474 0850270478 0000034-4-0407 0850270478 0000040-4-0146 MRUA0184 0000033-4-0145 MRUA0184 0000034-4-0139 MRUA0184 0000040-4-0045 0910274040 0000033-4-0046 0910274040 0000034-4-0041 0910274040 0000040-4-0069 UDN0121 0000033-4-0069 UDN0121 0000034-4-0066 UDN0121 0000040-4-0254 0920279028 0000033-4-0257 0920279028 0000034-4-0238 0920279028 0000040-4-0111 EM0230 0000033-4-0111 EM0230 0000034-4-0107 EM0230 0000040-4-0127 0850270409 0000033-4-0127 0850270409 0000034-4-0120 0850270409 0000040-4-0210 EM0239 0000033-4-0214 EM0239 0000034-4-0201 EM0239 0000040-4-0228 082427040146 0000033-4-0232 082427040146 0000034-4-0219 082427040146 0000040-4-0295 EM0258 0000033-4-0297 EM0258 0000034-4-0276 EM0258 0000040-4-0273 082427060052 0000033-4-0275 082427060052 0000034-4-0254 082427060052 0000040-4-0375 KW1100MR02 0000033-4-0378 KW1100MR02 0000034-4-0341 KW1100MR02 0000040-4-0285 082427100219 0000033-4-0286 082427100219 0000034-4-0270 082427100219 0000040-4-0449 M4035023 0000033-4-0452 M4035023 0000034-4-0391 M4035023 0000040-4-0290 082427110051 0000033-4-0291 082427110051 0000034-4-0273 082427110051 0000040-4-0419 MRUA0459 0000033-4-0422 MRUA0459 0000034-4-0370 MRUA0459 0000040-4-0163 082427120110 0000033-4-0165 082427120110 0000034-4-0159 082427120110 0000040-4-0181 QA1042MR01 0000033-4-0182 QA1042MR01 0000034-4-0176 QA1042MR01 0000040-4-0246 082427120116 0000033-4-0249 082427120116 0000034-4-0231 082427120116 0000040-4-0460 ZC0002 0000033-4-0462 ZC0002 0000034-4-0399 ZC0002 0000040-4-0269 082427120124 0000033-4-0273 082427120124 0000034-4-0252 082427120124 0000040-4-0147 EM0227 0000033-4-0147 EM0227 0000034-4-0141 EM0227 0000040-4-0309 082427120131 0000033-4-0306 082427120131 0000034-4-0288 082427120131 0000040-4-0174 EM0232 0000033-4-0175 EM0232 0000034-4-0168 EM0232 0000040-0-0303 082427130076 0000033-4-0309 082427130076 0000034-4-0291 082427130076 0000040-4-0170 EM0236 0000033-4-0171 EM0236 0000034-4-0164 EM0236 0000040-4-0156 83092403928 0000033-4-0156 83092403928 0000034-4-0148 83092403928 0000040-4-0221 EM0242 0000033-4-0222 EM0242 0000034-4-0202 EM0242 0000040-4-0165 83096535911 0000033-4-0187 83096535911 0000034-4-0178 83096535911 0000040-4-0231 EM0256 0000033-4-0234 EM0256 0000034-4-0220 EM0256 0000040-4-0279 EM0265 0000033-4-0280 EM0265 0000034-4-0266 EM0265 0000040-4-0082 0850270414 0000033-4-0082 0850270414 0000034-4-0077 0850270414 0000040-4-0282 EM0275 0000033-4-0283 EM0275 0000034-4-0265 EM0275 0000040-4-0211 83096536248 0000033-4-0215 83096536248 0000034-4-0203 83096536248 0000040-4-0405 EM0281 0000033-4-0407 EM0281 0000034-4-0377 EM0281 0000040-4-0260 83104653897 0000033-4-0263 83104653897 0000034-4-0246 83104653897 0000040-4-0346 KA1023MR02 0000033-4-0345 KA1023MR02 0000034-4-0319 KA1023MR02 0000040-4-0241 83105089955 