According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by GE Medical Systems.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 104 03 12 000
  • Event Initiated Date
    2012-03-25
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    During scanning of a patient, if an improperly torqued gradient connection exists, it may cause the gradient lead to produce excessive heat. and if a second and compounding issue of improperly routed rear pedestal air supply hose is also present, routed in such a way to cause the hose to come in contact with the overheated gradient lead cable, then the hose may generate smoke.

Device

  • Model / Serial
    Model number: 5410917 MR450W
  • Product Description
    MRI, systems, RF
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    GE Healthcare, Riyadh, (01) 4600530
  • Source
    SFDA