translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Synthes Inc.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 063 11 12 000
  • Event Initiated Date
    2012-11-17
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCAR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    The non-sterile power module is inserted into the sterile hand piece and then secured and locked by the sterile lid to maintain sterility of the device. synthes received one complaint stated, that power module (05.001.202) came out of the trs recon hand piece (05.001.240) as the lid (05.001.241) opened up. the locking mechanism of the lid did loosen which resulted in an unintended release of the power module during one surgical procedure in june 2012.

Device

  • Model / Serial
    Serial no.: 1001-1046 Recon Saw
  • Product Description
    surgical drilling, power module, Rechargeable
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Isam Economic Co., Riyadh, (01) 4600846
  • Source
    SFDA