Recall of Candela Smoothbeam Laser System || Model Numbers: 9914-00-0820 || 9914-06-0820 || 9914-07-0820 || 9914-08-0820 || 9914-10-0820 || Units are the same except for color

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Candela Laser Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27954
  • Event Risk Class
    Class 2
  • Event Number
    Z-0291-04
  • Event Initiated Date
    2003-11-25
  • Event Date Posted
    2004-01-01
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2006-12-06
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Powered Laser Surgical Instrument - Product Code GEX
  • Reason
    Laser may pulse and not dispense cryogen resulting in patient injury.
  • Action
    Candela notified direct customers and distributors by letter dated 11125/03 and issued Certified Mail. Accounts are requested to perform two tests on the unit to determine if the unit has the problem. If a unit fails, Candela's Field Service will make the repair. Distributors are requested to notify their customers.

Device

  • Model / Serial
    Serial Numbers:  9914-0820-0101 thru 9914-0820-0602 9914-0820-P2 9914-0820-P4 9914-0820-P5 9914-0820-P6
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide Foreign: Canada,Thailand, Germany, Spain , Japan, Italy, Greece, Saudi Arabia, UK, France, Switzerland, Mexico, Ukraine, Kuwait, Netherlands.
  • Product Description
    Candela Smoothbeam Laser System || Model Numbers: 9914-00-0820 || 9914-06-0820 || 9914-07-0820 || 9914-08-0820 || 9914-10-0820 || Units are the same except for color