Recall of Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex,

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Philips Medical Systems (Cleveland) Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    68245
  • Event Risk Class
    Class 2
  • Event Number
    Z-1722-2014
  • Event Initiated Date
    2014-04-01
  • Event Date Posted
    2014-06-04
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2018-06-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, x-ray, tomography, computed - Product Code JAK
  • Reason
    A customer reported that the table top had become free floating. a field service engineer (fse) confirmed that the service latch had become disengaged.
  • Action
    Philips Healthcare sent an Urgent Medical Device Correction letter dated April 1, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed If they experience a horizontal, free-floating couch motion, to contact their Field Service Engineer immediately. Customers were instructed to retain a copy of this Field Safety Notice with the equipment Instructions for USER Use (IFU). Customers with questions or needing further information were instructed to contact their local Philips representataive or local Philips Healthcare office. For North America, contact the Customer Care Solutions Center at 1-800-722-9377, Option 5. For questions regarding this recall call 440-483-3032.

Device

  • Model / Serial
    Brilliance CT 6: M/N: 728256; S/N: 3001, 3002, 3005, 3007, 3014, 3016, 3019-3021, 3025, 3026, 3028, 3032-3036, 3045, 3046, 3049, 3051, 3052, 3057, 3058, 3060, 3062, 3064-3066, 3068, 3071, 3073, 3074, 3076-3081, 3083, 3084, 3090, 3093-3095, 3097, 3101, 3102, 3104, 3106-3108, 3112-3115, 3118-3121, 3124, 3126, 3128, 3129, 3131, 3132, 3134, 3136, 3137, 3139-3141, 3143-3145, 3147, 3148, 3151-3158, 3162, 3165-3169, 3171, 3173-3177, 3180, 3181, 3184-3190, 3192, 3193, 3196, 3200, 3201, 3203, 3204, 3207, 3208, 3210-3218, 3220, 3221, 3225, 3226, 3228, 3229, 3232, 3235, 3237, 3240, 3241, 3243-3245, 3248, 3249, 3253-3257, 3260-3264, 3266-3272, 3275-3282, 3284, 3285, 3287, 3290, 3292, 3296, 3299, 3301, 3303-3305, 3307-3311, 3313, 3315, 3317-3323, 3325-3328, 3332-3334, 3336, 3338, 3339, 3341, 3343-3345, 3347-3349, 3351, 3352, 3354, 3357-3364, 3366-3368, 3370-3374, 3377-3384, 3386, 3388-3391, 3394, 3396-3400, 3402-3404, 3408, 3409, 3412, 3413, 3415-3418, 3423-3426, 3428-3432, 3434-3437, 3439, 3443, 3444, 3447, 3449, 3451-3453, 3455, 3457, 3459, 3460, 3461, 3463, 3465, 3467-3475, 3477-3483, 3485, 3486, 3488, 3489, 3491-3506, 3508-3511, 3513, 3515, 3517-3519, 3522-3524, 3526, 3527, 3532-3535, 3537-3549, 3551-3553, 3556-3559, 3562, 3564, 3566, 3568, 3570, 3571, 3573-3575, 3577, 3578, 3580, 3582-3588, 3590-3595, 3597-3608, 3611-3616, 3618-3622, 3625-3632, 3634-3637, 3639-3646, 3648, 3649, 3651-3661, 3663-3665, 3667-3670, 3672, 3674-3677, 3679-3693, 3695-3698, 3703-3712, 3714, 3715, 3717-3720, 3722, 6039, 6040, 30003, 30005, 30007-30009, 30011-30018, 30021, 30022, 30024, 30026-30031, 30033-30040, 30042, 30043, 30045-30048, 30050, 30052-30054, 30056-30062, 30064-30066, 30068, 30069, 30072, 30073, 30075, 30077-30080, 30082-30090, 30092, 30094, 30100, 30101, 30104-30108, 30111, 30113, 30117-30119, 30121-30131 & 300043.  