0000033-4-0244 83105089955 0000034-4-0326 83105089955 0000040-4-0027 LB1331MR01 0000033-4-0030 LB1331MR01 0000034-4-0030 LB1331MR01 0000040-4-0262 83108956002 0000033-4-0265 83108956002 0000034-4-0243 83108956002 0000040-4-0298 M4018549 0000033-4-0299 M4018549 0000034-4-0281 M4018549 0000040-4-0311 83112923777 0000033-4-0310 83112923777 0000034-4-0292 83112923777 0000040-4-0083 MPX10000 0000033-4-0083 MPX10000 0000034-4-0078 MPX10000 000040-04-0429 85110163303 0000033-4-0432 85110163303 0000034-4-0371 85110163303 0000040-4-0310 MRUA0343 0000033-4-0307 MRUA0343 0000034-4-0268 MRUA0343 0000040-4-0453 083027020204715 0000033-4-0427 083027020204715 0000033-4-0456 083027020204715 0000034-4-0378 083027020204715 0000040-4-0424 083027020204715 0000040-4-0136 O009MR03 0000033-4-0136 O009MR03 0000034-4-0128 O009MR03 0000040-4-0187 083027200034313 0000033-4-0189 083027200034313 0000034-4-0184 083027200034313 0000040-4-0297 PL3277MR01 0000033-4-0296 PL3277MR01 0000034-4-0289 PL3277MR01 0000040-4-0452 06008MRS03 0000033-4-0455 06008MRS03 0000034-4-0395 06008MRS03 0000040-4-0188 T002MR01 0000033-4-+190 T002MR01 0000034-4-0183 T002MR01 0000040-4-0026 120810MR02 0000033-4-0028 120810MR02 0000034-4-0028 120810MR02 0000040-4-0383 X41944010 0000033-4-0385 X41944010 0000034-4-0346 X41944010 0000040-4-0166 120810MR03 0000033-4-0166 120810MR03 0000034-4-0161 120810MR03 0000040-4-0133 EM0221 0000033-4-0132 EM0221 0000034-4-0126 EM0221 0000040-4-0065 140024MR01 0000033-4-0066 140024MR01 0000034-4-0065 140024MR01 0000040-4-0144 EM0225 0000033-4-0144 EM0225 0000034-4-0137 EM0225 0000040-4-0378 150250MR01 0000033-4-0399 150250MR01 0000034-4-0343 150250MR01 0000040-4-0164 EM0229 0000033-4-0168 EM0229 0000034-4-0162 EM0229 0000040-4-0498 157022MR01 0000033-4-0501 157022MR01 0000034-4-0425 157022MR01 0000040-4-0162 EM0231 0000033-4-0164 EM0231 0000034-4-0158 EM0231 0000040-4-0408 20312MRS01 0000033-4-0410 20312MRS01 0000034-4-0368 20312MRS01 0000040-4-0168 EM0233 0000033-4-0169 EM0233 0000034-4-0163 EM0233 0000040-4-0219 220078MR01 0000033-4-0221 220078MR01 0000034-4-0208 220078MR01 0000040-4-0171 EM0235 0000033-4-0172 EM0235 0000034-4-0165 EM0235 0000040-4-0339 220826MR02 0000033-4-0339 220826MR02 0000034-4-0308 220826MR02 0000040-4-0275 EM0237 0000033-4-0194 EM0237 0000034-4-0185 EM0237 0000040-4-0099 26001MRS01 0000033-4-0097 26001MRS01 0000034-4-0096 26001MRS01 0000040-4-0186 EM0240 0000033-4-0188 EM0240 0000034-4-0182 EM0240 0000040-4-0464 34026MRS05 0000033-4-0466 34026MRS05 0000034-4-0403 34026MRS05 0000040-4-0238 EM0243 0000033-4-0241 EM0243 0000034-4-0222 EM0243 0000040-4-0078 34348MRS01 0000033-4-0076 34348MRS01 0000034-4-0072 34348MRS01 0000040-4-0234 EM0255 0000033-4-0237 EM0255 0000034-4-0224 EM0255 0000040-4-0134 387040MR10 0000033-4-0134 387040MR10 