Field Safety Notices about Endo GIA Auto Suture Universal Loading Unit

According to ALIMS (Agencija za lekove i medicinska sredstva), this field safety notices involved a device in Serbia that was produced by Covidien IIc.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Event Date Posted
    2013-02-20
  • Event Country
  • Event Source
    ALIMSA
  • Event Source URL
  • Notes / Alerts
    Serbian data is current through February 2018. All of the data comes from ALIMS (Agencija za lekove i medicinska sredstva), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Serbia.
  • Extra notes in the data
  • Reason
    The manufacturer has received a report of an adverse reaction to a medical device in which each linear stapler cartridge is filled with two buckles. this can lead to difficulty in using the stapler and difficulty removing the agent from the application site, which may require additional medical intervention.

Device

  • Model / Serial
    N2F0563LX, N2F0341LX, N2F0725LX, N2E0605LX
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    ALIMSA