Brilliance CT 10: M/N: 728251; S/N: 3008, 3098, 4001, 4003, 4005, 4006, 4009-4012, 4014, 4016, 4020, 4021, 4023, 4026, 4027, 4029-4031, 4035, 4040, 4041, 4044, 4047-4050, 4052-4054, 4058, 4059, 4061, 4063-4067, 30099, 40006-40012, 40014-40018, 40021, 40023, 40024, 40028, 40030 & 40032-40035.  Brilliance CT 16: M/N: 728246; S/N: 3011, 3013, 3018, 3027, 3031, 3038, 3041, 3043, 3070, 3072, 3082, 3085, 3117, 3127, 3149, 3160, 3205, 3216, 3227, 3293-3295, 3302, 3306, 3324, 3336, 3376, 3393, 3466, 3490, 3514, 3521, 3563, 3565, 3624, 3647, 3673, 3721, 4004, 4017, 4018, 4019, 4024, 4025, 4036, 4043, 4055, 5001, 5004-5008, 5010-5021, 5023-5026, 5028-5032, 5034-5038, 5040-5042, 5044-5046, 5049-5055, 5058-5062, 5064, 5066-5070, 5072-5076, 5078, 5081-5086, 5088-5095, 5097-5099, 5101-5105, 5107, 5108, 5110-5118, 5120-5128, 5130-5141, 5143, 5145-5152, 5155, 5156, 5158, 5160-5164, 5166-5170, 5173-5181, 5183-5202, 5204, 5207-5209, 5211-5219, 5221-5226, 5228-5240, 5242, 5243, 5248, 5249, 5252-5254, 5258, 5264-5267, 5269-5275, 5278, 5279, 5281-5283, 5285-5297, 5301-5309, 5311-5313, 5315-5320, 5322, 5323, 5325-5332, 5334, 5336, 5338-5344, 5346-5351, 5353, 5355, 5357, 5359-5362, 5365-5369, 5371, 5372, 5375-5380, 5382, 5384, 5386-5389, 5394, 5396-5400, 5402-5410, 5414-5429, 5432-5435, 5437-5443, 5451, 5452, 5456, 5460, 5468, 5469, 5485, 5489, 5515, 5522, 5531, 5535, 5537, 5543, 5552, 5555, 5557, 5567, 5581-5584, 5586-5588, 5600, 5607, 5621, 5624, 5630, 5642, 5644, 5645, 5651, 5654, 5658, 5663, 5664, 5667, 5675, 5679, 5684, 5685, 5697, 5707, 5709, 5716, 5724, 5725, 5735, 5738, 5743-5745, 5748-5750, 5752, 5754, 5756, 5772, 5774, 5776, 5777, 5780, 5782, 5790, 5793, 5794, 5796, 5801, 5802, 5806, 5807, 5812-5815, 5823-5827, 5830-5834, 5840-5849, 5851, 5852, 5854, 5855, 5857, 5858, 5860, 5861, 5866, 5871-5873, 5880, 5881, 5882, 5884-5886, 5892, 5899, 5901, 5904, 5915, 5921-5923, 5939-5942, 5947, 5948, 5961-5963, 5966-5968, 5976, 5980, 6011, 6012, 6014, 6015, 6028, 6033, 6034, 6036-6038, 6045, 6050, 6068-6071, 6082, 6090, 6095, 6104, 6105, 6108, 6112, 6115, 6126, 6132, 6133, 6161-6165, 6173, 6180, 6221, 6240, 6241-6243, 6245-6257, 6259-6262, 6264-6273, 6275-6277, 6279-6283, 6285, 6287-6304, 6306-6317, 6319-6325, 6327-6364, 6366-6370, 6373-6380, 6382-6387, 6389, 6392-6395, 6397-6416, 6418-6421, 6423-6444, 6447-6458, 6460-6493, 6497-6500, 6502-6532, 6534-6563, 6565-6577, 6580-6590, 6592-6616, 6618-6636, 6638-6640, 6643-6645, 6647, 6648, 6650-6655, 6657, 6658, 6660-6671, 6673, 6675-6718, 6720-6726, 6729, 6733, 6738, 6740-6743, 6745, 6751, 6753-6755, 6757- 6760, 6762-6768, 30019, 30041, 30061, 30069, 30081, 30091, 30093, 30097, 30109, 32287, 40002, 40003, 40020, 40022, 40027, 50002, 50003, 50005, 50007-50013, 50015, 50016, 50018, 50021-50023, 50025-50036, 50038, 50040-50042, 50044-50059, 50061-50070, 50073-50088, 50091-50099, 50101-50107, 50109-50113, 50115, 50116, 50118-50122, 50124-50129, 50131, 50133-50138, 50141-50166, 50168, 50170, 50172, 50175-50182, 50184-50195, 50197-50200, 50202, 50204, 50206-50208, 50210 & 270001.  