0000034-4-0127 387040MR10 0000040-4-0256 EM0257 0000033-4-0259 EM0257 0000034-4-0239 EM0257 0000040-4-0153 600042MR02 0000033-4-0154 600042MR02 0000034-4-0146 600042MR02 0000040-4-0230 EM0260 0000033-4-0233 EM0260 0000034-4-0218 EM0260 0000040-4-0018 70431MRS04 0000033-4-0026 70431MRS04 0000034-4-0026 70431MRS04 0000040-4-0393 EM0267 0000033-4-0394 EM0267 0000034-4-0279 EM0267 0000040-4-0263 725100MR02 0000033-4-0266 725100MR02 0000034-4-0244 725100MR02 0000040-4-0445 EM0274 0000033-4-0448 EM0274 0000034-4-0390 EM0274 0000040-4-0257 77266MRS03 0000033-4-0260 77266MRS03 0000034-4-0241 77266MRS03 0000040-4-0428 EM0280 0000033-4-0431 EM0280 0000034-4-0381 EM0280 0000040-4-0014 850020MR02 0000033-4-0017 850020MR02 0000034-4-0017 850020MR02 0000040-4-0463 EX0130 0000033-4-0465 EX0130 0000034-4-0402 EX0130 0000040-4-0414 A4493006 0000033-4-0418 A4493006 0000034-4-0363 A4493006 0000040-4-0178 IE1010MR01 0000033-4-0179 IE1010MR01 0000034-4-0172 IE1010MR01 0000040-4-0307 AC001MR02 0000033-4-0308 AC001MR02 0000034-4-0286 AC001MR02 0000040-4-0244 KW1019MR03 0000033-4-0247 KW1019MR03 0000034-4-0229 KW1019MR03 0000040-4-0400 AP7645MR01 0000033-4-0402 AP7645MR01 0000034-4-0353 AP7645MR01 0000040-4-0139 LB1160MR02 0000033-4-0139 LB1160MR02 0000034-4-0132 LB1160MR02 0000040-4-0371 B5129925 0000033-4-0374 B5129925 0000034-4-0339 B5129925 0000040-4-0128 M004MR01 0000033-4-0128 M004MR01 0000034-4-0119 M004MR01 0000040-4-0266 B5174327 0000033-4-0269 B5174327 0000034-4-0247 B5174327 0000040-4-0388 M345477701 0000033-4-0390 M345477701 0000034-4-0350 M345477701 0000040-4-0215 BH1017MR01 0000033-4-0226 BH1017MR01 0000034-4-0215 BH1017MR01 0000040-4-0411 M40345211 0000033-4-0412 M40345211 0000034-4-0360 M40345211 0000040-4-0251 C001MR01 0000033-4-0254 C001MR01 0000034-4-0237 C001MR01 0000040-4-0305 M54606125 0000033-4-0303 M54606125 0000034-4-0285 M54606125 0000040-4-0074 C4200438 0000033-4-0073 C4200438 0000034-4-0069 C4200438 0000040-4-0161 MRUA0122 0000033-4-0161 MRUA0122 0000034-4-0152 MRUA0122 4040058 DPN26332 3340060 DPN26332 3440059 DPN26332 0000040-4-0258 MRUA0191 0000033-4-0261 MRUA0191 0000034-4-0242 MRUA0191 0000040-4-0423 DPN36500 0000033-4-0426 DPN36500 0000034-4-0366 DPN36500 0000040-4-0308 MRUA0353 0000033-4-0305 MRUA0353 0000034-4-0287 MRUA0353 0000040-4-0034 E001MR02 0000033-4-0062 E001MR02 0000034-4-0062 E001MR02 0000040-4-0293 N002MR03 0000033-4-0293 N002MR03 0000034-4-0277 N002MR03 0000040-4-0213 EM0025 0000033-4-0217 EM0025 0000034-4-0205 EM0025 0000040-4-0138 PC3003MR01 0000033-4-0138 PC3003MR01 0000034-4-0131 PC3003MR01 0000040-4-0061 EM0201 0000033-4-0011 EM0201 0000034-4-0061 EM0201 0000040-4-0176 PL3116MR01 0000033-4-0181 PL3116MR01 0000034-4-0175 PL3116MR01 0000040-4-0085 EM0207 0000033-4-0085 EM0207 0000034-4-0079 EM0207 