Recall of Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata,

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Siemens Medical Solutions USA, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27661
  • Event Risk Class
    Class 3
  • Event Number
    Z-0175-03
  • Event Initiated Date
    2003-10-16
  • Event Date Posted
    2003-12-03
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2004-08-25
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, Nuclear Magnetic Resonance Imaging - Product Code LNH
  • Reason
    Equipment sar monitor does not accept new patient data, incorrect rf levels may be administered to the patient.
  • Action
    The recalling firm field representatives are visiting each location to correct the problem through a software upgrade.

Device

  • Model / Serial
    Model numbers 4772906, 7106995, 7104693, 5751438, 7106714, 7104719, 7106425, 7104594, 4772971, and 7106557.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    The products were shipped nationwide to hospitals and medical imagining centers.
  • Product Description
    Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata Upgrade, Syngo Symphony, Syngo Symphony Q, and Syngo Symphony Upgrade.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Siemens Medical Solutions USA, Inc, 51 Valley Stream Pkwy, Malvern PA 19355-1406
  • Manufacturer Parent Company (2017)
  • Source
    USFDA