Brilliance CT 16 Power: M/N: 728240; S/N: 3004, 3046, 3100, 3109, 3236, 5003, 5043, 5047, 5071, 5345, 5356, 5374, 5390, 5395, 5413, 5430, 5455, 5467, 6007, 6009-6011, 6013, 6024, 6026-6028, 6031-6038, 6041-6043, 6045-6049, 6051-6060, 6062-6067, 6069-6072, 6075, 6077-6090, 6093, 6095, 6096, 6099-6113, 6115-6117, 6119, 6121-6126, 6129-6137, 6139, 6141-6151, 6153-6156, 6159, 6161-6163, 50039, 50073, 60001-60006, 60008-60014, 60017, 60022, 60024, 60025 & 60027-60029.  Brilliance CT 40: M/N: 728235; S/N: 4012, 9031, 9032, 9034, 9036, 9039, 9046, 9047, 9050, 9053, 9055, 9057, 9060, 9061, 9076, 9085, 9088, 9094, 9095, 9097, 9103, 9109, 9111, 9113-9116, 9119, 9122, 9124, 9125, 9128-9134, 9136, 9141, 9144-9147, 9149, 9150, 9153, 9154, 9157, 9158, 9161, 9164, 9165, 9167-9170, 9172, 9174, 9176, 9178-9183 9185, 9186, 9188, 9189, 9191-9194, 9198-9200, 9203,9204, 9207,9209-9212, 9219, 9222, 9225-9233, 9237-9241, 9244, 9245, 9247-9251, 9253, 9254, 9256-9258, 9260-9273, 9275, 9276, 9278-9282, 9284-9286, 29001, 29004, 29006-29008, 29010-29013, 29015, 29019-29024, 29028-29031, 29033, 29036, 29038-29041, 29043, 29046-29048, 29050, 29051, 29055, 29058-29061, 29063-29067, 29069, 29071-29073, 29075, 29077, 29078, 29080, 29082-29087, 29089-29091, 29093-29096, 29098-29106, 29108, 29109, 29113, 29114, 29116-29123, 29125, 29127-29132, 29134-29137, 29139,29140-29145, 29147-29150, 29152, 29153, 90034, 90035, 90040, 90043, 90044, 90049, 90051-90054, 90056, 90059, 90061, 90064, 90065, 90067, 90069, 90071, 90075, 90076, 90083, 90084, 90087, 90090, 90093, 90095, 90101, 90103, 90108-90110, 90121, 90122, 90125, 90126, 90130, 90138, 90142, 90144, 90146, 90148, 90149, 90155, 90157-90159, 90164-90166, 90169, 90171-90174, 90178, 90179, 90181, 90183, 90184, 90186, 90189, 90190, 90192-90194, 90196-90198, 90200, 90202-90204, 90206-90215, 90219, 90220, 90222, 90224, 90226, 90228, 90229, 90231-90237 & 92034.  Brilliance CT 64: M/N: 728231; S/N: 4001, 4003, 4016, 4020, 5006, 9029, 9030, 9033-9035, 9037, 9038, 9040, 9041, 9043-9045, 9048, 9049, 9051, 9052, 9054, 9056, 9058, 9059, 9062, 9064-9075, 9077, 9080, 9083, 9086, 9087, 9089, 9092-9094, 9098, 9099, 9104-9108, 9112, 9117, 9118, 9120, 9121, 9126, 9127, 9135, 9137, 9138, 9140-9143, 9148, 9151, 9152, 9155, 9156, 9159, 9160, 9162, 9163, 9166, 9171, 9173, 9175, 9177, 9187, 9190, 9195-9197, 9201, 9202, 9205, 9206, 9208, 9213-9218, 9235, 9236, 9243, 9255, 9259, 9501-9505, 9507-9512, 9514-9531, 9533-9536, 9538-9543, 9545-9552, 9554-9566, 9568, 9569, 9570, 9572-9585, 9587-9596, 9598-9608, 9610, 9612-9620, 9622-9655, 9657-9659, 9662-9668, 9670-9674, 9676-9680, 9683-9695, 9697-9703, 9705-9714, 9716, 9718-9720, 9722-9726, 9728-9732, 9734-9742, 9745-9761, 9763, 9765-9769, 9772-9782, 