Field Safety Notices about Extraction Screw for PFNA Blade

According to Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), this field safety notices involved a device in Slovenia that was produced by Synthes GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Event Country
  • Event Source
    AMPMDRS
  • Notes / Alerts
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Extra notes in the data
  • Reason
    Through a complaint investigation, it was discovered that the affected lots of the extraction screw could break or disassemble from the instrument. this may interfere with removal of the pfna and pfna-ll blade from the patient.

Device

  • Model / Serial
    2089467; 2090743; 2101443; 2103560; 2107022; 2107094; 2109115; 2114311; 2119302; 2125711; 2127225; 2135962; 2145558; 2155648; 2162089; 2163636; 2168449; 2186602; 2187918; 2193852; 2203709; 2216008; 2225226; 2225992; 2234804; Extraction Screw for PFNA Blade 356.825 2238532; 2238533; 2257480; 2257481; 2263698; 2265892; 2311442; 2324437; 2354073; 2357959; 2370086; 2389952; 2401175; 2416428; 2442559; 2443570; 2450764; 2458359; 2475991; 2484015; 2485474; 2498270
  • Manufacturer

Manufacturer