0000040-4-0012 PNUE08MR01 0000033-4-0025 PNUE08MR01 0000034-4-0006 PNUE08MR01 0000040-4-0106 EM0209 0000033-4-0106 EM0209 0000034-4-0102 EM0209 0000040-4-0267 SA1255MR03 0000033-4-0270 SA1255MR03 0000034-4-0248 SA1255MR03 0000040-4-0100 EM0212 0000033-4-0100 EM0212 0000034-4-0097 EM0212 0000040-4-0455 TP0008MR07 0000033-4-0467 TP0008MR07 0000034-4-0404 TP0008MR07 0000040-4-0108 EM0214 0000033-4-0108 EM0214 0000034-4-0104 EM0214 0000040-4-0296 UE929526 0000033-4-0298 UE929526 0000034-4-0280 UE929526 0000040-4-0140 EM0218 0000033-4-0092 EM0218 0000034-4-0087 EM0218 0000040-4-0479 ZC0001 0000033-4-0481 ZC0001 0000034-4-0412 ZC0001 0000040-4-0120 EM0219 0000033-4-0119 EM0219 0000034-4-0114 EM0219 0000040-4-0115 EM0220 0000033-4-0116 EM0220 0000034-4-0111 EM0220 0000040-4-0151 EM0228 0000033-4-0152 EM0228 0000034-4-0145 EM0228 0000040-4-0355 RU8937MR01 0000033-4-0356 RU8937MR01 0000040-4-0050 N4192951 0000033-4-0056 N4192951 0000040-4-0199 MR308183MR3 0000033-4-0201 MR308183MR3 0000040-4-0236 83104203907 0000033-4-0239 83104203907 0000040-4-0456 BG4680MR01 0000033-4-0458 BG4680MR01 0000040-4-0401 EM0264 0000033-4-0403 EM0264 0000040-4-0253 EM0249 0000033-4-0256 EM0249 0000040-4-0377 EM0266 0000033-4-0380 EM0266 0000034-4-0136 082427040135 0000033-4-0143 082427040135 0000040-4-0409 EM0268 0000033-4-0411 EM0268 0000040-4-0442 082427160149 0000033-4-0441 082427160149 0000040-4-0084 0910273043 0000033-4-0084 0910273043 0000040-4-0232 RU8615MR01 0000033-4-0235 RU8615MR01 0000040-4-0441 EM0276 0000033-4-0445 EM0276 0000040-4-0214 EM0247 0000033-4-0218 EM0247 0000040-4-0422 EM0277 0000033-4-0425 EM0277 0000034-4-0023 X59007321 48 X59007321 0000040-4-0117 EM0213 0000033-4-0117 EM0213 0000040-4-0450 A1765918 0000033-4-0453 A1765918 0000040-4-0049 EM0215 0000033-4-0110 EM0215 0000040-4-0382 A5719869 0000033-4-0383 A5719869 0000040-4-0124 EM0217 0000033-4-0123 EM0217 0000040-4-0207 EM0238 0000033-4-0210 EM0238 0000040-4-0062 541010MR04 0000033-4-0063 541010MR04 0000040-4-0220 130083MR05 0000034-4-0209 130083MR05 0000040-4-0152 RU5599MR01 0000033-4-0153 RU5599MR01 0000040-4-0119 M001MR02 0000034-4-0113 M001MR02 0000040-4-0306 RU8792MR01 0000033-4-0304 RU8792MR01 0000040-4-0459 M2850013 0000033-4-0461 M2850013 0000040-4-0154 SA1223MR01 0000034-4-0154 SA1223MR01 0000040-4-0259 0850270446 0000033-4-0262 0850270446 0000040-4-0240 EM0248 0000033-4-0243 EM0248 0000040-4-0126 EM0226 0000033-4-0126 EM0226 0000040-4-0252 EM0254 0000033-4-0255 EM0254 0000040-4-0385 M4172341 0000033-4-0386 M4172341 0000040-4-0053 M45052106 0000034-4-0052 M45052106 0000040-4-0439 MRUA0394A 0000033-4-0443 MRUA0394A 00040-4-0008-2 082427040135AW2 0000033-4-0315 A5547422 0000034-4-0149 85092252519 0000034-4-0191 EM0005 0000034-4-0258 0853270141 0000034-4-0053 0850270386 0000040-4-0294 