translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Elekta Limited.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 125 05 14 000
  • Event Initiated Date
    2014-05-27
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Mosaiq does not re-calculate the agent volume when the user changes the ordering dose (to other than what was calculated) prior to saving the order. incorrect agent volume information is displayed in the dispensing detail widget. this information could lead to the patient being under or overdosed. there is a remote probability of a mistreatment resulting in serious injury.

Device

  • Model / Serial
    all released builds of version 2.50 from 2.50.04D5 and higher.
  • Product Description
    Radiation Treatment Planning System Software
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Analysis Device For Medical and Scientific Service ( ADMSS )
  • Source
    SFDA