Recall of Animas IR 1250 Insulin Pump

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Animas Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    31760
  • Event Risk Class
    Class 2
  • Event Number
    Z-0929-05
  • Event Initiated Date
    2005-04-11
  • Event Date Posted
    2005-06-25
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2006-01-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Pump, Infusion, Insulin - Product Code LZG
  • Reason
    Pump failed to respond to keypad button pressess after using the food database feature.
  • Action
    The recalling firm issued a recall notice dated 4/11/05 to all patients and healthcare providers via first class mail. The letter instructs the patient to reset the pump, thus correcting any problems caused by the software bug, and to discontinue use of the food database. All distributors were contacted by phone on 4/12/05 and instructed to return all pumps in their inventory. On 4/25/05, the recalling began emailing and telephoning patients to arrange for replacement and return of the devices.

Device

  • Model / Serial
    Model IR1250
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    The products were shipped to distributors in FL, TX, UT, and VA. The products were also shipped to patients nationwide.
  • Product Description
    Animas IR 1250 Insulin Pump.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Animas Corporation, 200 Lawrence Dr, West Chester PA 19380-3428
  • Source
    USFDA