According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by EOS Imaging.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 025 08 18 000
  • Event Initiated Date
    2018-08-07
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    When the operator applies a horizontal flip to an image belonging to a series of more than two images, the information relating to the symmetry is not correctly recorded when saved. this impacts the 3d calibration matrix used to display the projections of objects onto the images. projection errors may lead to errors in the calculation of 3d clinical parameters when adjusting the position, size and shape of the 3d objects used.

Device

  • Model / Serial
    The issue concerns all sterEOS software versions.
  • Product Description
    Radiology DICOM image processing application software
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Bio Standards
  • Source
    SFDA