translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by Hospira Inc.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 013 10 11 000
  • Event Initiated Date
    2011-10-08
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    Hospira states that it has received reports of incidents in which the above infusion pumps continuously rebooted and/or recycled when the "backlight intensity" and/or "display contrast" setting for the liquid crystal display (lcd) display was adjusted from the original default setting.

Device

  • Model / Serial
    list 11005, serial numbers 0016792207 through 0018807700 a) list 11971- serial numbers 0013120799 through 0099072300; b) list 11973 - serial numbers 0012570058 through 0012573663; c) list 12391 - serial numbers 0000073363 through 0999948294; d) list 20679 - serial numbers 0015441179 through 0015459047; e) list 20792 - serial numbers 0016061555 through 0026347395 a) list 12348 - serial numbers 0013740398 through 0013741025; b) list 12618 - serial numbers 0014040007 through 0018857420; c) list 20678 - serial numbers 0015543071 through 0019655471
  • Product Description
    Infusion Pumps, General-Purpose
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    AL-KAMAL Import, Riyadh, (01) 2181122
  • Source
    SFDA