Recall of BD Biosciences FACSDiva Software Versions 4.0.1 to 5.0.2

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by BD Biosciences.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    38337
  • Event Risk Class
    Class 2
  • Event Number
    Z-1525-2008
  • Event Initiated Date
    2007-04-05
  • Event Date Posted
    2008-08-16
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2009-09-30
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Differential Cell Counter - Product Code GKZ
  • Reason
    When data file containing one or no fluoresence parameters is exported, the software will automatically apply compensation to this file and all subsequently exported files.
  • Action
    The firm has issued a customer Important Product Information letter via mail on 04/20/2007 to inform them of the software issues. The letter states that to avoid this problem, make sure each tube or well has at least two fluorescence parameters before exporting FCS 3.0 data files. Manually exporting tubes or wells one at a time or restarting the software after the export of an FCS data file containing one or no fluorescence parameters will prevent the problem from occurring. Customers are requested to fill out and return the response form and to contact BD Customer Support Center if there are any additional questions. A software fix is being developed in version 6.0 and a free maintenance release with no product enhancments (version 5.0.3).

Device

  • Model / Serial
    Software versions 4.0.1 to 5.0.2.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution --- USA including states of NY, NM, MI, GA, MD, MA, AL, CA, VA, NC, OH, TX, CO, NJ, PA, MI, CT, IA, MO, NV, KY, IL, TN, FL, MN, FL, UT, WA, OK, CT, WI, SD, AZ, and IN, and countries of Buenos Aires, BE, Brazil, Canada, Chile, China, Hong Kong, Japan, Singapore, Taiwan, Venezuela, Mexico, Trinidad and Tobago, and Columbia.
  • Product Description
    FACSDiva Software Versions 4.0.1 to 5.0.2, Flow Cytometry Acquisition & Analysis software. Catalog number 641609 || Flexible data acquistion and analysis package specifically designed for digital-based flow cytometers.
  • Manufacturer

Manufacturer