9784-9795, 9797, 9800, 9801, 9803-9813, 9815-9822, 9824, 9826, 9829, 9830, 9832-9835, 9837-9847, 9849-9853, 9858-9871, 9873-9879, 9881-9885, 9888-9896, 9898, 9899, 9901-9906, 9909-9911, 9913-9927, 9929-9934, 9936-9938, 9940-9942, 9944-9948, 9950-9956, 9958-9961, 9963-9971, 9974-10009, 10011-10028, 10031-10037, 10039-10042, 10044-10054, 10056-10073, 10075, 10077-10081, 10083-10090, 10092, 10094-10108, 10110-10132, 10134-10137, 10139-10146, 10148-10158, 10160-10162, 10164-10168, 10170-10174, 10176-10198, 10200-10235, 10237-10239, 10241-10245, 10247, 10248, 10250-10283, 10286-10288, 10291-10305, 10307-10319, 10321-10327, 10330-10339, 10341-10353, 10355, 10357-10381, 10383-10386, 10388-10402, 10404-10422, 10424-10432, 10435-10492, 10494-10500, 10502-10523, 10525-10528, 10530-10535, 10537-10543, 10545-10619, 10621-10671, 10673, 10675-10681, 10683-10689, 10691-10708, 10710, 10712-10715, 10717, 10718, 10720-10745, 10747-10762, 10765, 10767-10786, 10789, 10790, 10792, 10793, 10795-10804, 29002, 29003, 29005, 29014, 29016-29018, 29025, 29026, 29032, 29034, 29035, 29037, 29044, 29045, 29049, 29053, 29054, 29056, 29057, 29062, 29066, 29068, 29070, 29092 ,29101, 29107, 29113, 29115, 29124, 29138, 29151, 32129, 64015, 64016-64020, 64022-64039, 64041-64080, 64082-64094, 64096-64104, 64106-64112, 64115-64118, 64121-64131, 64133-64140, 64142, 64143-64154, 64157, 64158, 64160, 64162-64169, 64171-64175, 65006, 90029-90031, 90036, 90037, 90041, 90042, 90046-90048, 90057, 90058, 90060, 90062, 90063, 90066, 90070, 90072, 90074, 90077-90082, 90084-90089, 90091, 90092, 90094, 90096-90100, 90105-90107, 90117-90120, 90123, 90124, 90129, 90131, 90132, 90135-90137, 90139-90141, 90143, 90147, 90150-90154, 90156, 90160, 90161, 90163, 90167-90170, 90175-90177, 90180, 90182, 90187, 90195, 90199, 90205-90208, 90213, 90218, 90221-90223, 91003, 95003, 95008, 95009, 95011-95024, 95029-95052, 95054-95069, 95071-95076, 95078-95085, 95088-95095, 95097-95104, 95106-95114, 95116-95119, 95121-95123, 95125-95129, 95132-95143, 95147, 95149, 95151, 95153, 95154, 95157, 95161, 95162, 95165-95170, 95177, 95178, 95181-95186, 95189-95191, 95194, 95195, 95197, 95199, 95201-95204, 95207-95210, 95214, 95216-95219, 95221-95223, 95227, 95228, 95231-95234, 95236-95239, 95242, 95243, 95245-95247, 95249-95254, 95257-95259, 95261, 95262, 95264-95266, 95268-95271, 95273-95287, 95289-95294, 95296-95298, 95300, 95301, 95303-95305, 95308-95310, 95312-95335, 95338, 95342, 95344-95346, 95348-95351, 95353-95366, 95369, 95370, 95374-95376, 95378-95400, 95402-95404, 95406, 95408-95411, 95413-95436, 95438-95440, 95442-95458, 95460-95492, 95494-95529, 95531-95584, 95586, 95588-95592, 95595-95641, 95643-95662, 95664-95700, 95702-95727, 95729-95761, 95764-95770, 95772-95812, 95814-95832, 95835-95845, 95848-95858, 95860-95862, 95864-95876, 95878, 95879, 95884, 95920, 96494, 100062, 110058, 110391 & 999999.   