translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by GE Healthcare.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 009 03 11 000
  • Event Initiated Date
    2011-03-02
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company
  • Reason
    The plastic enclosure on the external power supply for the trusat pulse oximeter may break, exposing the electrical board and line voltage. the exposed electrical board and line voltage poses a risk to the user or caregiver if they touch the power supply while it is plugged in. touching the internal parts of a broken power supply while plugged in may result in a burn or electrical shock.

Device

  • Model / Serial
    All external power supplies used with TruSat Pulse Oximeters sold under the GE or Datex-Ohmeda brand name.
  • Product Description
    Pulse oximeter, line-powered
  • Manufacturer

Manufacturer