Recall of Biomet KWire/Trocar Point

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Biomet, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    53580
  • Event Risk Class
    Class 2
  • Event Number
    Z-0367-2010
  • Event Initiated Date
    2009-08-06
  • Event Date Posted
    2009-11-19
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2010-10-06
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    K-wire, Trocar Point - Product Code FZX
  • Reason
    Packages labeled as having a .045 inch diameter trocar point actually contain a diamond point with a diameter of .062 inches.
  • Action
    An "Urgent Medical Device Recall Notice" dated August 6, 2009 was issued to consignees. The letter described the affected product, issue and actions for consignees. Customers were instructed to complete and return the "FAX back Response Form" (fax number 1-574-372-1683) and all affected product in inventory to Biomet Orthopedics. Direct questions related to the recall notification by calling at 1-800-348- 9500 or 1-574-372-3983, Monday through Friday, 8AM to 5PM.

Device

  • Model / Serial
    Lot Number 087190.
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    United States (Florida and Pennsylvania).
  • Product Description
    Biomet K-Wire/Trocar Point, both ends/Dia. 045, length: 9 in., certified stainless steel, sterile, Biomet Orthopedics, Inc., Warsaw, IN; REF 35-351534. || Internal fixation devices intended to aid in the alignment and stabilization of fractures to the skeletal system until healing has occurred. The devices can also be used as instrument accessories to stabilize cut blocks and can serve as guide wires while using reamers.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Biomet, Inc., 56 E Bell Dr, Warsaw IN 46582
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    USFDA