Brilliance CT Big Bore Oncology: M/N: 728243; S/N: 7006-7009, 7013-7051, 7053-7062, 7064-7120, 7122-7142, 7144-7166, 7168, 7169, 7171, 7173-7175, 7177-7179, 7181, 7183-7186, 7188, 7189, 7191, 7192, 7194-7199, 7201-7214, 7216-7243, 7245-7247, 7249, 7250, 7253-7263, 7265-7267, 7269, 7270, 7272-7289, 7291-7295, 7297-7300, 7302, 7304, 7306-7316, 7319, 7320, 7322-7374, 7376-7379, 7381,7382, 7387-7389, 7392, 7393, 7395, 7397-7401, 7403-7406, 7409,7410,7411-7422, 7424-7426, 7428, 7429, 7431-7433, 7435-7438, 7440-7451, 7454-7457, 7459, 7461-7471, 7473-7475, 7477-7485, 7487-7509, 7511-7513, 7515-7524, 7526-7528, 7530-7532, 7534, 7535, 7537, 7539-7542, 7545, 7547-7556 ,7558-7561, 7563-7569, 7573, 7574,7577-7582, 7584, 7585, 7588-7593, 7596-7599, 7601-7603, 7605, 7606, 7608, 7609, 7611-7613, 7615-7618, 7620, 7622-7624, 7626-7630, 7634, 7635, 7637-7650, 7652, 7655-7659, 7661, 7662, 7665-7667, 7671, 7672, 7674, 7675, 7677-7681, 7684, 7685, 7687, 7689-7691, 7693, 7695, 7696, 7698-7707, 7709, 7715, 7716, 7718, 7720, 7724-7727, 7729-7731, 7735-7737, 7739-7741, 7743-7745, 7749-7755, 7757, 7759, 7761-7763, 7765, 7769, 7770, 7774-7777, 7779, 7782-7788, 7791, 7793-7797, 7799-7803, 7806-7808, 7810-7812, 7814, 7815, 7817, 7818, 7820, 7822-7824, 7827-7829, 7831-7838, 7840-7842, 7846-7852, 7855-7858, 7860, 7861, 7864-7868, 7872, 7873, 7875-7877, 7881, 7884, 7886, 7891-7893, 7895, 7896, 7898, 7900-7902, 7904, 7905, 7907-7911, 7914, 7916, 7918, 7920, 7921, 7923-7925, 7927-7929, 7931-7933, 7935-7942, 7944-7946, 7948-7950, 7955, 7956, 7963, 7968, 7970, 7972, 7973, 7976, 7984, 7987, 7990, 7991 & 7993.  Brilliance CT Big Bore Radiology: M/N: 728244; S/N: 7052, 7063, 7121, 7155, 7172, 7176, 7180, 7182, 7190, 7200, 7244, 7252, 7268, 7271, 7301, 7318, 7375, 7380, 7383, 7384, 7386, 7390, 7396, 7402, 7407, 7408, 7427, 7430, 7439, 7452, 7453, 7458, 7486, 7510, 7514, 7525, 7529, 7533, 7538, 7543, 7546, 7570, 7572, 7575, 7583, 7586, 7587, 7594, 7595, 7604, 7610, 7614, 7619, 7621, 7625, 7631, 7632, 7633, 7636, 7651, 7653, 7654, 7660, 7668-7670, 7676, 7683, 7686, 7688, 7692, 7694, 7697, 7708, 7710-7714, 7717, 7719, 7721-7723, 7732-7734, 7738, 7742, 7746, 7747, 7756, 7760, 7764, 7766, 7767, 7771-7773, 7778, 7780, 7781, 7789, 7790, 7798, 7804, 7809, 7813, 7819, 7825, 7826, 7830, 7839, 7844, 7845, 7853, 7859, 7862, 7869, 7870, 7874, 7878, 7879, 7880, 7882, 7887, 7888, 7894, 7897, 7899, 7903, 7906, 7912, 7913, 7915, 7917, 7919, 7922, 7926, 7930, 7934, 7943, 7952,7958, 7959, 7964, 7967, 7986, 7988 & 7992.  Brilliance iCT: M/N: 728306; S/N: 85000, 100016, 100019-100025, 100028, 100031, 100032, 100037-100040, 100042-100074, 100077-100079, 100081-100084, 100086-100106, 100108-100135, 100137-100142, 100144-100155, 100157-100182, 100184-100236, 100238, 100240, 100242-100254, 100256-100268, 100270-100278, 100280-100283, 100285-100297, 100299-100325, 100327, 100329-100332, 100334, 100336, 100338-100343, 100345-100350, 100352-100358, 100360-100365, 100368-100380, 100382-100389, 100391, 100393-100402, 100404-100412, 100414-100418, 100423-100426, 100429, 100430, 100433-100435, 100438-100442, 100449-100458, 100460-100462, 100464-100470, 100473, 100476, 100477, 100480-100482, 100484-100486, 100488, 100490, 100493, 100494, 100497-100499, 100503, 100505-100508, 100511, 200005, 200083, 200098 & 987654.  