6073MR0003A 0000034-4-0213 EM0244 0000034-4-0212 EM0245 0000034-4-0206 EM0246 0000034-4-0106 0853270119 0000034-4-0122 EM0216 00000034-4-230 083027860052014 0000004-4-0242 MRUA0272 0000033-4-0037 83027840038612 0000033-4-0080 AE1016MR01 0000033-4-0141 0850270427 0000033-4-0160 SA1223MR01 0000033-4-0184 KW1005MR04 0000033-4-0192 RU8438MR01 0000033-4-0193 RU8438MR03 0000033-4-0213 130083MR05 0000033-4-0225 10692MRS01 0000033-4-0240 00230MRS01 0000033-4-0248 083027860052014 0000033-4-0278 A5111746 0000033-4-0379 083027806724415 0000033-4-0387 53008MRS01 0000033-4-0388 06162MRS01 0000033-4-0389 EG1526MR05 0000033-4-0396 083027046705715 0000033-4-0397 20312MRS01 0000033-4-0408 DZ1060MR04 0000033-4-0435 082427200086 0000033-4-0439 0850270459 0000033-4-0460 210002MR01 0000033-4-0475 082427100254 0000033-4-0478 MRWB0843 0000033-4-0488 082427010087 0000033-4-0491 00440MRS08 0000033-4-0497 10722MRS01 0000034-4-0398 M2850013 0000040-4-0041 083027840038612 00000032E40288 83096322711 0000032-E-0389 082427130076 000022E-4-0544 M007MR01 000032E-4-0015 PNUE08MR01 000032E-4-0021 LB1331MR01 000032E-4-0027 HU1098MR02 000032E-4-0031 X59007321 000032E-4-0037 70431MRS04 000032E-4-0039 120810MR02 000032E-4-0048 083027840038612 000032E-4-0050 0920279031 000032E-4-0051 0910274040 000032E-4-0056 M2322246 000032E-4-0058 N4192951 000032E-4-0060 0850270386 000032E-4-0068 EM0201 000032E-4-0069 850020MR01 000032E-4-0071 541010MR04 000032E-4-0073 140024MR01 000032E-4-0075 UDN0121 000032E-4-0079 C4200438 000032E-4-0081 0850270411 000032E-4-0085 EM0212 000032E-4-0086 34348MRS01 000032E-4-0088 0850270414 000032E-4-0090 PC9595MR01 000032E-4-0092 MPX10000 000032E-4-0093 M001MR02 000032E-4-0097 0910273043 000032E-4-0098 EM0207 000032E-4-0105 EM0218 000032E-4-0107 EM0223 000032E-4-0109 26001MRS01 000032E-4-0113 EM0209 000032E-4-0119 EM0214 000032E-4-0120 EM0215 000032E-4-0122 0853270119 000032E-4-0123 EM0230 000032E-4-0128 EM0220 000032E-4-0129 EM0213 000032E-4-0132 EM0219 000032E-4-0136 EM0226 000032E-4-0138 EM0217 000032E-4-0141 0850270409 000032E-4-0142 M004MR01 000032E-4-0144 EM0216 000032E-4-0148 EM0221 000032E-4-0149 387040MR10 000032E-4-0150 O009MR03 000032E-4-0152 PC3003MR01 000032E-4-0153 LB1160MR02 000032E-4-0158 082427040135 000032E-4-0159 EM0225 000032E-4-0160 MRUA0184 000032E-4-0162 EM0227 000032E-4-0168 EM0228 000032E-4-0169 RU5599MR01 000032E-4-0174 600042MR02 000032E-4-0175 83092403928 000032E-4-0176 85092252519 000032E-4-0179 0850270430 000032E-4-0181 MRUA0122 000032E-4-0183 SA1223MR01 000032E-4-0190 EM0231 000032E-4-0191 082427120110 000032E-4-0194 120810MR03 000032E-4-0196 EM0229 000032E-4-0197 EM0233 000032E-4-0198 EM0236 000032E-4-0199 EM0235 000032E-4-0202 EM0232 000032E-4-0203 EM0234 000032E-4-0206 IE1010MR01 000032E-4-0208 PL3116MR01 