Recall of C1141 Processing Container

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Steris Corp.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    26151
  • Event Risk Class
    Class 2
  • Event Number
    Z-0829-03
  • Event Initiated Date
    2003-04-23
  • Event Date Posted
    2003-05-15
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2004-02-05
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Sterilizer, Chemical - Product Code MLR
  • Reason
    The container flow port is inverted, which restricts the flow of sterilant to the endoscope and affects the sterilization process.
  • Action
    The firm notified their customers by telephone and letter on 4/23/2003.

Device

  • Model / Serial
    All codes.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    The product was shipped to end-users located nationwide and worldwide.
  • Product Description
    C1141 Flexible Endoscope Processing Container, designed to fit the Steris System I Processor. The unit may be sold individually or with the processing tray assembly (C1140).
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Steris Corp, 5960 Heisley Rd, Mentor OH 44060
  • Source
    USFDA