translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Philips Healthcare.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 088 04 14 000
  • Event Initiated Date
    2014-04-15
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Under very specific error conditions the system executes an unintended exposure: - the handswitch is erroneously stuck in pressed position, while the system is switched off, - the system is powered up with the handswitch in that erroneous condition, - the user logs in, selects a patient for examination, gets a green ready light in the examination context, if the tube arm is then moved out of the zero position, the system is doing an exposure.

Device

  • Model / Serial
    MobileDiagnost wDR worldwide except US
  • Product Description
    X-ray, digital acquisition systems, direct digital radiography.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Al-Faisaliah Medical System, Riyadh, (01) 2119948
  • Source