translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Ortho-Clinical Diagnostics.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 111 06 11 000
  • Event Initiated Date
    2011-06-27
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Ortho clinical diagnostics (ocd) has received customer complaints of intermittent imprecision, outliers, or shifts in control fluid values when using various lots of vitros chemistry products phbr slides (phenobarbital) from coating 0053. ocd investigation has confirmed that biased values for patient samples could be obtained that may or may not be detected by quality control fluids.

Device

  • Model / Serial
    VITROS Chemistry Products PHBR Slides Lots Manufactured from Coating 0053 (Product Code 8221384). Lot No: 2532-0053-0144 , 2532-0053-0149 , 2532-0053-0217 , 2532-0053-0615 , 2532-0053-0616 , 2532-0053-1472 , 2532-0053-1488 , 2532-0053-1489 , 2532-0053-1490 , 2532-0053-4961 , 2532-0053-4984 , 2532-0053-4996 , 2532-0053-4997.
  • Product Description
    VITROS Chemistry Products PHBR Slides (phenobarbital) from Coating 0053
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Samir Photographic Supplies Co. Ltd.
  • Source
    SFDA