Field Safety Notices about SOMATOM Definition AS

According to Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), this field safety notices involved a device in Slovenia that was produced by Siemens Healthcare GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Date
    2016-10-04
  • Event Country
  • Event Source
    AMPMDRS
  • Notes / Alerts
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Extra notes in the data
  • Reason
    We have found that the laser beam window (see figures 1a and 1b), which is used for laser positioning and is integrated into the front bezel of your ct device, can in rare cases become loose and potentially fall off. in this case, there is a risk that the person reaching the window opening may touch rotating or electrical parts. gantry. such a person may be seriously injured, e.G. electrocution or damage to your fingers.

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    AMPMDRS