000032E-4-0209 QA1042MR01 000032E-4-0212 T002MR01 000032E-4-0214 EM0271 000032E-4-0215 83096535911 000032E-4-0216 EM0240 000032E-4-0217 083027200034313 000032E-4-0224 EM0237 000032E-4-0226 0850270436 000032E-4-0228 EM0005 000032E-4-0233 MR308183MR3 000032E-4-0244 EM0238 000032E-4-0247 EM0239 000032E-4-0249 83096536248 000032E-4-0251 EM0025 000032E-4-0253 EM0246 000032E-4-0256 220078MR01 000032E-4-0257 130083MR05 000032E-4-0258 EM0242 000032E-4-0261 BH1017MR01 000032E-4-0264 EM0245 000032E-4-0265 EM0244 000032E-4-0272 082427040146 000032E-4-0273 EM0260 000032E-4-0275 RU8615MR01 000032E-4-0277 EM0256 000032E-4-0281 83104203907 000032E-4-0285 EM0255 000032E-4-0286 EM0243 000032E-4-0287 EM0248 000032E-4-0289 EM0247 000032E-4-0293 83105089955 000032E-4-0294 KW1019MR03 000032E-4-0296 082427120116 000032E-4-0298 EM0253 000032E-4-0305 C001MR01 000032E-4-0306 EM0249 000032E-4-0307 EM0254 000032E-4-0311 EM0257 000032E-4-0313 MRUA0191 000032E-4-0314 0850270446 000032E-4-0315 83108956002 000032E-4-0318 725100MR02 000032E-4-0320 83104653897 000032E-4-0322 B5174327 000032E-4-0323 SA1255MR03 000032E-4-0328 082427120124 000032E-4-0332 082427060052 000032E-4-0334 0850270455 000032E-4-0337 0853270141 000032E-4-0341 E001MR02 000032E-4-0349 EM0265 000032E-4-0352 EM0275 000032E-4-0357 082427100219 000032E-4-0364 082427110051 000032E-4-0368 N002MR03 000032E-4-0371 EM0258 000032E-4-0373 UE929526 000032E-4-0376 M4018549 000032E-4-0378 6073MR0007 000032E-4-0381 M54606125 000032E-4-0382 RU8792MR01 000032E-4-0383 AC001MR02 000032E-4-0384 MRUA0353 000032E-4-0385 PL3277MR01 000032E-4-0386 082427120131 000032E-4-0387 MRUA0343 000032E-4-0390 83112923777 000032E-4-0403 EM0262 000032E-4-0406 M365865301 000032E-4-0407 082427180028 000032E-4-0418 EM0259 000032E-4-0421 EM0261 000032E-4-0422 EM0270 000032E-4-0425 220826MR02 000032E-4-0426 KA1023MR02 000032E-4-0427 082427040167 000032E-4-0428 83112923995 000032E-4-0430 082427100225 000032E-4-0440 EM0263 000032E-4-0441 083027244278914 000032E-4-0443 0920279028 000032E-4-0446 RU8937MR01 000032E-4-0455 082427120122 000032E-4-0457 082427120135 000032E-4-0460 082427010080 000032E-4-0467 B5129925 000032E-4-0471 KW1100MR02 000032E-4-0474 EM0266 000032E-4-0479 A5719869 000032E-4-0480 X41944010 000032E-4-0481 M4172341 000032E-4-0485 EG1526MR05 000032E-4-0486 M345477701 000032E-4-0492 EM0267 000032E-4-0494 259130MR12 000032E-4-0496 150250MR01 000032E-4-0499 AP7645MR01 000032E-4-0502 EM0264 000032E-4-0511 EM0268 000032E-4-0513 M40345211 000032E-4-0515 20312MRS01 000032E-4-0518 A4493006 000032E-4-0525 MRUA0459 000032E-4-0528 EM0003 000032E-4-0530 EM0277 000032E-4-0531 DPN36500 000032E-4-0532 EM0281 000032E-4-0539 85110163303 000032E-4-0545 EM0280 000032E-4-0554 MRUA0394A 000032E-4-0556 EM0276 000032E-4-0561 082427160149 000032E-4-0563 