Brilliance iCT SP: M/N: 728311; S/N: 100039, 100062, 200003, 200006-200016, 200018-200027, 200029-200058, 200060-200065, 200067, 200068-200071, 200073-200079, 200084-200097, 200099-200117, 200119, 200120, 200123-200131, 200133-200137.  Ingenuity CT: M/N: 728326; S/N: 6372, 6395, 6569, 10607, 300005, 300011, 300012, 300014, 300016-300022, 300024, 300025, 300027-300042, 300044-300066, 300068, 300069, 300071-300075, 300077-300082, 300084-300114, 300116-300133, 300135-300140, 300142-300145, 300147-300151, 300156-300163, 300164 & 310058.  Ingenuity Core: M/N: 728321; S/N: 6563, 7700, 10614, 10621, 10633, 50171, 50204, 95732, 100507, 310001, 310004, 310006, 310008, 310009, 310010, 310012-310015, 310017, 310018, 310020-310022, 310024-310026, 310028-310035, 310037-310057, 310059-310143, 310146-310148, 310156, 310157, 310159, 333001-333003, 333005-333012, 333016 & 3100016.  Ingenuity Core128: M/N: 728323; S/N: 100315, 320002-320011, 320013-320015, 320017-320112, 320115, 320119, 320123, 320126, 320131-320133, 320135, 320136, 320140, 336001-336003, 336006-336008 & 336014.  MX8000 Dual v. EXP: M/N: 728130; S/N: 2110, 8010, 8016, 8018, 8034, 8035, 8039-8041, 8043, 8048, 8053, 8055, 8056, 8058, 8061, 8065, 8067, 8068, 8070, 8073-8077, 8079, 8081, 8082, 8084-8086, 8092, 8095, 8097, 8098, 8104, 8105-8108, 8111, 8112, 8114-8120, 8123-8125,8127-8131, 8134-8138, 8140, 8142-8147, 8149-8151, 8153-8156, 8159, 8160, 8163-8165, 8167-8169, 8171-8175, 8177, 8178, 8181, 8184-8186, 8188, 8190-8192, 8194-8196, 8198, 8199, 8201-8208, 8211-8213, 8215-8218, 8220, 8221, 8223, 8225-8227, 8232-8240, 8242, 8243, 8245, 8247-8250, 8253, 8256-8258, 8260-8289, 8292-8296, 8298-8306, 8308, 8310-8315, 8317, 8318, 8320-8324, 8326, 8328-8335, 8337-8342, 8344-8347, 8349, 8351-8355, 8357-8359, 8361-8373, 8375-8396, 8398, 8400-8407, 8409-8413, 8442, 9528, 80002-80005, 80007, 80009-80019, 80021 & 80022.  Ingenuity Flex: M/N728317; S/N: 260002, 260003, 340002, 340003, 345002, 345003 & 345004.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution: US nationwide including PR and DC, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Canada, Cambodia, Cape Verde, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dubai, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Greenland, Guadeloupe, Herzegovina, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, LAO People's Dem, Rep., Latvia, Lebanon, Libya, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nahar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam & Yemen.
  • Product Description
    Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, and the MX8000 Dual v. EXP scanners)
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Philips Medical Systems (Cleveland) Inc, 595 Miner Rd, Cleveland OH 44143-2131
  • Manufacturer Parent Company (2017)
  • Source
    USFDA