Field Safety Notices about SANO Liftkar PT electric stairclimber

According to Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), this field safety notices involved a device in Slovenia that was produced by SANO GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Date
    2017-01-09
  • Event Country
  • Event Source
    AMPMDRS
  • Notes / Alerts
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Extra notes in the data
  • Reason
    Due to an incident with one handle (cracks in welding area of handle unit, which resulted in a breakage) decision was taken by sano to inspect handles of above defined serial numbers. in order to keep downtime of device low, sano will send sufficient new handles to its partners for replacement. inspection of replaced handles needs to be done by sano after return of parts to its premises in austria.

Device

  • Model / Serial
    from 20061377 to 20105770
  • Manufacturer

Manufacturer

  • Source