M4035023 000032E-4-0567 EM0274 000032E-4-0569 EM0273 000032E-4-0572 A1765918 000032E-4-0577 083027020204715 000032E-4-0579 BG4693MR01 000032E-4-0580 EM0269 000032E-4-0584 M2850013 000032E-4-0585 ZC0002 000032E-4-0588 EX0130 000032E-4-0589 34026MRS05 000032E-4-0592 TP0008MR07 000032E-4-0595 EX0131 000032E-4-0598 0850270478 000032E-4-0610 ZC0001 000032E-4-0634 34225MRS02 000032E-4-0637 157022MR01 000032E-4-0014 140024MR01 000032E-4-0010 EM0223 000032E-4-0220 EM0237 000032E-4-0271 EM0260 000032E-4-0274 EM0256 000032E-4-0280 EM0255 000032E-4-0283 EM0243 000032E-4-0252 EM0247 000032E-4-0290 83105089955 000032E-4-0309 EM0257 000032E-4-0316 83104653897 000032E-4-0329 0850270455 000032E-4-0344 EM0265 000032E-4-0348 EM0275 000032E-4-0366 N002MR03 000032E-4-0367 EM0258 000032E-4-0375 M4018549 000032E-4-0380 M54606125 000032E-4-0374 PL3277MR01 00000032E-0386 082427120131 000032E-4-0356 MRUA0343 000032E-4-0405 M365865301 000032E-4-0408 EM0259 000032E-4-0415 220826MR02 000032E-4-0462 EM0266 000032E-4-0478 X41944010 000032E-4-0450 M4172341 000032E-4-0372 EM0267 278 150250MR01 000032E-4-0495 AP7645MR01 000032E-4-0483 20312MRS01 000032E-4-0529 EM0277 000032E-4-0521 DPN36500 000032E-4-0505 EM0281 000032E-4-0526 85110163303 000032E-4-0538 EM0280 000032E-4-0560 EM0274 000032E-4-0533 083027020204715 000032E-4-0559 EM0269 000032E-4-0582 ZC0002 000032E-4-0413 EX0130 000032E-4-0590 TP0008MR07 000032E-4-0607 ZC0001 000032E-4-0633 157022MR01 000032E-4-0509 20312MRS01 032E-4-0330 506648MR4 032E-4-0397 412914MR 033-04-0319 920738MR750 032E-4-0412 716740MR1 032E-4-0379 614293MR9 032e-4-0189 RADNETMR621 032E-4-0442 248533750W 000032E-4-0438 336832MR750 032E-4-0433 206CH750MR 032E-4-0487 734845VA750MR 000032E-4-0112 BOSTONMEDMR1 000032E-4-0193 BRIGHAM750WMR1 000032E-4-0447 MSK53MR02 032E-4-0477 970624MR1 032E-4-0358 415UCBHMR3 032E-4-0354 415UCBHMR1 032E-4-0362 415UCBHMR2 032e-4-0411 503251PMR3 032E-4-0465 718204MR750W 032E-4-0392 MR229049 032E-4-0488 269DVMR 032E-4-0207 201445750W 032e-4-0359 516NS7113T 032E-4-0368 904483MR3T 000032E-4-0476 608265MR750W 032E-4-0514 847316MR3T 032E-4-0497 608VAMR750W 032E-4-0546 720848MR4 032E-4-0522 480342MR3 032E-4-0547 423493MR750W 032E-4-0452 318443MR2 032E-4-0448 RADNETMR613 032E-4-0139 510449MR750W 032E-4-0509 212288MR4 notfnd MAYOPETMR01 000032E-4-0463 608890MR750 032E-4-0461 614722MR3 032E40527 617726MR1A2 032E-4-0537 205592750 032E-4-0520 215829MR750W 032E-4-0562 860446MR750W 032E-4-0570 859334MR 032E-4-0574 650PH3T 032E-4-0417 608HERIMR 032E-4-0555 773665SJMR2 032E-4-0394 E416249MR2 032E-4-0594 CHOPPETMR 032E-4-0613 303492MR 901813 864529MR 032E-4-0593 910907WAMR1 032E-4-0644 336397P3T 032E-4-0548 310264MR2 032E-4-0617 815968MR 00000032E40229 702382MR750W 0000034-4-0234 317865MR4 000022E-4-0190 305854MR1 