translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by DiaMed GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 069 05 13 000
  • Event Initiated Date
    2013-05-20
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    An error relative to the liquid reagents management (test cells or anti-serum vials) might occur if the following conditions are present: 1. ih-1000 is in status "ready to start" and, 2. the user requests the unloading of a reagent rack while the system is creating the internal batch, the time slot during which the error could occur is about 2 seconds. 3. another rack is loaded immediately afterwards in the same position. if all three conditions are present, the ih-1000 may not use the full set of liquid reagents required to perform the tests of the ongoing batch. due to a software failure, the system may only use the vial placed in the first position of the re-ioaded rack for all the tests of the ongoing batch.

Device

  • Model / Serial
    Cat.No.001000 SW version : All SN : see attached
  • Product Description
    Fully Automated Blood Grouping Machine IH-1000
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    Al Nahir Trading co , Riyadh , (011)4770444
  • Source
    SFDA