Recall of Cytomics FC500 MPL with MXP Software versions 1.1 and 2.0

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    32733
  • Event Risk Class
    Class 2
  • Event Number
    Z-1377-05
  • Event Initiated Date
    2005-07-06
  • Event Date Posted
    2005-08-17
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-04-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Counter, Differential Cell - Product Code GKZ
  • Reason
    Software anomally. if the laser is shutdown or fluctuates outside of the specification limits, the software continues to run and will not indicate a laser failure, and could therefore cause erroneous results.
  • Action
    Letters issued to users on July 13, 2005 which also included modified operating instructions, and response cards.