According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Medtronic Navigation Inc.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 132 07 13 000
  • Event Initiated Date
    2013-07-21
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Medtronic navigation, inc. has identified that the use of high resolution exams (greater than 256 x 256) under certain use conditions may cause the look‐ahead view labels to show user selected depths (displayed), but the corresponding images are shown at a depth twice the selected distance. under these conditions, visualization of the planned trajectory could show the image at the incorrect depth. further internal investigation of similar or related views discovered that the use of high resolution exams (greater than 256 x 256) under certain use conditions may cause a similar discrepancy between the exam image and the displayed measurement or overlay.

Device

  • Model / Serial
    The part numbers included in this corrective action are as follows: (1) Part Number 9734047 FrameLink S7 Applications Kit (2)Part Number 9734066 FrameLink S7 Planning Station Kit (3)Part Number 9733986 FrameLink Application Software
  • Product Description
    Navigation system, Planning Station
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Manufacturers representative
    Al Amin Medical Instruments Co. Ltd., Jeddah, (012) 6601149
  • Source
    SFDA