Field Safety Notices about ADVANCED Proximal Femoral Nailing System (TFNA)

According to Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), this field safety notices involved a device in Slovenia that was produced by Synthes GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Date
    2015-04-17
  • Event Country
  • Event Source
    AMPMDRS
  • Notes / Alerts
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Extra notes in the data
  • Reason
    There is the potential that the tfna nails received before march 31, 2015 from the affected lots were assembled with a locking mechanism too close to the top of the nail. the locking mechanism could: . prevent the connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the insertion handle, or . cause the bottom of the connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.

Manufacturer