According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Philips Healthcare.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 025 01 15 000
  • Event Initiated Date
    2015-01-05
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    The mrx could reboot once in any operating mode if the following sequence occurs: 1. in service mode, the status log* is cleared 2. prior to exiting service mode, the mrx generates an entry in the status log. 3. service mode is exited. the mrx is then in a state that could trigger a reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed. the mrx will restart within 8 seconds.

Device

  • Model / Serial
    Product: Philips HeartStart MRx Monitor/Defibrillators Units Affected: MRx units with a serial number within the following ranges: -Model M3535A: US00100100 to US00578696 -Model M3536A: US00100902 to US00576650 -Model M3536J: US00209838 to US00332675 -Model M3536M: US00500002 to US00553553 -Model M3536MC: US00500001 to US00500087 -Model M3536M2: US00554176, US00554177, US00554178 -Model M3536M4: US00500003 to US00574869 -Model M3536M5: US00500001 to US00562935 -Model M3536M6: US00554358 to US00576619
  • Product Description
    Monitor/Defibrillator
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Al-Faisaliah Medical System, Riyadh, (01) 2119948