  • Type of Event
    Field Safety Notice
  • Event Country
  • Event Source
    AMPMDRS
  • Notes / Alerts
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Extra notes in the data
  • Reason
    Axiom artis semiautomatic focal point switch with vb35e software version and megalix x-ray tubes may not work properly if the "little" and "micro" focal points are damaged when the footswitch is pressed repeatedly. this particular type of focus failure that prevents semi-automatic focus shifting is more likely with a longer x-ray tube life.