Recall of ACCU-CHEK Inform meter; catalog No. 3035123.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Diagnostics Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    30462
  • Event Risk Class
    Class 2
  • Event Number
    Z-0372-05
  • Event Initiated Date
    2004-11-19
  • Event Date Posted
    2004-12-28
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2005-06-07
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, Test, Blood Glucose, Over The Counter - Product Code NBW
  • Reason
    A firmware bug may allow test results from other systems, recorded into the inform system via the ote functionality, to be reversed from positive to negative, and vice versa.
  • Action
    Consignees were notified via a letter dated 11/19/04, which instructed them to discontinue using the OTE functionality.