Recall of Biomet Orthopedics Apical Hole Plug

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Biomet, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    44812
  • Event Risk Class
    Class 2
  • Event Number
    Z-0102-2008
  • Event Initiated Date
    2007-08-09
  • Event Date Posted
    2007-10-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2009-06-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Apical Hole Plug - Product Code LPH
  • Reason
    Incorrect or lack of threading: the plug component was manufactured incorrectly and does not fit with the screw hole as intended.
  • Action
    The distributor was notified via phone on or about 8/9/07 and by letter dated 8/31/07 and requested to return the product. Returns included subaccounts

Device

  • Model / Serial
    Lot 522080.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Nationwide
  • Product Description
    Biomet Orthopedics Apical Hole Plug, 3/8 - 24 threaded, TI-6AL-4V alloy, sterile; REF 123741, Biomet Orthopedics, Inc., Warsaw, IN 46581
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Biomet, Inc., 56 E Bell Dr, Warsaw IN 46581-0587
  • Source
    USFDA