000032E-4-0033 815395MR2 000032E-4-0041 604875MR750 000032E-4-0043 608WIMRMR2 032e-4-0065 RADNETMR1631 000032E-4-0074 808545MR 000032E-4-0076 305461MR1 000032E-4-0084 757827MR2 000032E-4-0099 RADNETMR1639 000032E-4-0101 281274MR2 000032E-4-0108 708346MR3 032E-4-0558 616774MR750W 000032E-4-0111 330344MR750W 000032E-4-0115 RADNETMR1648 000032E-4-0117 703698MR4 000032E-4-0124 856581MR3T 000032E-4-0126 RADNETMR1652 000032E-4-0127 847872ZMR2 000032E-4-0130 270798MR2 000032E-4-0133 757983MR 000032E-4-0134 671FHPMR750W 032E-40051 414647MR1 000032E-4-0146 VAWESTROX750W 000032E-4-0161 217876DMH3T 000032E-4-0163 312355DVMR 000032E-4-0164 808YMMR 032E-4-0065 808873MR750W 000032E-4-0171 6033343TW 000032E-4-0173 856247MR1 000032E-4-0177 301570MR 000032E-4-0180 337289MR1 000032E-4-0185 858622HDMR1 000032E-4-0186 808485MR2 000032E-4-0200 212543MR750 032e-4-0165 713798MR1 000032E-4-0210 415CBMRP 000032E-4-0213 608CHILDMR1 000032E-4-0218 COR526967 000032E-4-0227 513636MR4 000032E-4-0234 217528MR1 000032E-4-0236 602863MR 000032E-4-0242 913322MISHMRW 000032E-4-0243 937641MR2 032E-4-0110 843815750W 000032E-4-0262 715847MR3T 000032E-4-0266 410328MR3T 000032E-4-0292 949760MR3 000032E-4-0300 954355MR750W 000032E-4-0301 843750WGEM 000032E-4-0303 207MMC3TMR 000032E-4-0319 914750W 000032E-4-0321 309655MR750W 000032E-4-0326 703478MR2DV 000032E-4-0333 702731MR750W 000032E-4-0343 386FNSBMR3T 000032E-4-0347 6142343T 000032E-4-0360 317887MR3 000032E-4-0361 865984MR750 000032E-4-0365 703573MR1DV 000032E-4-0377 608262PETMR1 032E-4-0336 506648MR4 032E-4-0399 920738MR750 032E-4-0456 206CH750MR 000032E-4-0401 MSK53MR02 032E-4-0400 503251PMR3 032E-4-0404 MR229049 032E-4-0491 269DVMR 032E-4-0535 720848MR4 032E-4-0503 480342MR3 032E-4-0542 423493MR750W 032E-4-0463 608890MR750 032E-4-0541 859334MR 000032E-4-0009 815395MR2 000032E-4-0239 937641MR2 000032E-4-0299 207MMC3TMR 000032E-4-0325 703478MR2DV 000032E-4-0339 386FNSBMR3T 040-4-0321 920738MR750
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    US Nationwide Distribution
  • Product Description
    NeoCoil 3.0T GEM Flex Coil. || The 3.0T GEM Flex coil system includes three sizes of coil arrays and one interface box. || 3.0T GEM Flex, Interface 16ch Fixed, P -Connector, NC032204, GEHC part number 5430008-5. || 3.0T GEM Flex Coil16-L Array, NC040200, GEHC part number 5430008-2. || 3.0T GEM Flex Coil 16-M Array, NC033000, GEHC part number 5430008-3. || 3.0T GEM Flex Coil 16-S Array, NC034200, GEHC part number5430008-4. || The 3.0T GEM Flex Coil (interface box and arrays) is used in conjunction with a 3.0T GE Magnetic Resonance Scanner
  • Manufacturer

Manufacturer

  • Manufacturer Address
    NeoCoil, LLC, N27W23910 Paul Rd Ste A, Pewaukee WI 53072-6204
  • Manufacturer Parent Company (2017)
  • Source
    USFDA