Recall of Coulter AcT 5diff Open Vial Hematology Analyzers

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27251
  • Event Risk Class
    Class 3
  • Event Number
    Z-0039-04
  • Event Initiated Date
    2003-09-08
  • Event Date Posted
    2003-10-15
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2004-03-30
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Counter, Differential Cell - Product Code GKZ
  • Reason
    Samples with high fat content will not count platelets accurately.
  • Action
    Firm issued an advisory letter to users to retain as part of the Quality Manual for the instruments. It cautions to check specimens for interfering lipemia. Letter was sent 9/8/2003.

Device

  • Model / Serial
    All instruements are affected.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    United States and Canada
  • Product Description
    Coulter AcT 5diff Open Vial Hematology Analyzers || Part Numbers: 6605580, 6605581, 6605604