Field Safety Notices about iPlan RT Image (below v.4.1.2) / iPlan RT; and BrainSCAN (discontinued)

According to Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), this field safety notices involved a device in Slovenia that was produced by Brainlab.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Date
    2015-09-07
  • Event Country
  • Event Source
    AMPMDRS
  • Notes / Alerts
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Extra notes in the data
  • Reason
    We are writing to advise you of a requirement of the brainlab radiation treatment planning software. except for ct or pet normalized to suv (standard uptake value), the brainlab software supports only image sets that contain a uniform calibration of pixel values assigned by the imaging device (scanner) for all images, i.E. for all slices of the set. .

Manufacturer