Field Safety Notices about SynReam Flexible shaft

According to Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), this field safety notices involved a device in Slovenia that was produced by Synthes GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Date
    2016-11-11
  • Event Country
  • Event Source
    AMPMDRS
  • Notes / Alerts
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Extra notes in the data
  • Reason
    Synthes has determined there is a potential the affected lots of the synream flexible shaft (352.040 and 352.044) may have been manufactured with a non-symmetric hexagonal coupling which may result in the reamer shaft not fitting into the reamer heads. the synream flexible shafts are intended for use as an optional reaming step in intramedullary nailing systems. .

Device

  • Model / Serial
    SynReam Flexible Shaft 352.040 9946822, 9947895, 9956527, L084498 Flexible Shaft Ø 7.0mm, long 352.044 9916436, 9916503, 9916508, L000158, L000160, L000573, L000603, L000604, L000605, L000607, L000632, L009405, L030364, L044313, L082974
  • Manufacturer

Manufacturer